A mathematical simulation of the tip-apex distance and the calcar-referenced tip-apex distance for intertrochanteric fractures reduced with lag screws

被引:23
|
作者
Li, Shuang [1 ]
Chang, Shi-Min [1 ]
Jin, Yan-Min [2 ]
Zhang, Ying-Qi [1 ]
Niu, Wen-Xin [3 ]
Du, Shou-Chao [1 ]
Zhang, Li-Zhi [1 ]
Ma, Hui [4 ]
机构
[1] Tongji Univ, Sch Med, Yangpu Hosp, Dept Orthoped Surg, 450 Tengyue Rd, Shanghai 200090, Peoples R China
[2] Tongji Univ, Coll Surveying & Geoinformat, Shanghai 200090, Peoples R China
[3] Tongji Univ, Sch Med, Yang Zhi Rehabil Hosp, Shanghai 200090, Peoples R China
[4] First Rehabil Hosp Shanghai, Dept Orthopaed Surg, Shanghai, Peoples R China
关键词
Tip-apex distance; TAD; Calcar-referenced tip-apex distance; CalTAD; Intertrochanteric fracture; PROXIMAL FEMORAL NAIL; TROCHANTERIC FRACTURES; GAMMA-NAIL; HIP SCREW; FIXATION; 3RD-GENERATION; ANTIROTATION; PARAMETERS; POSITION; FAILURE;
D O I
10.1016/j.injury.2016.03.029
中图分类号
R4 [临床医学];
学科分类号
1002 ; 100602 ;
摘要
Background: As a predictor of the risk of lag screw cutout, it was recommended that keeping tip-apex distance (TAD) < 25 mm and placing the screw centrally or inferiorly, but positioning the lag screw too inferiorly in the head would produce TAD > 25 mm. We aim to simulate various positions of the lag screw in the femoral head and identify whether 25 mm is a suitable cut-off value that favours all sizes of femoral heads with intertrochanteric fractures of the hip. Methods: Using a general mathematical software, the positions of the screw tip points were simulated. The virtual anterior-posterior and lateral views were then visualised, and the locus of the screw tips was projected into a Cartesian coordinate system according to the TAD and calcar-referenced tip-apex distance (CalTAD) formulas. Each original virtual anterior-posterior and lateral image was zoomed and compiled to match a calculated average image. The screw tip points were recorded, traced and compiled into volumes which could be used to visualise the screw's movements and positioning within the femoral head. The extracted volumes were calculated when 10 mm < TAD < 25 mm and 10 mm < CalTAD < 25 mm, and the region where these two volumes overlapped was also calculated. Suitable positions for the screw tip were then assessed. Results: For the TAD calculation, the shape of the traced screw tip points had a pancake-like appearance, while the CalTAD plot produced a teardrop-shaped region. The volume ratios of TAD, CalTAD and overlapping region relative to the femoral head volume were respectively 3.51 +/- 1.30%, 5.19 +/- 1.62% and 2.64 +/- 1.32%. The volumes of the traced TAD, CalTAD and overlapping regions increased slower than the volume of an idealised sphere. Conclusion: Positioning the lag screw should address geometrical effects of both tip-apex distance and femoral head size, with an emphasis on measuring the position of the screw tip for the suitable zone by volume ratio. The previous 25 mm TAD cut-off value should be adjusted according to the individual femoral head size. (C) 2016 Elsevier Ltd. All rights reserved.
引用
收藏
页码:1302 / 1308
页数:7
相关论文
共 50 条
  • [21] Impact of tip-apex distance and femoral head lag screw position on treatment outcomes of unstable intertrochanteric fractures using cephalomedullary nails
    Lee, Cheng-Hung
    Su, Kuo-Chih
    Chen, Kun-Hui
    Pan, Chien-Chou
    Wu, Yun-Che
    JOURNAL OF INTERNATIONAL MEDICAL RESEARCH, 2018, 46 (06) : 2128 - 2140
  • [22] Audit of Tip-Apex Distance in Dynamic Hip Screw Fixation of Proximal Femoral Fractures
    Potter, M.
    Hickey, B.
    BRITISH JOURNAL OF SURGERY, 2024, 111
  • [23] The tip-apex distance in the X-Bolt dynamic plating system
    Fernandez, M. A.
    Aquilina, A.
    Achten, J.
    Parsons, N.
    Costa, M. L.
    Griffin, X. L.
    BONE & JOINT RESEARCH, 2017, 6 (04): : 204 - 207
  • [24] Tip-apex distance as a risk factor for cut-out in cephalic double-screw nailing of intertrochanteric femur fractures
    Coviello, M.
    Abate, A.
    Maccagnano, G.
    Ippolito, F.
    Nappi, V.
    Abbaticchio, A. M.
    Caiaffa, E.
    Caiaffa, V.
    BONE & JOINT OPEN, 2024, 5 (06): : 457 - 463
  • [25] Clinical evaluation of PFNAA® and relationship between the tip-apex distance and mechanical failure
    Kraus, M.
    Krischak, G.
    Wiedmann, K.
    Riepl, C.
    Gebhard, F.
    Joeckel, J. A.
    Scola, A.
    UNFALLCHIRURG, 2011, 114 (06): : 470 - 478
  • [26] Measurement of Tip Apex Distance and Migration of Lag Screws and Novel Blade Screw Used for the Fixation of Intertrochanteric Fractures
    Yang, Jesse Chieh-Szu
    Chen, Hsin-Chang
    Lai, Yu-Shu
    Cheng, Cheng-Kung
    PLOS ONE, 2017, 12 (01):
  • [27] Accuracy of the surgeon's eye: Use of the tip-apex distance in clinical practice
    Wright, Jonathan
    Kahane, Steven
    Moeed, Abdul
    MacDowell, Andrew
    INJURY-INTERNATIONAL JOURNAL OF THE CARE OF THE INJURED, 2015, 46 (07): : 1346 - 1348
  • [28] Measuring tip-apex distance using a picture archiving and communication system (PACS)
    Johnson, Luke J.
    Cope, Marcus R.
    Shahrokhi, Shahram
    Tarnblyn, Peter
    INJURY-INTERNATIONAL JOURNAL OF THE CARE OF THE INJURED, 2008, 39 (07): : 786 - 790
  • [29] Intertrochanteric fracture fixation with Dynamic Hip Screw: Is tip-apex distance measurement useful for predicting fixation failure?
    Goh, K. L.
    Zamzuri, Z.
    Ariff, Mohd S.
    Azril, Mohamed M. A.
    INTERNATIONAL MEDICAL JOURNAL MALAYSIA, 2016, 15 (01) : 31 - 34
  • [30] Late postoperative analysis of the tip-apex distance (TAD) in pertrochanteric fractures: is there an accommodation of the implant within the bone?
    Aihara, Leandro Jun
    Nanni, Rafael Augusto
    Carvalho, Marina Sousa
    Zamboni, Caio
    Durigan, Jorge R.
    Hungria Neto, Jose S.
    Mercadante, Marcelo T.
    Christian, Ralph W.
    Hungria, Jose O. S.
    INJURY-INTERNATIONAL JOURNAL OF THE CARE OF THE INJURED, 2017, 48 : S54 - S56