X-ray performance of a 1242 cm3 LaCl3:Ce scintillation spectrometer

被引:24
|
作者
Owens, Alan [1 ]
Bos, A. J. J.
Brandenburg, S.
Dathy, C.
Dorenbos, P.
Kraft, S.
Ostendorf, R. W.
Ouspenski, V.
Quarati, F.
机构
[1] Estec, ESA, Sci Payload & Adv Concepts Off, Noordwijk, Netherlands
[2] Delft Univ Technol, Delft, Netherlands
[3] Univ Groningen, Kernfys Versneller Inst, NL-9747 AA Groningen, Netherlands
[4] St Gobain Cristaux, F-77794 Nemours, France
[5] Cosine Res BV, NL-2333 CA Leiden, Netherlands
关键词
lanthanum chloride; scintillator detectors; gamma-ray spectroscopy;
D O I
10.1016/j.nima.2007.01.091
中图分类号
TH7 [仪器、仪表];
学科分类号
0804 ; 080401 ; 081102 ;
摘要
We present the results of gamma-ray measurements on a large 4 '' x 6 '' LaCl3:Ce crystal, characterized using radioactive sources over the range 14-3220 keV. The response of the crystal was found to be largely linear over the upper 90% of its dynamic range-the deviations at lower energies can be attributed to the shortcomings of the photomultiplier tube (PMT) and is in fact a generic problem when using high-gain PMTs with lanthanum halide scintillators. At 662 keV, a measured energy resolution of 4.1% FWHM was recorded at room temperature. In comparison to earlier generations of LaCl3 detectors, the internal background spectrum shows significantly less X-ray, beta, gamma-ray and alpha contamination. The integral background count rate in the energy region 20 keV-3 MeV was determined to be 1.8 cm(-3) s(-1) of which most (1.5 cm(-3) s(-1)) can be attributed to the beta continuum from La-138 and only similar to 0.04 cm(-3) s(-1) to the alpha complex from Ac-227 and daughters. Using these data, we have determined the light output of alpha particles relative to the equivalent energy electrons (or alternately gamma-rays) to be, alpha/gamma = 0.35 +/- 0.02. (c) 2007 Elsevier B.V. All rights reserved.
引用
收藏
页码:110 / 114
页数:5
相关论文
共 50 条
  • [21] LaCl3(Ce) scintillation detector applications for environmental gamma-ray measurements of low to high dose rates
    Tsutsumi, M
    Tanimura, Y
    NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 2006, 557 (02): : 554 - 560
  • [22] LONG-DELAYED FLUORESCENCE OF ND3+ IN PURE LACL3 AND IN LACL3 CONTAINING CE3+
    ASAWA, CK
    PHYSICAL REVIEW, 1967, 155 (02): : 188 - &
  • [23] INFRARED ZEEMAN EFFECT FOR CE3+ IN LACL3
    BAGGULEY, DM
    VELLACOL.G
    JOURNAL OF APPLIED PHYSICS, 1968, 39 (2P2) : 973 - &
  • [24] LaCl3:Ce coincidence signatures to calibrate gamma-ray detectors
    McIntyre, Justin I.
    Schrom, Brian T.
    Cooper, Mathew W.
    Day, Anthony
    Hayes, James C.
    Heimbigner, Tom R.
    Hubbard, Charles W.
    Ripplinger, Michael D.
    Suarez, Reynold
    NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 2011, 652 (01): : 201 - 204
  • [25] Proton induced activation of LaBr3:Ce and LaCl3:Ce
    Buis, Ernst-Jan
    Quarati, F.
    Brandenburg, S.
    Bos, A. J. J.
    Dathy, C.
    Dorenbos, P.
    Kraft, S.
    Maddox, E.
    Ostendorf, R.
    Owens, A.
    NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 2007, 580 (02): : 902 - 905
  • [26] Radiation damage effects of LaCl3:Ce crystal
    Pei, Yu
    Chen, Xiaofeng
    Yao, Dongmin
    Ren, Guohao
    RADIATION MEASUREMENTS, 2007, 42 (03) : 407 - 412
  • [27] Test of a 3" x 3" LaCl3:Ce crystal and neutron sensitivity of lanthanum halide scintillators
    Martinez, T.
    Cano-Ott, D.
    Guerrero, C.
    2008 IEEE NUCLEAR SCIENCE SYMPOSIUM AND MEDICAL IMAGING CONFERENCE (2008 NSS/MIC), VOLS 1-9, 2009, : 495 - 497
  • [28] Characterization Of LaCl3:Ce Detectors For Prompt Fission Gamma-Ray Measurements
    Oberstedt, A.
    Billnert, R.
    Karlsson, J.
    Oberstedt, S.
    Geerts, W.
    Hambsch, F. -J.
    4TH INTERNATIONAL WORKSHOP ON NUCLEAR FISSION AND FISSION-PRODUCT SPECTROSCOPY, 2009, 1175 : 257 - +
  • [29] Magnetic resonance investigations of LaCl3:Ce3+ scintillators
    Rogulis, U
    Schweizer, S
    Spaeth, JM
    Van Loef, EVD
    Dorenbos, P
    Van Eijk, CWE
    Krämer, K
    Güdel, HU
    RADIATION EFFECTS AND DEFECTS IN SOLIDS, 2002, 157 (6-12): : 951 - 955
  • [30] Scintillation properties of LaCl3:Ce3+ crystals:: Fast, efficient, and high-energy resolution scintillators
    van Loef, EVD
    Dorenbos, P
    Krämer, K
    Güdel, HU
    IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 2001, 48 (03) : 341 - 345