Distal Locking Screws for Intramedullary Nailing of Tibial Fractures

被引:8
|
作者
Agathangelidis, Filon [1 ]
Petsatodis, Georgios [1 ]
Kirkos, John [2 ]
Papaapadopoulos, Pericles [1 ]
Karataglis, Dimitrios [3 ]
Christodoulou, Anastasios [1 ]
机构
[1] Aristotle Univ Thessaloniki, Dept Orthopaed 1, G Papanikolaou Gen Hosp, GR-54006 Thessaloniki, Greece
[2] Aristotle Univ Thessaloniki, Dept Orthopaed 3, Papageorgiou Gen Hosp, GR-54006 Thessaloniki, Greece
[3] Euromed Blue Cross Clin, Thessaloniki, Greece
关键词
INTERLOCKING; STABILITY; MODEL;
D O I
10.3928/01477447-20160129-02
中图分类号
R826.8 [整形外科学]; R782.2 [口腔颌面部整形外科学]; R726.2 [小儿整形外科学]; R62 [整形外科学(修复外科学)];
学科分类号
摘要
Recently introduced tibial intramedullary nails allow a number of distal screws to be used to reduce the incidence of malalignment and loss of fixation of distal metaphyseal fractures. However, the number of screws and the type of screw configuration to be used remains obscure. This biomechanical study was performed to address this question. Thirty-six Expert tibial nails (Synthes, Oberdorf, Switzerland) were introduced in composite bone models. The models were divided into 4 groups with different distal locking configurations ranging from 2 to 4 screws. A 7-mm gap osteotomy was performed 72 mm from the tibial plafond to simulate a 42-C3 unstable distal tibial fracture. Each group was divided in 3 subgroups and underwent nondestructive biomechanical testing in axial compression, coronal bending, and axial torsion. The passive construct stiffness was measured and statistically analyzed with one-way analysis of variance. Although some differences were noted between the stiffness of each group, these were not statistically significant in compression (P=.105), bending (P=.801), external rotation (P=.246), and internal rotation (P=.370). This in vitro study showed that, when using the Expert tibial nail for unstable distal tibial fractures, the classic configuration of 2 parallel distal screws could provide the necessary stability under partial weight-bearing conditions.
引用
收藏
页码:E253 / E258
页数:6
相关论文
共 50 条
  • [41] INTRAMEDULLARY NAILING OF TIBIAL SHAFT FRACTURES
    KUNER, EH
    ORTHOPADE, 1984, 13 (04): : 266 - 270
  • [42] BLIND INTRAMEDULLARY NAILING FOR TIBIAL FRACTURES
    DAUBIGNE, RM
    MAURER, P
    ZUCMAN, J
    MASSE, Y
    CLINICAL ORTHOPAEDICS AND RELATED RESEARCH, 1974, (105) : 267 - 275
  • [43] Intramedullary nailing of open tibial fractures
    Court-Brown, CM
    CURRENT ORTHOPAEDICS, 2003, 17 (03): : 161 - 166
  • [44] SECONDARY INTRAMEDULLARY NAILING OF TIBIAL FRACTURES
    OLERUD, S
    KARLSTRO.G
    JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME, 1972, A 54 (07): : 1419 - &
  • [45] INTRAMEDULLARY NAILING OF TIBIAL SHAFT FRACTURES
    SOLHEIM, K
    BO, O
    ACTA ORTHOPAEDICA SCANDINAVICA, 1973, 44 (03): : 323 - 334
  • [46] Intramedullary nailing of humeral diaphyseal fractures. Is distal locking really necessary
    Tyllianakis, Minos
    Tsoumpos, Pantelis
    Anagnostou, Kostas
    Konstantopoulou, Anna
    Panagopoulos, Andreas
    INTERNATIONAL JOURNAL OF SHOULDER SURGERY, 2013, 7 (02): : 65 - 69
  • [47] Is Distal Locking Necessary? A Biomechanical Investigation of Intramedullary Nailing Constructs for Intertrochanteric Fractures
    Gallagher, David
    Adams, Brock
    El-Gendi, Hebah
    Patel, Ashish
    Grossman, Lauren
    Berdia, Jay
    Korshunov, Yevgeniy
    Goldman, Ariel
    JOURNAL OF ORTHOPAEDIC TRAUMA, 2013, 27 (07) : 373 - 378
  • [48] Tibial unreamed intramedullary nailing using Schanz screws in displaced diaphyseal segmental fractures
    Kim, Kyung Cheon
    Lee, June Kyu
    Hwang, Deuk Soo
    Yang, Jun Young
    Kim, Young Mo
    ORTHOPEDICS, 2007, 30 (11) : 906 - 908
  • [49] Do Poller screws effect union in tibial shaft fractures treated with intramedullary nailing?
    Peat, Fidel
    Ordas-Bayon, Alejandro
    Krkovic, Matija
    INJURY-INTERNATIONAL JOURNAL OF THE CARE OF THE INJURED, 2021, 52 (10): : 3132 - 3138
  • [50] Treatment of distal tibial metaphyseal fractures: Plating versus shortened intramedullary nailing
    Yang, SW
    Tzeng, HM
    Chou, YJ
    Teng, HP
    Liu, HH
    Wong, CY
    INJURY-INTERNATIONAL JOURNAL OF THE CARE OF THE INJURED, 2006, 37 (06): : 531 - 535