Technical Note: Confirming the prescribed angle of CT localizer radiographs and c-arm projection acquisitions

被引:1
|
作者
Szczykutowicz, Timothy P. [1 ,2 ,3 ]
Labby, Zacariah E. [4 ]
Rubert, Nicholas [2 ]
Wallace, Charles [4 ]
机构
[1] Univ Wisconsin, Dept Radiol, Madison, WI 53705 USA
[2] Univ Wisconsin, Dept Med Phys, Madison, WI 53705 USA
[3] Univ Wisconsin, Dept Biomed Engn, Madison, WI 53706 USA
[4] Univ Wisconsin, Dept Human Oncol, Madison, WI 53706 USA
关键词
quality assurance; quality control; computed tomography; cone beam CT; COMPUTED-TOMOGRAPHY; RADIATION-THERAPY; QUALITY-ASSURANCE;
D O I
10.1118/1.4940124
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Purpose: Accurate CT radiograph angle is not usually important in diagnostic CT. However, there are applications in radiation oncology and interventional radiology in which the orientation of the x-ray source and detector with respect to the patient is clinically important. The authors present a method for measuring the accuracy of the tube/detector assembly with respect to the prescribed tube/detector position for CT localizer, fluoroscopic, and general radiograph imaging using diagnostic, mobile, and c-arm based CT systems. Methods: A mathematical expression relating the x-ray projection of two metal BBs is related to gantry angle. Measurement of the BBs at a prescribed gantry (i.e., c-arm) angle can be obtained and using this relation the prescribed versus actual gantry angle compared. No special service mode or proprietary information is required, only access to projection images is required. Projection images are available in CT via CT localizer radiographs and in the interventional setting via fluorography. Results: The technique was demonstrated on two systems, a mobile CT scanner and a diagnostic CT scanner. The results confirmed a known issue with the mobile scanner and accurately described the CT localizer angle of the diagnostic system tested. Conclusions: This method can be used to quantify gantry angle, which is important when projection images are used for procedure guidance, such as in brachytherapy and interventional radiology applications. (C) 2016 American Association of Physicists in Medicine.
引用
收藏
页码:865 / 869
页数:5
相关论文
共 37 条
  • [1] Technical note: A comparison of physician doses in C-Arm and CT fluoroscopy procedures
    Troville, Jonathan
    Knott, Emily
    Reynoso-Mejia, Carlos Alberto
    Wagner, Martin
    Lee, Fred T.
    Szczykutowicz, Timothy P.
    JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, 2024, 25 (05):
  • [2] C-arm free reduction for thoracolumbar fracture: A technical note
    Tanaka, Masato
    Sonawane, Sumeet
    Sharma, Sagar
    Fujiwara, Yoshihiro
    Uotani, Koji
    Yamauchi, Taro
    Arataki, Shinya
    Ikuma, Hisanori
    INTERDISCIPLINARY NEUROSURGERY-ADVANCED TECHNIQUES AND CASE MANAGEMENT, 2021, 25
  • [3] Technical Note: Unsupervised C-arm pose tracking with radiographic fiducial
    Fallavollita, P.
    Burdette, E. C.
    Song, D. Y.
    Abolmaesumi, P.
    Fichtinger, G.
    MEDICAL PHYSICS, 2011, 38 (04) : 2241 - 2245
  • [4] Technical Feasibility of CT Perfusion using a C-Arm CBCT System
    Xu, Min
    Tang, Jie
    Royalty, Kevin
    Chen, Guang-Hong
    MEDICAL IMAGING 2013: PHYSICS OF MEDICAL IMAGING, 2013, 8668
  • [5] Removal of the retained suture needle under C-arm fluoroscopy: a technical note
    Sencimen, Metin
    Bayar, Gurkan Rasit
    Gulses, Aydin
    DENTAL TRAUMATOLOGY, 2010, 26 (06) : 527 - 529
  • [6] Percutaneous C-Arm Free O-Arm Navigated Biopsy for Spinal Pathologies: A Technical Note
    Tanaka, Masato
    Sonawane, Sumeet
    Uotani, Koji
    Fujiwara, Yoshihiro
    Sessumpun, Kittipong
    Yamauchi, Taro
    Sugihara, Shinsuke
    DIAGNOSTICS, 2021, 11 (04)
  • [7] Technical note: Workload and transmission data for mobile C-arm fluoroscopy in gastrointestinal endoscopy
    Li, Xinhua
    Marschall, Theodore A.
    Yang, Kai
    Liu, Bob
    MEDICAL PHYSICS, 2024, 51 (04) : 2461 - 2467
  • [8] Closed reduction of mandibular condyle fractures using C-arm fluoroscopy: a technical note
    Imai, Tomoaki
    Michizawa, Masahiro
    Yamamoto, Naofumi
    Kai, Tatsuro
    ORAL SURGERY ORAL MEDICINE ORAL PATHOLOGY ORAL RADIOLOGY, 2013, 115 (01): : E4 - E9
  • [9] C-arm Free O-arm Navigated Posterior Atlantoaxial Fixation in Down Syndrome: A Technical Note
    Tanaka, Masato
    Sonawane, Sumeet
    Fujiwar, Yoshihiro
    Uotani, Koji
    Arataki, Shinya
    Yamauchi, Taro
    Ye, Youchen
    Misawa, Haruo
    ACTA MEDICA OKAYAMA, 2022, 76 (01) : 71 - 78
  • [10] Aortic arch injection with C-arm cone beam CT for radiosurgery treatment planning of cerebral arteriovenous malformations: technical note
    Radvany, Martin G.
    Ehtiati, Tina
    Huang, Judy
    Mahesh, Mahadevappa
    Gailloud, Philippe
    JOURNAL OF NEUROINTERVENTIONAL SURGERY, 2012, 4 (05)