A quality assurance tool for high-dose-rate brachytherapy

被引:26
|
作者
Rickey, Daniel W. [1 ,2 ,3 ]
Sasaki, David [1 ]
Bews, Jeff [1 ,2 ,3 ]
机构
[1] CancerCare Manitoba, Med Phys, Winnipeg, MB R3E 0V9, Canada
[2] Univ Manitoba, Dept Phys & Astron, Winnipeg, MB R3T 2N2, Canada
[3] Univ Manitoba, Dept Radiol, Winnipeg, MB R3A 1R9, Canada
关键词
high-dose-rate brachytherapy; quality assurance; radiochromic film; diode detector; RADIOCHROMIC FILM; HDR BRACHYTHERAPY; STEPPING SOURCE; CANCER; TIME;
D O I
10.1118/1.3425786
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Purpose: The purpose of this work was to develop a quality assurance (QA) tool for high-dose-rate (HDR) brachytherapy that would quickly and easily verify both source positioning (dwell positions) and durations (dwell times). Methods: The authors constructed a QA tool that combined radiochromic film to verify position with four photodiode detectors to verify dwell times. To characterize the temporal accuracy of the tool, a function generator powered four red light-emitting diodes that were optically coupled to the four photodiode detectors. The QA tool was used to verify the dwell positions and times of a commercial brachytherapy afterloader. Measurements of dwell time were independently verified by a one-dimensional optical camera that acquired 1000 lines/s. Results: The temporal accuracy of the QA tool was found to be about 1 ms. For visual assessment, the source position could be located within about 0.5 mm. Evaluating the accuracy and precision of an HDR brachytherapy afterloader, the authors found that the bias in dwell time can exceed 60 ms and the dwell time associated with the first dwell position had an unexpectedly large standard deviation of 30 ms. They found that the source locations were much easier to locate on the film if a plastic catheter was used instead of a metal treatment tube. Scanning the films enabled the dwell positions to be determined within about 0.2 mm. Conclusions: For pretreatment QA, the authors found that this tool allowed verification of dwell positions and dwell times in about 6 min. (C) 2010 American Association of Physicists in Medicine. [DOI: 10.1118/1.3425786]
引用
收藏
页码:2525 / 2532
页数:8
相关论文
共 50 条
  • [21] High-Dose-Rate Prostate Brachytherapy Consistently Results in High Quality Dosimetry
    White, Evan C.
    Kamrava, Mitchell R.
    Demarco, John
    Park, Sang-June
    Wang, Pin-Chieh
    Kayode, Oluwatosin
    Steinberg, Michael L.
    Demanes, D. Jeffrey
    INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2013, 85 (02): : 543 - 548
  • [22] Modern development of high-dose-rate brachytherapy
    Itami, Jun
    JAPANESE JOURNAL OF CLINICAL ONCOLOGY, 2020, 50 (05) : 490 - 501
  • [23] VERIFICATION OF HIGH-DOSE-RATE BRACHYTHERAPY PLANS
    MOWER, HW
    LO, TC
    GIRSHOVICH, L
    LAWSON, BJ
    COSTEY, GE
    RADIOLOGY, 1992, 185 : 129 - 130
  • [24] "Magic phantom" - a quality assurance system for high dose rate brachytherapy
    Espinoza, A.
    Petasecca, M.
    Fuduli, I.
    Howie, A.
    Corde, S.
    Bucci, J.
    Jackson, M.
    Lerch, M. L. F.
    Rosenfeld, A. B.
    RADIOTHERAPY AND ONCOLOGY, 2015, 115 : S139 - S140
  • [25] Feasibility Study of Patient-specific Quality Assurance System for High-dose-rate Brachytherapy in Patients with Cervical Cancer
    Lee, Boram
    Ahn, Sung Hwan
    Kim, Hyeyoung
    Han, Youngyih
    Huh, Seung Jae
    Kim, Jin Sung
    Kim, Dong Wook
    Sim, Jina
    Yoon, Myonggeun
    JOURNAL OF THE KOREAN PHYSICAL SOCIETY, 2016, 68 (08) : L1029 - L1036
  • [26] A Quality Assurance System with a Web Camera for High Dose Rate Brachytherapy
    Ueda, Y.
    Hirose, A.
    Oohira, S.
    Isono, M.
    Tsujii, K.
    Miyazaki, M.
    Kawaguchi, Y.
    Konishi, K.
    Teshima, T.
    MEDICAL PHYSICS, 2015, 42 (06) : 3346 - 3346
  • [27] Feasibility study of patient-specific quality assurance system for high-dose-rate brachytherapy in patients with cervical cancer
    Boram Lee
    Sung Hwan Ahn
    Hyeyoung Kim
    Youngyih Han
    Seung Jae Huh
    Jin Sung Kim
    Dong Wook Kim
    Jina Sim
    Myonggeun Yoon
    Journal of the Korean Physical Society, 2016, 68 : 1029 - 1036
  • [28] Quality of life after high-dose-rate brachytherapy monotherapy for prostate cancer
    Contreras, Jessika A.
    Wilder, Richard B.
    Mellon, Eric A.
    Strom, Tobin J.
    Fernandez, Daniel C.
    Biagioli, Matthew C.
    INTERNATIONAL BRAZ J UROL, 2015, 41 (01): : 40 - 45
  • [29] Low-dose-rate brachytherapy is superior to high-dose-rate brachytherapy for bladder cancer
    Pos, FJ
    Horenblas, S
    Lebesque, J
    Moonen, L
    Schneider, C
    Sminia, P
    Bartelink, H
    INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2004, 59 (03): : 696 - 705
  • [30] The Effect of Needle Number on the Quality of High-dose-rate Prostate Brachytherapy Implants
    Froehlich, Georgina
    Agoston, Peter
    Loevey, Jozsef
    Polgar, Csaba
    Major, Tibor
    PATHOLOGY & ONCOLOGY RESEARCH, 2010, 16 (04) : 593 - 599