A cine-EPID based method for jaw detection and quality assurance for tracking jaw in IMRT/VMAT treatments

被引:15
|
作者
Fuangrod, Todsaporn [1 ]
Rowshanfarzad, Pejman [2 ]
Greer, Peter B. [3 ,4 ]
Middleton, Richard H. [1 ]
机构
[1] Univ Newcastle, Fac Engn & Built Environm, Sch Elect Engn & Comp Sci, Callaghan, NSW 2308, Australia
[2] Univ Western Australia, Sch Phys, Crawley, WA 6009, Australia
[3] Univ Newcastle, Fac Sci & IT, Sch Math & Phys Sci, Callaghan, NSW 2308, Australia
[4] Calvary Mater Newcastle Hosp, Dept Radiat Oncol, Newcastle, NSW 2310, Australia
来源
关键词
Real-time verification; Jaws; QA; EPID; RADIOCHROMIC FILM; DOSIMETRY;
D O I
10.1016/j.ejmp.2014.11.002
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
A new tool with the potential to verify and track jaw position during delivery has been developed. The method should be suitable for independent quality assurance for jaw position during jaw tracking dynamic IMRT and VMAT treatments. The jaw detection and tracking algorithm developed consists of five main steps. Firstly, the image is enhanced by removing a normalised predicted EPID image (that does not include the collimator transmission) from each cine EPID image. Then, using a histogram clustering technique a global intensity threshold level was determined. This threshold level was used to classify each pixel of the image as either under the jaws or under the MLC. Additionally, the collimator angle was automatically detected and used to rotate the image to vertical direction. Finally, this rotation allows the jaw positions to be determined using vertical and horizontal projection profiles. Nine IMRT fields (with static jaws) and a single VMAT clinical field (with dynamic jaws) were tested by determining the root mean square difference between planned and detected jaw positions. The test results give a detection accuracy of +/-1 mm RMS error for static jaw IMRT treatments and +/-1.5 mm RMS error for the dynamic jaw VMAT treatment. This method is designed for quality assurance and verification in modern radiation therapy; to detect the position of static jaws or verify the position of tracking jaws in more complex treatments. This method uses only information extracted from EPID images and it is therefore independent from the linear accelerator. (C) 2014 Associazione Italiana di Fisica Medica. Published by Elsevier Ltd. All rights reserved.
引用
收藏
页码:16 / 36
页数:21
相关论文
共 35 条
  • [31] Cloud-Based Monte Carlo Patient-Specific Quality Assurance (QA) Method for Volumetric-Modulated Arc Therapy (VMAT): Clinical and Educational Impacts
    Chen, X.
    Xing, L.
    Mok, E.
    Luxton, G.
    Bush, K.
    [J]. MEDICAL PHYSICS, 2013, 40 (06)
  • [32] Comparison of two different EPID-based solutions performing pretreatment quality assurance: 2D portal dosimetry versus 3D forward projection method
    Bresciani, Sara
    Poli, Matteo
    Miranti, Anna
    Maggio, Angelo
    Di Dia, Amalia
    Bracco, Christian
    Gabriele, Pietro
    Stasi, Michele
    [J]. PHYSICA MEDICA-EUROPEAN JOURNAL OF MEDICAL PHYSICS, 2018, 52 : 65 - 71
  • [33] Signal quality monitoring-based spoofing detection method for Global Navigation Satellite System vector tracking structure
    Zhang, Xinran
    Li, Hong
    Yang, Chun
    Lu, Mingquan
    [J]. IET RADAR SONAR AND NAVIGATION, 2020, 14 (06): : 944 - 953
  • [34] A Novel Quality Assurance Method to Find and Verify the Linear and Rotational Limits of the Cyberknife M6 Fiducial Based 6D Synchrony Respiratory Tracking
    Gibbard, G.
    Shang, C.
    [J]. MEDICAL PHYSICS, 2017, 44 (06)
  • [35] High power quality maximum power point tracking-based islanding detection method for grid-connected photovoltaic systems
    Bakhshi-Jafarabadi, Reza
    Sadeh, Javad
    Rakhshani, Elyas
    Popov, Marjan
    [J]. INTERNATIONAL JOURNAL OF ELECTRICAL POWER & ENERGY SYSTEMS, 2021, 131