High-Precision Electrical Impedance Tomography Data Acquisition System for Brain Imaging

被引:100
|
作者
Shi, Xuetao [1 ]
Li, Weichen [1 ]
You, Fusheng [1 ]
Huo, Xuyang [2 ]
Xu, Canhua [1 ]
Ji, Zhenyu [1 ]
Liu, Ruigang [1 ]
Liu, Benyuan [1 ]
Li, Yandong [1 ]
Fu, Feng [1 ]
Dong, Xiuzhen [1 ]
机构
[1] Fourth Mil Med Univ, Sch Biomed Engn, Dept Biomed Engn, Xian 710032, Shaanxi, Peoples R China
[2] Jilin Med Coll, Fac Biomed Engn, Jilin 132013, Jilin, Peoples R China
基金
中国国家自然科学基金;
关键词
Electrical impedance; brain imaging; data acquisition; imaging monitoring; EIT SYSTEM; CALIBRATION; ELECTRODES; DESIGN; EEG;
D O I
10.1109/JSEN.2018.2836336
中图分类号
TM [电工技术]; TN [电子技术、通信技术];
学科分类号
0808 ; 0809 ;
摘要
In electrical impedance tomography (EIT), it is difficult to obtain the intracranial impedance due to the highly resistive skull enclosing the brain. Therefore, a high-precision data acquisition system is required for brain EIT. In this paper, we used a high-precision digital synthesis method and a digital demodulation technique with high noise immunity to eliminate random errors. Moreover, we focused on two problems encountered during EIT data acquisition: 1) the shunt effect on the excitation current due to the distributed capacitance between electrodes and ground and 2) high common-mode voltages in the boundary measurements. We designed a new electrode interface to reduce the influence of the distributed capacitance and a programmable current source to accurately compensate for the excited current. We also proposed a new voltmeter circuit with improved CMRR. Overall, this EIT data acquisition system can produce a programmable current with SNR greater than 89 dB. It can also measure the voltage difference precisely with CMRR higher than 75 dB with a 1-k Omega impedance imbalance. The results on a calibration model show that this system has a high SNR of 83 dB and a low reciprocity error of 0.125%. In addition, EIT imaging results were acquired using a brain physical phantom. The system can detect small disturbances of 0.35% in volume (1.99% of the cross-sectional area) and 17% in resistivity. Experiments on healthy volunteers also suggest that small intracranial impedance variations due to temporary occlusion and reperfusion of the unilateral carotid artery may be monitored by the system.
引用
收藏
页码:5974 / 5984
页数:11
相关论文
共 50 条
  • [41] Electrical impedance tomography imaging using a priori ultrasound data
    Soleimani, Manuchehr
    BIOMEDICAL ENGINEERING ONLINE, 2006, 5 (1)
  • [42] A unique data acquisition system for electrical resistance tomography
    Daily, W
    Ramirez, A
    Zonge, K
    PROCEEDINGS OF THE SYMPOSIUM ON THE APPLICATION OF GEOPHYSICS TO ENGINEERING AND ENVIRONMENTAL PROBLEMS, 1996, : 743 - 751
  • [43] Design and Construction of a High-Precision Electrical Parameters Monitoring System
    Rahman, Md Maksudur
    Morsalin, Sayidul
    Ahammad, Arif
    Mahmud, K.
    Reza, Syed Enam
    2016 5TH INTERNATIONAL CONFERENCE ON INFORMATICS, ELECTRONICS AND VISION (ICIEV), 2016, : 339 - 344
  • [44] High-Precision Probabilistic Imaging for Interface Debonding Monitoring Based on Electromechanical Impedance
    Du, Fei
    Wang, Guanghao
    Weng, Jiexin
    Fan, Haodong
    Xu, Chao
    AIAA JOURNAL, 2022, 60 (07) : 3950 - 3960
  • [45] ACT3 - A HIGH-SPEED, HIGH-PRECISION ELECTRICAL-IMPEDANCE TOMOGRAPH
    COOK, RD
    SAULNIER, GJ
    GISSER, DG
    GOBLE, JC
    NEWELL, JC
    ISAACSON, D
    IEEE TRANSACTIONS ON BIOMEDICAL ENGINEERING, 1994, 41 (08) : 713 - 722
  • [46] The acquisition hardware system with direct digital synthesis and filtered back-projection Imaging in electrical impedance tomography
    Xu, Guizhi
    Zhang, Shuai
    Wu, Huanli
    Yang, Shuo
    Geng, Duyan
    Yan, Weili
    Wang, Mingshi
    2005 27TH ANNUAL INTERNATIONAL CONFERENCE OF THE IEEE ENGINEERING IN MEDICINE AND BIOLOGY SOCIETY, VOLS 1-7, 2005, : 7758 - 7761
  • [47] Test and measurement requires high-accuracy and high-precision data acquisition solutions
    Pachchigar, Maithil
    Electronics World, 2023, 128 (2023): : 26 - 31
  • [48] A electrical impedance tomography system
    Peng, Y
    Mo, YL
    IEEE EMBS APBME 2003, 2003, : 280 - 281
  • [49] The study on image data acquisition system for electrical capacitance tomography system
    Li, Yan
    Jin, Tianbo
    Meng, Xiaoying
    Chen, Deyun
    Zhang, Liyong
    ICIEA 2007: 2ND IEEE CONFERENCE ON INDUSTRIAL ELECTRONICS AND APPLICATIONS, VOLS 1-4, PROCEEDINGS, 2007, : 1635 - 1638
  • [50] Electrical Impedance Tomography of brain function
    Holder, David S.