An In Vitro Assessment of Acoustic Radiation Force Impulse Imaging for Visualizing Cardiac Radiofrequency Ablation Lesions

被引:43
|
作者
Eyerly, Stephanie A. [1 ]
Hsu, Stephen J. [1 ]
Agashe, Shruti H. [1 ]
Trahey, Gregg E. [1 ,2 ]
Li, Yang [1 ]
Wolf, Patrick D. [1 ]
机构
[1] Duke Univ, Dept Biomed Engn, Durham, NC 27706 USA
[2] Duke Med Ctr, Dept Radiol, Durham, NC USA
关键词
acoustic radiation force impulse imaging; radiofrequency catheter ablation; atrial fibrillation; atrial arrhythmias; intracardiac echocardiography; ATRIAL-FIBRILLATION; CATHETER ABLATION; ARRHYTHMIAS; CONDUCTION; COMPLICATIONS; FEASIBILITY; TACHYCARDIA; PREDICTORS; EXPERIENCE; STIFFNESS;
D O I
10.1111/j.1540-8167.2009.01664.x
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Methods and Results: RFA lesions were created in vitro in porcine ventricular myocardium and imaged with an intracardiac ultrasound catheter transducer capable of acquiring spatially registered B-mode and ARFI images. The myocardium was sliced along the imaging plane and photographed. The maximum ARFI-induced displacement images of the lesion were normalized and spatially registered with the photograph by matching the surfaces of the tissue in the B-mode and photographic images. The lesion dimensions determined by a manual segmentation of the photographed lesion based on the visible discoloration of the tissue were compared to automatic segmentations of the ARFI image using 2 different calculated thresholds. ARFI imaging accurately localized and sized the lesions within the myocardium. Differences in the maximum lateral and axial dimensions were statistically below 2 mm and 1 mm, respectively, for the 2 thresholding methods, with mean percent overlap of 68.7 +/- 5.21% and 66.3 +/- 8.4% for the 2 thresholds used. Conclusion: ARFI imaging is capable of visualizing myocardial RFA lesion dimensions to within 2 mm in vitro. Visualizing lesions during transcatheter cardiac ablation procedures could improve the success of the treatment by imaging lesion line discontinuity and potentially reducing the required number of ablation lesions and procedure time. (J Cardiovasc Electrophysiol, Vol. 21, pp. 557-563, May 2010).
引用
收藏
页码:557 / 563
页数:7
相关论文
共 50 条
  • [31] Comparison of the Delineation of Ablation Regions with Electrode Displacement Elastography and Acoustic Radiation Force Impulse Imaging
    Yang, W.
    Varghese, T.
    Ziemlewicz, T.
    Alexander, M.
    Lubner, M.
    Hinshaw, J.
    Wells, S.
    Lee, F., Jr.
    MEDICAL PHYSICS, 2017, 44 (06) : 3238 - 3238
  • [32] Acoustic Radiation Force Impulse Imaging: A Literature Review
    Iyo, Allison Y.
    JOURNAL OF DIAGNOSTIC MEDICAL SONOGRAPHY, 2009, 25 (04) : 204 - 211
  • [33] Usefulness of Acoustic Radiation Force Impulse Imaging in the Differential Diagnosis of Benign and Malignant Liver Lesions
    Tian Shuang-Ming
    Zhou Ping
    Qian Ying
    Chen Li-Rong
    Zhang Ping
    Li Rui-Zhen
    ACADEMIC RADIOLOGY, 2011, 18 (07) : 810 - 815
  • [34] BREAST LESIONS EVALUATED BY COLOR-CODED ACOUSTIC RADIATION FORCE IMPULSE (ARFI) IMAGING
    Zhou, JianQiao
    Yang, ZhiFang
    Zhan, WeiWei
    Zhang, JingWen
    Hu, Na
    Dong, YiJie
    Wang, YingYing
    ULTRASOUND IN MEDICINE AND BIOLOGY, 2016, 42 (07): : 1464 - 1472
  • [35] Acoustic Radiation Force Impulse Imaging: a New Technology for the Noninvasive Assessment of Liver Fibrosis?
    Castera, Laurent
    JOURNAL OF GASTROINTESTINAL AND LIVER DISEASES, 2009, 18 (04) : 411 - 412
  • [36] ACOUSTIC RADIATION FORCE IMPULSE IMAGING IN THE ASSESSMENT OF LIVER FIBROSIS IN PEDIATRIC PATIENTS.
    Noruegas, Maria J.
    Matos, Hugo C.
    Goncalves, Isabel
    Cipriano, Maria A.
    Sanches, Maria C.
    PEDIATRIC TRANSPLANTATION, 2011, 15 : 59 - 59
  • [37] Noninvasive Assessment of Liver Fibrosis Using Acoustic Radiation Force Impulse (ARFI) Imaging
    Bajwa, A.
    Wood, C.
    Berggruen, S.
    Miller, F.
    Ganger, D.
    Grant, T.
    AMERICAN JOURNAL OF ROENTGENOLOGY, 2012, 198 (05)
  • [38] Erratum to: Acoustic radiation force impulse-imaging in the assessment of liver fibrosis in children
    Maria José Noruegas
    Hugo Matos
    Isabel Gonçalves
    Maria Augusta Cipriano
    Conceição Sanches
    Pediatric Radiology, 2012, 42 : 641 - 641
  • [39] Comparison of Acoustic Radiation Force Impulse Elastography and Transient Elastography for the Prediction of Hepatocellular Carcinoma Recurrence after Radiofrequency Ablation
    Yoon, Jun Sik
    Kang, Gyoun Eun
    Seo, Sang Kyung
    Jeong, Bina
    Lee, Yu Rim
    Kweon, Young-Oh
    Tak, Won Young
    Jang, Se Young
    Park, Soo Young
    Hur, Keun
    Park, Jung Gil
    Lee, Hye Won
    HEPATOLOGY, 2017, 66 : 869A - 869A
  • [40] Liver stiffness measured by acoustic radiation force impulse elastography predicted prognoses of hepatocellular carcinoma after radiofrequency ablation
    Pei-Chang Lee
    Yi-You Chiou
    Nai-Chi Chiu
    Ping-Hsien Chen
    Chien-An Liu
    Wei-Yu Kao
    Teh-Ia Huo
    Yi-Hsiang Huang
    Ming-Chih Hou
    Han-Chieh Lin
    Jaw-Ching Wu
    Chien-Wei Su
    Scientific Reports, 10