Comparison of hemodynamic responses to endotracheal intubation with the GlideScope video laryngoscope and Macintosh laryngoscope in patients undergoing cardiovascular surgery

被引:4
|
作者
Inangil, Gokhan [1 ]
Cansiz, Kadir Hakan [1 ]
Gurbuz, Fuat [1 ]
Bakal, Omer [1 ]
Merih Gokben, Fatma [1 ]
Sen, Huseyin [1 ]
机构
[1] Univ Hlth Sci, Sultan Abdulhamid Han Training & Res Hosp, Dept Anesthesiol & Reanimat, Istanbul, Turkey
关键词
Cardiovascular surgery; GlideScope; hemodynamic response; intubation; laryngoscopy; TRACHEAL INTUBATION; OROTRACHEAL INTUBATION; CLINICAL-ASSESSMENT; ANESTHESIA; FENTANYL; VIDEOLARYNGOSCOPE; PREDICTORS; LIDOCAINE; AIRWAY; TRIAL;
D O I
10.5606/tgkdc.dergisi.2018.15716
中图分类号
R61 [外科手术学];
学科分类号
摘要
Background: This study aimed to compare hemodynamic response to endotracheal intubation using GlideScope and Macintosh laryngoscopes in patients who underwent cardiovascular surgery. Methods: A total of 74 patients were enrolled in the study. Patients were randomly assigned into two groups either a GlideScope (n=37) or a Macintosh laryngoscope (n=37). Laryngoscopy time, heart rate, invasive arterial pressure and rate pressure product were compared during induction and until five minutes after laryngoscopy at 11 time points. Results: Seventy patients completed the study. The intubation time was similar in both groups (14.1 +/- 4.0 sec vs. 13.2 +/- 4.2 sec; p=0.22). Hemodynamic values including heart rate, systolic, diastolic and mean arterial pressure and rate pressure product were similar at all-time points. The laryngoscopy time was also similar between groups. Conclusion: The GlideScope video laryngoscope did not show any advantage on hemodynamic response in patients undergoing cardiovascular surgery when compared to Macintosh laryngoscope.
引用
收藏
页码:386 / 393
页数:8
相关论文
共 50 条
  • [1] Comparison of the GlideScope video laryngoscope and Macintosh laryngoscope in simulated tracheal intubation scenarios
    Kim, H. J.
    Chung, S. P.
    Park, I. C.
    Cho, J.
    Lee, H. S.
    Park, Y. S.
    [J]. EMERGENCY MEDICINE JOURNAL, 2008, 25 (05) : 279 - 282
  • [2] Comparison of hemodynamic responses to orotracheat intubation with the GlideScope® videolaryngoscope and the Macintosh direct laryngoscope
    Xue, Fu S.
    Zhang, Guo H.
    Li, Xuan Y.
    Sun, Hai T.
    Li, Ping
    Li, Cheng W.
    Liu, Kun P.
    [J]. JOURNAL OF CLINICAL ANESTHESIA, 2007, 19 (04) : 245 - 250
  • [3] Comparing GlideScope Video Laryngoscope and Macintosh Laryngoscope Regarding Hemodynamic Responses During Orotracheal Intubation: A Randomized Controlled Trial
    Pournajafian, Ali Reza
    Ghodraty, Mohammad Reza
    Faiz, Seyed Hamid Reza
    Rahimzadeh, Poupak
    Goodarzynejad, Hamidreza
    Dogmehchi, Enseyeh
    [J]. IRANIAN RED CRESCENT MEDICAL JOURNAL, 2014, 16 (04)
  • [4] A Randomized Comparison Simulating Face to Face Endotracheal Intubation of Pentax Airway Scope, C-MAC Video Laryngoscope, Glidescope Video Laryngoscope, and Macintosh Laryngoscope
    Choi, Hyun Young
    Oh, Young Min
    Kang, Gu Hyun
    Kang, Hyunggoo
    Jang, Yong Soo
    Kim, Wonhee
    Kim, Euichung
    Cho, Young Soon
    Choi, Hyukjoong
    Kim, Hyunjong
    Kim, Gyoung Yong
    [J]. BIOMED RESEARCH INTERNATIONAL, 2015, 2015
  • [5] A Comparison of the GlideScope With the Macintosh Laryngoscope for Nasotracheal Intubation in Patients With Ankylosing Spondylitis
    Xu Lili
    Hu Zhiyong
    Shen Jianjun
    [J]. JOURNAL OF NEUROSURGICAL ANESTHESIOLOGY, 2014, 26 (01) : 27 - 31
  • [6] Circulatory responses to nasotracheal intubation:: comparison of GlideScope® videolaryngoscope and Macintosh direct laryngoscope
    Xue Fu-Shan
    Li Xuan-Ying
    Liu Qian-Jin
    Liu He-Ping
    Yang Quan-Yong
    Xu Ya-Chao
    Liao Xu
    Liu Yi
    [J]. CHINESE MEDICAL JOURNAL, 2008, 121 (14) : 1290 - 1296
  • [7] HEMODYNAMIC COMPARISON OF ENDOTRACHEAL INTUBATION - MACINTOSH LARYNGOSCOPE VERSUS AUGUSTINE-GUIDE
    KOVAC, A
    MAYE, J
    DEVANE, G
    CALKINS, J
    [J]. ANESTHESIA AND ANALGESIA, 1994, 78 (02): : U124 - U124
  • [8] A comparison of the video laryngoscope to the direct laryngoscope for endotracheal intubation in emergency department patients
    Sun, Yongbo
    Zhang, Shiming
    Lin, Bin
    [J]. INTERNATIONAL JOURNAL OF CLINICAL AND EXPERIMENTAL MEDICINE, 2020, 13 (11): : 9151 - 9158
  • [9] Comparison of the GlideScope® with the Macintosh laryngoscope for tracheal intubation in patients with simulated difficult airway
    Lim, Y
    Yeo, SW
    [J]. ANAESTHESIA AND INTENSIVE CARE, 2005, 33 (02) : 243 - 247
  • [10] Comparison of GlideScope video laryngoscopy and Macintosh laryngoscope in ear–nose and throat surgery
    G. Misirlioglu
    O. Sen
    [J]. Irish Journal of Medical Science (1971 -), 2016, 185 : 729 - 733