Considerations on the miniaturization of detectors for in vivo dosimetry in radiotherapy: A Monte Carlo study

被引:9
|
作者
Cunha, Micaela [1 ,2 ]
Testa, Etienne [1 ,2 ]
Beuve, Michael [1 ,2 ]
Balosso, Jacques [3 ]
Chaikh, Abdulhamid [3 ]
机构
[1] Univ Lyon, F-69622 Lyon, France
[2] CNRS, IN2P3, Inst Phys Nucl Lyon, Paris, France
[3] Grenoble Alpes Univ Hosp CHUGA, Dept Radiat Oncol & Med Phys, Grenoble, France
关键词
In vivo dosimetry; Nanodosimeter; Microdosimeter; Monte Carlo simulation; Radiotherapy; LIQUID WATER; TRACK-STRUCTURE; SUBEXCITATION ELECTRONS; MULTIPLE IONIZATION; HYDRATED ELECTRON; ENERGY DEPOSITION; CROSS-SECTIONS; SOLID WATER; RADIOLYSIS; SIMULATION;
D O I
10.1016/j.nimb.2017.03.078
中图分类号
TH7 [仪器、仪表];
学科分类号
0804 ; 080401 ; 081102 ;
摘要
The evolution of technology in radiotherapy nowadays allows us to deliver much higher doses to the target volumes, thanks to better absorbed dose distribution accuracy and conformation, while better sparing healthy tissues. In photon radiotherapy, higher precision usually entails employing small and moving beams. This emphasizes the role of in vivo dosimetry, which assesses internal'absorbed dose rather than entrance dose. This, along with advances in materials science, results in a tendency towards the miniaturization of dosimeters. However, the stochastic nature of the radiation-matter interaction takes on greater importance at smaller scales, resulting in fluctuations in energy deposition whose effect may not be negligible. Miniaturization needs to take this into account. We estimated such fluctuations by Monte Carlo simulations considering energy deposition in cylindrical volumes of different sizes and for several absorbed dose values. We not only present and discuss the probability distributions of absorbed doses for a large range of target sizes (0.1-10 mu m) and clinically relevant doses (0.1-10 Gy), but also derive an estimation of the risk of measuring an absorbed dose with a value outside a given interval of tolerance (3, 5 and 10%) around the expected dose. The distributions presented features consistent with the theory of microdosimetry. Those for dosimeter sizes smaller than 0.3 p.m showed a very high dispersion in specific energy, while those for 10 gm dosimeters tended to become Gaussian and narrower with increasing absorbed dose. The probability of measuring an absorbed dose outside the defined interval of tolerance is close to 100% for the smallest size, regardless of the dose and the interval width considered. It decreases with increasing dose, dosimeter size and width of the interval of tolerance. The best results were obtained with 10 ism dosimeters, for which the probability of doses outside the tolerance range is always zero for absorbed doses equal or greater than 1 Gy and for all the absorbed, doses in the case of the largest interval of tolerance. The ability of a small dosimeter to give measurements within a suitable range of values depends on its size and on the absorbed dose to be measured. An excessively small dosimeter size yields a distribution of specific energies spread over a very large range of values, which makes it harder to obtain accurate measurements of absorbed dose. On the other hand, a dosimeter of about 10 tm in size may allow for good accuracy of the measurements in a wider range of doses and with different intervals of tolerance. (C) 2017 Elsevier B.V. All rights reserved.
引用
收藏
页码:20 / 27
页数:8
相关论文
共 50 条
  • [1] PET dosimetry in proton radiotherapy: a Monte Carlo study
    Del Guerra, A
    Di Domenico, G
    Mukhopadhayay, D
    [J]. APPLIED RADIATION AND ISOTOPES, 1997, 48 (10-12) : 1617 - 1624
  • [2] Monte Carlo simulations in radiotherapy dosimetry
    Pedro Andreo
    [J]. Radiation Oncology, 13
  • [3] Monte Carlo simulations in radiotherapy dosimetry
    Andreo, Pedro
    [J]. RADIATION ONCOLOGY, 2018, 13
  • [4] Monte Carlo application in radiation dosimetry and radiotherapy
    Ma, CM
    Jiang, SB
    Pawlicki, T
    Li, JS
    Deng, J
    Kapur, A
    Mora, G
    Lee, MC
    Mok, E
    Findley, D
    Boyer, AL
    [J]. PROCEEDINGS OF THE EIGHTH INTERNATIONAL CONFERENCE ON CALORIMETRY IN HIGH ENERGY PHYSICS, 2000, : 681 - 692
  • [5] Experimental characterization and Monte Carlo simulation of scintillator detectors in online electron FLASH radiotherapy dosimetry
    Morrocchi, Matteo
    Pensavalle, Jake Harold
    Ciarrocchi, Esther
    Di Martino, Fabio
    Felici, Giuseppe
    Galante, Federica
    Gasparini, Alessia
    Grasso, Luigi
    Linsalata, Stefania
    Massa, Maurizio
    Moggi, Andrea
    Pacitti, Matteo
    Vanreusel, Verdi
    Verellen, Dirk
    Bisogni, Maria Giuseppina
    [J]. JOURNAL OF INSTRUMENTATION, 2022, 17 (09)
  • [6] Monte Carlo approaches to important problems in radiotherapy dosimetry
    Nahum, AE
    [J]. COMPUTING RADIATION DOSIMETRY - CRD 2002, WORKSHOP PROCEEDINGS, 2004, : 99 - 114
  • [7] IN VIVO DOSIMETRY FROM EPID AND MONTE CARLO SIMULATION
    Badel, J. N.
    Ginestet, C.
    Sarrut, D.
    [J]. RADIOTHERAPY AND ONCOLOGY, 2008, 88 : S386 - S386
  • [8] Monte Carlo dosimetry for synchrotron stereotactic radiotherapy of brain tumours
    Boudou, C
    Balosso, J
    Estève, F
    Elleaume, H
    [J]. PHYSICS IN MEDICINE AND BIOLOGY, 2005, 50 (20): : 4841 - 4851
  • [9] Monte Carlo verification of gel dosimetry measurements for stereotactic radiotherapy
    Kairn, T.
    Taylor, M. L.
    Crowe, S. B.
    Dunn, L.
    Franich, R. D.
    Kenny, J.
    Knight, R. T.
    Trapp, J. V.
    [J]. PHYSICS IN MEDICINE AND BIOLOGY, 2012, 57 (11): : 3359 - 3369
  • [10] Dosimetry of lead shield in electron radiotherapy: A Monte Carlo evaluation
    Chow, J.
    Grigorov, G.
    [J]. MEDICAL PHYSICS, 2009, 36 (09) : 4308 - 4308