Phantoms for noninvasive blood glucose sensing with near infrared transmission spectroscopy

被引:0
|
作者
Burmeister, JJ [1 ]
Chung, HI [1 ]
Arnold, MA [1 ]
机构
[1] Univ Iowa, Dept Chem, Iowa City, IA 52242 USA
关键词
D O I
10.1562/0031-8655(1998)067<0050:PFNBGS>2.3.CO;2
中图分类号
Q5 [生物化学]; Q7 [分子生物学];
学科分类号
071010 ; 081704 ;
摘要
In vivo spectra from human subjects can be simulated with a phantom composed of different layers of water, fat and muscle tissue, All three components are necessary to simulate in vivo spectra collected over the combination spectral region (5000-4000 cm(-1)). Muscle tissue is not required, however, to accurately simulate overtone spectra (6600-5400 cm(-1)), The near-IR spectral characteristics of fat and muscle tissue from several animal sources are essentially identical to those found for human tissue, hence, the animal source for these phantom components is not critical, Thickness of each tissue layer can be determined by a regression analysis where the in vivo spectrum of interest is regressed against standard absorbance spectra of the necessary model components (water, fat and muscle), In general, in vivo overtone spectra collected across human webbing tissue with a thickness of 6.7 mm can be simulated with water layer thicknesses ranging from 5.0 to 6.4 mm combined with fat layer thicknesses from 1.4 to 4.2 mm.
引用
收藏
页码:50 / 55
页数:6
相关论文
共 50 条
  • [1] Evaluation of measurement sites for noninvasive blood glucose sensing with near-infrared transmission spectroscopy
    Burmeister, JJ
    Arnold, MA
    [J]. CLINICAL CHEMISTRY, 1999, 45 (09) : 1621 - 1627
  • [2] Noninvasive blood glucose sensing with near-infrared spectroscopy based on interstitial fluid
    [J]. Lu, Q. (luqipeng@126.com), 1600, Chinese Optical Society (33):
  • [3] Research on the Effective Signal Extraction in the Noninvasive Blood Glucose Sensing by Near Infrared Spectroscopy
    Ding Hai-quan
    Lu Qi-peng
    Wang Dong-min
    Chen Xing-dan
    [J]. SPECTROSCOPY AND SPECTRAL ANALYSIS, 2010, 30 (01) : 50 - 53
  • [4] Accuracy of Noninvasive Glucose Sensing Based on Near-Infrared Spectroscopy
    Liu, Jin
    Liu, Rong
    Xu, Kexin
    [J]. APPLIED SPECTROSCOPY, 2015, 69 (11) : 1313 - 1318
  • [5] Progress in noninvasive glucose monitoring with near infrared transmission spectroscopy.
    Arnold, MA
    [J]. ABSTRACTS OF PAPERS OF THE AMERICAN CHEMICAL SOCIETY, 2002, 224 : U114 - U114
  • [6] On Feasibility of Near-Infrared Spectroscopy for Noninvasive Blood Glucose Measurements
    Vural, N. Mert
    Yoleri, Yigit
    Torun, Hamdi
    [J]. OPTICAL DIAGNOSTICS AND SENSING XIX: TOWARD POINT-OF-CARE DIAGNOSTICS, 2019, 10885
  • [7] New advancement in noninvasive measurement of blood glucose by near-infrared spectroscopy
    Li, QB
    Wang, Y
    Xu, KX
    [J]. OPTICS IN HEALTH CARE AND BIOMEDICAL OPTICS: DIAGNOSTICS AND TREATMENT, 2002, 4916 : 457 - 464
  • [8] NONINVASIVE BLOOD-GLUCOSE SENSORS BASED ON NEAR-INFRARED SPECTROSCOPY
    HEISE, HM
    MARBACH, R
    KOSCHINSKY, T
    GRIES, FA
    [J]. ARTIFICIAL ORGANS, 1994, 18 (06) : 439 - 447
  • [9] Noninvasive Blood Glucose Measurements by Means of Near Infrared Spectroscopy: State of the Art
    Paez Roa, Arbey Alexis
    Villamizar Mejia, Rodolfo
    [J]. UIS INGENIERIAS, 2012, 11 (01): : 21 - 33
  • [10] Noninvasive Blood Glucose Analysis Based on Near-Infrared Reflectance Spectroscopy
    Lu Xiao-feng
    Zhang Ting-lin
    Xiao Feng
    Li Guang
    Wang You
    [J]. SPECTROSCOPY AND SPECTRAL ANALYSIS, 2016, 36 (07) : 2312 - 2317