Finite element analysis comparison between superior clavicle locking plate with and without screw holes above fracture zone in midshaft clavicular fracture

被引:13
|
作者
Pengrung, Nachapan [1 ]
Lakdee, Natthaphop [2 ]
Puncreobutr, Chedtha [3 ,4 ]
Lohwongwatana, Boonrat [3 ,4 ]
Sa-ngasoongsong, Paphon [1 ]
机构
[1] Mahidol Univ, Fac Med, Dept Orthoped, Ramathibodi Hosp, 270,Rama 6 Rd, Bangkok 10400, Thailand
[2] Chulalongkorn Univ, Meticuly Co Ltd, Biomech Res Ctr, Bangkok, Thailand
[3] Chulalongkorn Univ, Biomed Engn Res Ctr, Bangkok, Thailand
[4] Chulalongkorn Univ, Fac Engn, Dept Met Engn, Bangkok, Thailand
关键词
Superior clavicle locking plate; Midshaft clavicular fracture; Plate fixation; Screw hole; Implant failure; Peak stress; NONOPERATIVE TREATMENT; FIXATION; METAANALYSIS; STRENGTH;
D O I
10.1186/s12891-019-2847-y
中图分类号
R826.8 [整形外科学]; R782.2 [口腔颌面部整形外科学]; R726.2 [小儿整形外科学]; R62 [整形外科学(修复外科学)];
学科分类号
摘要
Background Midshaft clavicular fractures are common fractures and generally treated conservatively. Among the surgical options, plate fixation is the most popular and has been biomechanically and clinically proven in numerous studies. However, implant failures caused by plate deformations or breakage still occur in up to 16.7% of cases, and recent studies showed that screw holes above fracture zone (SHFZ) might be the at-risk location. Using finite element analysis, this study aimed to test the biomechanical property of the superior clavicle locking plate (SCLP) with and without SHFZ in comminuted midshaft clavicular fracture. Methods Finite element models of comminuted midshaft clavicular fracture fixed with standard 8-hole titanium SCLP with screw holes (SHFZ plate) and without screw holes above fracture zone (No-SHFZ plate) were built. Both groups were tested under three different loading models (100-N cantilever bending, 100-N axial compression, and 1-Nm torsion). The average peak stress on medial clavicle, fracture zone, and lateral clavicle, and the peak stress on each screw hole (or the same position in the No-SHFZ plate) were measured and compared. Results The highest average peak stress on the fracture zone was higher than those on medial and lateral clavicles under all loading conditions in both plates. However, the No-SHFZ plate significantly reduced the average peak stress value on the fracture zone, compared to the SHFZ plate (45.0% reduction in cantilever bending, 52.2% reduction in axial compression, and 54.9% reduction in axial torsion). The peak stress value on the maximal stress point in the SHFZ and No-SHFZ plates with cantilever bending, axial compression, and torsion loads were 1257.10 MPa vs. 647.21 MPa, 186.42 MPa vs. 131.63 MPa, and 111.86 MPa vs. 82.41 MPa, respectively. Conclusion The weakest link of the SCLP construct in comminuted midshaft clavicular fracture fixation is the SHFZ, especially in the cantilever bending load. Additionally, the biomechanical property of the SCLP without SHFZ model (No-SHFZ plate) is superior to the standard SCLP model (SHFZ plate), with a significantly lower peak stress on the SHFZ location in all loading conditions. We recommend a new SCLP design with SHFZ to prevent implant failure and improve surgical outcomes.
引用
收藏
页数:7
相关论文
共 50 条
  • [1] Finite element analysis comparison between superior clavicle locking plate with and without screw holes above fracture zone in midshaft clavicular fracture
    Nachapan Pengrung
    Natthaphop Lakdee
    Chedtha Puncreobutr
    Boonrat Lohwongwatana
    Paphon Sa-ngasoongsong
    BMC Musculoskeletal Disorders, 20
  • [2] Conceptual finite element study for comparison among superior, anterior, and spiral clavicle plate fixations for midshaft clavicle fracture
    Huang, Teng-Le
    Chen, Wen-Chuan
    Lin, Kun-Jhih
    Tsai, Cheng-Lun
    Lin, Kang-Ping
    Wei, Hung-Wen
    MEDICAL ENGINEERING & PHYSICS, 2016, 38 (10) : 1070 - 1075
  • [3] Screw Stress Distribution in a Clavicle Fracture with Plate Fixation: A Finite Element Analysis
    Alito, Angelo
    Fenga, Domenico
    Tropeano, Giada
    Milardi, Demetrio
    Leonetti, Danilo
    Migliorato, Alba
    Tisano, Adriana
    D'Andrea, Danilo
    Filardi, Vincenzo
    BIOENGINEERING-BASEL, 2023, 10 (12):
  • [4] Biomechanical comparison of the Locking Compression superior anterior clavicle plate with seven and ten hole reconstruction plates in midshaft clavicle fracture stabilisation
    Eden, Lars
    Doht, Stefanie
    Frey, Soenke P.
    Ziegler, Dirk
    Stoyhe, Jan
    Fehske, Kai
    Blunk, Torsten
    Meffert, Rainer H.
    INTERNATIONAL ORTHOPAEDICS, 2012, 36 (12) : 2537 - 2543
  • [5] Biomechanical comparison of the Locking Compression superior anterior clavicle plate with seven and ten hole reconstruction plates in midshaft clavicle fracture stabilisation
    Lars Eden
    Stefanie Doht
    Sönke P. Frey
    Dirk Ziegler
    Jan Stoyhe
    Kai Fehske
    Torsten Blunk
    Rainer H. Meffert
    International Orthopaedics, 2012, 36 : 2537 - 2543
  • [6] Finite element analysis of spiral plate and Herbert screw fixation for treatment of midshaft clavicle fractures
    Zhang, Xiaojuan
    Cheng, Xiaodong
    Yin, Bing
    Wang, Jianzhao
    Li, Sheng
    Liu, Guobin
    Hu, Zusheng
    Wu, Weiwei
    Zhang, Yingze
    MEDICINE, 2019, 98 (34)
  • [7] Management of midshaft clavicle fracture and ipsilateral acromioclavicular joint dislocation through combination of reconstruction locking plate and clavicular hook locking plate: an exposition of case series and literature review
    Zhang, Jian-Ping
    Zhang, Jin-Hui
    Liu, Hui
    Luo, De-Qing
    Xu, Wei-Zhen
    Wu, Jin
    INTERNATIONAL JOURNAL OF SURGERY OPEN, 2024, 62 (04) : 384 - 389
  • [8] Comparison of plate-screw systems used in mandibular fracture reduction: Finite element analysis
    Lovald, Scott T.
    Khraishi, Tariq
    Wagner, Jon
    Baack, Bret
    Kelly, James
    Wood, John
    JOURNAL OF BIOMECHANICAL ENGINEERING-TRANSACTIONS OF THE ASME, 2006, 128 (05): : 654 - 662
  • [9] Finite Element Analysis of a Novel Anterior Locking Plate for Thoracolumbar Burst Fracture
    Ren, Pengcheng
    Cheng, Xiaodong
    Lu, Chongyao
    Wu, Haotian
    Yao, Shuangquan
    Yang, Sidong
    Song, Zhaohui
    BIOMED RESEARCH INTERNATIONAL, 2021, 2021
  • [10] Finite element analysis of a novel anatomical locking plate for scapular neck fracture
    Xiang, Feifan
    Xiao, Yukun
    Wei, Daiqing
    Tan, Xiaoqi
    He, Siyuan
    Luo, Liang
    Yang, Yunkang
    JOURNAL OF ORTHOPAEDIC SURGERY AND RESEARCH, 2023, 18 (01)