Full-pupil versus divided-pupil confocal line-scanners for reflectance imaging of human skin in vivo

被引:0
|
作者
Gareau, Dan [1 ]
Abeytunge, Sanjeewa [1 ]
Rajadhyaksha, Milind [1 ]
机构
[1] Mem Sloan Kettering Canc Ctr, 1275 York Ave, New York, NY 10021 USA
关键词
D O I
10.1117/12.700563
中图分类号
TH7 [仪器、仪表];
学科分类号
0804 ; 080401 ; 081102 ;
摘要
A full-pupil confocal line-scanning microscope is under development for imaging human skin in vivo in reflectance. The new design potentially offers an alternative to current point- and line-scanners that may simplify the optics, electronics and mechanics, and lead to simpler and smaller confocal microscopes. With a combination of a cylindrical lens and an objective lens, the line-scanner creates a focused line of laser light in the object plane within tissue. An oscillating galvanometric mirror scans the focused line transverse to its axis. The backscattered light from the tissue is de-scanned and focused onto a linear CMOS detector array. Preliminary measurements of the axial line-spread function, with a 30x, 0.9-NA water immersion objective lens and illumination wavelength of 633 mn, determined the optical sectioning to be 10 pm. The new design is simple, requiring only eight optical components. However, the disadvantage is non-confocality in one dimension that results in 20% weaker sectioning than with a pointscanner, and reduced contrast in scattering tissue. The images of standard reflective targets such as a mirror and grating as well as dermis-like scattering target such as paper offer a preliminary glimpse into the performance of the line-scanner. A similar alternative design is the divided-pupil (theta) line-scanner, which provides 50% weaker sectioning than with a point scanner, but better contrast and less speckle due to the theta configuration. Such line scanners may prove useful for routine imaging of humans in clinical settings.
引用
收藏
页数:9
相关论文
共 12 条
  • [1] Line-scanning reflectance confocal microscopy of human skin: comparison of full-pupil and divided-pupil configurations
    Gareau, Daniel S.
    Abeytunge, Sanjee
    Rajadhyaksha, Milind
    OPTICS LETTERS, 2009, 34 (20) : 3235 - 3237
  • [2] Full-Pupil Line-Scanning Confocal Microscope for Imaging Weakly Scattering Tissues: Comparison to Divided-Pupil
    Gareau, Daniel S.
    Abeytunge, Sanjee
    Rajadhyaksha, Milind
    IMAGING, MANIPULATION, AND ANALYSIS OF BIOMOLECULES, CELLS, AND TISSUES VII, 2009, 7182
  • [3] Performance of full-pupil line-scanning reflectance confocal microscopy in human skin and oral mucosa in vivo
    Larson, Bjorg
    Abeytunge, Sanjeewa
    Rajadhyaksha, Milind
    BIOMEDICAL OPTICS EXPRESS, 2011, 2 (07): : 2055 - 2067
  • [4] Compact divided-pupil line-scanning confocal microscope for investigation of human tissues
    Glazowski, Christopher
    Peterson, Gary
    Rajadhyaksha, Milind
    PHOTONIC THERAPEUTICS AND DIAGNOSTICS IX, 2013, 8565
  • [5] Comparison of divided and full pupil configurations for line-scanning confocal microscopy in human skin and oral mucosa
    Larson, Bjorg
    Abeytunge, Sanjee
    Glazowski, Chris
    Rajadhyaksha, Milind
    PHOTONIC THERAPEUTICS AND DIAGNOSTICS VIII, PTS 1 AND 2, 2012, 8207
  • [6] Background and speckle suppression with a divided pupil and Nomarski prism for reflectance line-scanning confocal microscopy of human tissues
    Glazowski, Christopher
    Zavislan, James
    Rajadhyaksha, Milind
    THREE-DIMENSIONAL AND MULTIDIMENSIONAL MICROSCOPY: IMAGE ACQUISITION AND PROCESSING XIX, 2012, 8227
  • [7] Confocal reflectance theta line scanning microscope for imaging human skin in vivo
    Dwyer, PJ
    DiMarzio, CA
    Zavislan, JM
    Fox, WJ
    Rajadhyaksha, M
    OPTICS LETTERS, 2006, 31 (07) : 942 - 944
  • [8] Perfectly registered multiphoton and reflectance confocal video rate imaging of in vivo human skin
    Wang, Hequn
    Lee, Anthony M. D.
    Frehlick, Zack
    Lui, Harvey
    McLean, David I.
    Tang, Shuo
    Zeng, Haishan
    JOURNAL OF BIOPHOTONICS, 2013, 6 (04) : 305 - 309
  • [9] In vivo confocal scanning laser microscopy:: comparison of the reflectance and fluorescence mode by imaging human skin
    Meyer, Lars E.
    Otberg, Nina
    Sterry, Wolfram
    Lademann, Juergen
    JOURNAL OF BIOMEDICAL OPTICS, 2006, 11 (04)
  • [10] Image segmentation for integrated multiphoton microscopy and reflectance confocal microscopy imaging of human skin in vivo
    Chen, Guannan
    Lui, Harvey
    Zeng, Haishan
    QUANTITATIVE IMAGING IN MEDICINE AND SURGERY, 2015, 5 (01) : 17 - 22