Reduction in Blood Culture Contamination Through Use of Initial Specimen Diversion Device

被引:62
|
作者
Rupp, Mark E. [1 ]
Cavalieri, R. Jennifer [1 ]
Marolf, Cole [1 ]
Lyden, Elizabeth [2 ]
机构
[1] Univ Nebraska Med Ctr, Div Infect Dis, Omaha, NE USA
[2] Univ Nebraska Med Ctr, Dept Epidemiol, Omaha, NE USA
关键词
blood culture; contamination; initial specimen diversion device; EMERGENCY-DEPARTMENT; RATES; IMPACT;
D O I
10.1093/cid/cix304
中图分类号
R392 [医学免疫学]; Q939.91 [免疫学];
学科分类号
100102 ;
摘要
Background. Blood culture contamination is a clinically significant problem that results in patient harm and excess cost. Methods. In a prospective, controlled trial at an academic center Emergency Department, a device that diverts and sequesters the initial 1.5-2 mL portion of blood (which presumably carries contaminating skin cells and microbes) was tested against standard phlebotomy procedures in patients requiring blood cultures due to clinical suspicion of serious infection. Results. In sum, 971 subjects granted informed consent and were enrolled resulting in 904 nonduplicative subjects with 1808 blood cultures. Blood culture contamination was significantly reduced through use of the initial specimen diversion device T (ISDD) compared to standard procedure: (2/904 [0.22%] ISDD vs 16/904 [1.78%] standard practice, P = .001). Sensitivity was not compromised: true bacteremia was noted in 65/904 (7.2%) ISDD vs 69/904 (7.6%) standard procedure, P = .41. No needlestick injuries or potential bloodborne pathogen exposures were reported. The monthly rate of blood culture contamination for all nurse-drawn and phlebotomist-drawn blood cultures was modeled using Poisson regression to compare the 12-month intervention period to the 6 month before and after periods. Phlebotomists (used the ISDD) experienced a significant decrease in blood culture contamination while the nurses (did not use the ISDD) did not. In sum, 73% of phlebotomists completed a post-study anonymous survey and widespread user satisfaction was noted. Conclusions. Use of the ISDD was associated with a significant decrease in blood culture contamination in patients undergoing blood cultures in an Emergency Department setting.
引用
收藏
页码:201 / 205
页数:5
相关论文
共 50 条
  • [1] Reducing blood culture contamination using an initial specimen diversion device
    Zimmerman, Frederic S.
    Assous, Marc, V
    Zevin, Shoshana
    Wiener-Well, Yonit
    AMERICAN JOURNAL OF INFECTION CONTROL, 2019, 47 (07) : 822 - 826
  • [2] Initial Specimen Diversion Device® reduces blood culture contamination and vancomycin use in academic medical centre
    Nielsen, L. E.
    Nguyen, K.
    Wahl, C. K.
    Huss, J. L.
    Chang, D.
    Ager, E. P.
    Hamilton, L.
    JOURNAL OF HOSPITAL INFECTION, 2022, 120 : 127 - 133
  • [3] The Impact of Initial Specimen Diversion Systems on Blood Culture Contamination
    Al Mohajer, Mayar
    Lasco, Todd
    OPEN FORUM INFECTIOUS DISEASES, 2023, 10 (05):
  • [4] Innovation for Reducing Blood Culture Contamination: Initial Specimen Diversion Technique
    Patton, Richard G.
    Schmitt, Timothy
    JOURNAL OF CLINICAL MICROBIOLOGY, 2010, 48 (12) : 4501 - 4503
  • [5] Effect of the Initial Specimen Diversion Technique on Blood Culture Contamination Rates
    Binkhamis, Khalifa
    Forward, Kevin
    JOURNAL OF CLINICAL MICROBIOLOGY, 2014, 52 (03) : 980 - 981
  • [6] Effect of an initial specimen diversion device on blood-culture contamination rates and vancomycin usage: A quasi-experimental study
    Wilber, Eli P. P.
    Babiker, Ahmed
    Howard-Anderson, Jessica
    Holdsworth, Jill E. E.
    Burd, Eileen M. M.
    Eldridge, M. Jeremy
    Jacob, Jesse T. T.
    INFECTION CONTROL AND HOSPITAL EPIDEMIOLOGY, 2024, 45 (01): : 100 - 102
  • [7] Initial Specimen Diversion Device Utilization Mitigates Blood Culture Contamination Across Regional Community Hospital and Acute Care Facility
    Povroznik, Mark D.
    AMERICAN JOURNAL OF MEDICAL QUALITY, 2022, 37 (05) : 405 - 412
  • [8] Decreasing Blood Culture Contamination Rates Using A Specimen Diversion Device: A Quasi-Experimental Study
    Babiker, Ahmed
    Ramakrishnan, Aditi
    Howard-Anderson, Jessica
    Holdsworh, Jill
    Jacob, Mini
    Jacob, Jesse
    INFECTION CONTROL AND HOSPITAL EPIDEMIOLOGY, 2020, 41 : S468 - S468
  • [9] Decreasing blood culture contamination rates using a specimen diversion device: a quasi-experimental study
    Wilber, E.
    Babiker, A.
    Howard-Anderson, J.
    Eldridge, J.
    Burd, E.
    Holdsworth, J.
    Jacob, J.
    AMERICAN JOURNAL OF THE MEDICAL SCIENCES, 2023, 365 : S373 - S373
  • [10] Evaluation of the initial specimen diversion technique (ISDT) for reducing blood culture contamination in children: a prospective randomized controlled trial
    Lalezari, A.
    Rayya, Z. A.
    Lellouche, J.
    Yana, M.
    Sharon, N.
    JOURNAL OF HOSPITAL INFECTION, 2022, 128 : 96 - 97