Volume-of-interest, (VOI) imaging in C-arm flat-detector CT for high image quality at reduced dose

被引:57
|
作者
Kolditz, Daniel [1 ]
Kyriakou, Yiannis [1 ]
Kalender, Willi A. [1 ]
机构
[1] Univ Erlangen Nurnberg, Inst Med Phys, D-91052 Erlangen, Germany
关键词
volume-of-interest (VOI) imaging; flat-detector CT (FDCT); C-arm CT; dose; image quality; BEAM BREAST CT; GUIDED RADIATION-THERAPY; COMPUTED-TOMOGRAPHY; PANEL DETECTOR; RECONSTRUCTION; REGION; FEASIBILITY; RESOLUTION; ALGORITHM; GUIDANCE;
D O I
10.1118/1.3427641
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Purpose: A novel method for flat-detector computed tomography was developed to enable volume-of-interest (VOI) imaging at high resolution, low noise, and reduced dose. For this, a full low-dose overview (OV) scan and a local high-dose scan of a VOI are combined. Methods: The first scan yields an overview of the whole object and enables the selection of an arbitrary VOI. The second scan of that VOI assures high image quality within the interesting volume. The combination of the two consecutive scans is based on a forward projection of the reconstructed OV volume that was registered to the VOI. The artificial projection data of the OV scan are combined with the measured VOI data in the raw data domain. Different projection values are matched by an appropriate transformation and weighting. The reconstruction is performed with a standard Feldkamp-type algorithm. In simulations, the combination of OV scan and VOI scan was investigated on a mathematically described phantom. In measurements, spatial resolution and noise were evaluated with image quality phantoms. Modulation transfer functions and noise values were calculated. Measurements of an anthropomorphic head phantom were used to validate the proposed method for realistic applications, e. g., imaging stents. In Monte Carlo simulations, 3D dose distributions were calculated and dose values were assessed quantitatively. Results: By the proposed combination method, an image is generated which covers the whole object and provides the VOI at high image quality. In the OV image, a resolution of 0.7 lp/mm (line pairs per millimeter) and noise of 63.5 HU were determined. Inside the VOI, resolution was increased to 2.4 lp/mm and noise was decreased to 18.7 HU. For the performed measurements, the cumulative dose was significantly reduced in comparison to conventional scans by up to 93%. The dose of a high-quality scan, for example, was reduced from 97 to less than 7 mGy, while keeping image quality constant within the VOI. Conclusions: The proposed VOI application with two scans is an effective way to ensure high image quality within the VOI while simultaneously reducing the cumulative patient dose. (C) 2010 American Association of Physicists in Medicine. [DOI: 10.1118/1.3427641]
引用
收藏
页码:2719 / 2730
页数:12
相关论文
共 50 条
  • [31] Image quality and effective dose of a robotic flat panel 3D C-arm vs computed tomography
    Kraus, Michael
    Fischer, Eric
    Gebhard, Florian
    Richter, Peter H.
    [J]. INTERNATIONAL JOURNAL OF MEDICAL ROBOTICS AND COMPUTER ASSISTED SURGERY, 2016, 12 (04): : 743 - 750
  • [32] Change your Angle of ViewSinusoidal C-Arm Movement in Cranial Flat-panel CT to Improve Image Quality
    Hannes Luecking
    Philip Hoelter
    Stefan Lang
    Manuel Schmidt
    Felix Eisenhut
    Arnd Doerfler
    [J]. Clinical Neuroradiology, 2022, 32 : 1109 - 1115
  • [33] Image Quality Evaluation for a Robot-Driven C-Arm CT System
    Prell, D.
    Kyriakou, Y.
    Kalender, W. A.
    [J]. WORLD CONGRESS ON MEDICAL PHYSICS AND BIOMEDICAL ENGINEERING, VOL 25, PT 2 - DIAGNOSTIC IMAGING, 2009, 25 : 394 - 397
  • [34] Trabecular structure analysis using C-arm CT: comparison with MDCT and flat-panel volume CT
    Catherine M. Phan
    Eric A. Macklin
    Miriam A. Bredella
    Monica Dadrich
    Paul Flechsig
    Albert J. Yoo
    Joshua A. Hirsch
    Rajiv Gupta
    [J]. Skeletal Radiology, 2011, 40 : 1065 - 1072
  • [35] Trabecular structure analysis using C-arm CT: comparison with MDCT and flat-panel volume CT
    Phan, Catherine M.
    Macklin, Eric A.
    Bredella, Miriam A.
    Dadrich, Monica
    Flechsig, Paul
    Yoo, Albert J.
    Hirsch, Joshua A.
    Gupta, Rajiv
    [J]. SKELETAL RADIOLOGY, 2011, 40 (08) : 1065 - 1072
  • [36] Percutaneous foot joint needle placement using a C-arm flat-panel detector CT
    Martin Wiewiorski
    Martin Thanh Long Takes
    Victor Valderrabano
    Augustinus Ludwig Jacob
    [J]. International Journal of Computer Assisted Radiology and Surgery, 2012, 7 : 177 - 180
  • [37] Percutaneous foot joint needle placement using a C-arm flat-panel detector CT
    Wiewiorski, Martin
    Takes, Martin Thanh Long
    Valderrabano, Victor
    Jacob, Augustinus Ludwig
    [J]. INTERNATIONAL JOURNAL OF COMPUTER ASSISTED RADIOLOGY AND SURGERY, 2012, 7 (02) : 177 - 180
  • [38] Image Quality Assessment for Extended-Volume C-arm CT Using a Multi-Turn Reverse Helix
    Yu, Z.
    Maier, A.
    Lauritsch, G.
    Vogt, F.
    Schoenborn, M.
    Koehler, C.
    Hornegger, J.
    Noo, F.
    [J]. 2014 IEEE NUCLEAR SCIENCE SYMPOSIUM AND MEDICAL IMAGING CONFERENCE (NSS/MIC), 2014,
  • [39] Hepatic Blood Volume Imaging with the Use of Flat-Detector CT Perfusion in the Angiography Suite: Comparison with Results of Conventional Multislice CT Perfusion
    Zhuang, Zhi-guo
    Zhang, Xue-bin
    Han, Jing-feng
    Beilner, Janina
    Deuerling-Zheng, Yu
    Chi, Jia-chang
    Wang, Ji
    Qian, Li-jun
    Zhou, Yan
    Xu, Jian-rong
    [J]. JOURNAL OF VASCULAR AND INTERVENTIONAL RADIOLOGY, 2014, 25 (05) : 739 - 746
  • [40] Volume CT with a flat-panel detector on a mobile, isocentric C-arm: Pre-clinical investigation in guidance of minimally invasive surgery
    Siewerdsen, JH
    Moseley, DJ
    Burch, S
    Bisland, SK
    Bogaards, A
    Wilson, BC
    Jaffray, DA
    [J]. MEDICAL PHYSICS, 2005, 32 (01) : 241 - 254