Titanium Alloy Gamma Nail versus Biodegradable Magnesium Alloy Bionic Gamma Nail for Treating Intertrochanteric Fractures: A Finite Element Analysis

被引:15
|
作者
Li, Ming [1 ]
Zhao, Kuo [1 ]
Ding, Kai [1 ]
Cui, Yun-wei [1 ]
Cheng, Xiao-dong [1 ,2 ]
Yang, Wei-jie [1 ]
Hou, Zhi-yong [1 ]
Zhang, Ying-ze [1 ,2 ]
Chen, Wei [1 ,2 ]
Hu, Pan [3 ]
Zhu, Yan-bin [1 ,2 ]
机构
[1] Hebei Med Univ, Hosp 3, Dept Orthopaed Surg, 139 Ziqiang Rd, Shijiazhuang 050051, Hebei, Peoples R China
[2] Key Lab Biomech Hebei Prov, Shijiazhuang, Hebei, Peoples R China
[3] Peking Univ, Dept Orthopaed Surg, Peoples Hosp, 11 Xizhimen South St, Beijing 100044, Peoples R China
基金
中国国家自然科学基金;
关键词
Bionic internal fixation; Finite element analysis; Gamma nail; Intertrochanteric fracture; Trabecular bone; DYNAMIC HIP SCREW; PROXIMAL FEMUR; INTERNAL-FIXATION; BONE; OSTEOPOROSIS; IMPLANTS;
D O I
10.1111/os.12973
中图分类号
R826.8 [整形外科学]; R782.2 [口腔颌面部整形外科学]; R726.2 [小儿整形外科学]; R62 [整形外科学(修复外科学)];
学科分类号
摘要
Objective To using finite element analysis to investigate the effects of the traditional titanium alloy Gamma nail and a biodegradable magnesium alloy bionic Gamma nail for treating intertrochanteric fractures. Methods Computed tomography images of an adult male volunteer of appropriate age and in good physical condition were used to establish a three-dimensional model of the proximal femur. Then, a model of a type 31A1 intertrochanteric fracture of the proximal femur was established, and the traditional titanium alloy Gamma nails and biodegradable magnesium alloy bionic Gamma nails were used for fixation, respectively. The von Mises stress, the maximum principal stress, and the minimum principal stress were calculated to evaluate the effect of bone ingrowth on stress distribution of the proximal femur after fixation. Results In the intact model, the maximum stress was 5.8 MPa, the minimum stress was -11.7 MPa, and the von Mises stress was 11.4 MPa. The maximum principal stress distribution of the cancellous bone in the intact model appears in a position consistent with the growth direction of the principal and secondary tensile zones. After traditional Gamma nail healing, the maximum stress was 32 MPa, the minimum stress was -23.5 MPa, and the von Mises stress was 31.3 MPa. The stress concentration was quite obvious compared with the intact model. It was assumed that the nail would biodegrade completely within 12 months postoperatively. The maximum stress was 18.7 MPa, the minimum stress was -12.6 MPa, and the von Mises stress was 14.0 MPa. For the minimum principal stress, the region of minimum stress value less than -10 MPa was significantly improved compared with the traditional titanium alloy Gamma nail models. Meanwhile, the stress distribution of the bionic Gamma nail model in the proximal femur was closer to that of the intact bone, which significantly reduced the stress concentration of the implant. Conclusion The biodegradable magnesium alloy bionic Gamma nail implant can improve the stress distribution of fractured bone close to that of intact bone while reducing the risk of postoperative complications associated with traditional internal fixation techniques, and it has promising clinical value in the future.
引用
收藏
页码:1513 / 1520
页数:8
相关论文
共 50 条
  • [41] Biomechanical optimization of the magnesium alloy bionic cannulated screw for stabilizing femoral neck fractures: a finite element analysis
    Cui, Yunwei
    Ding, Kai
    Lv, Hongzhi
    Cheng, Xiaodong
    Fan, Zixi
    Sun, Dacheng
    Zhang, Yifan
    Chen, Wei
    Zhang, Yingze
    FRONTIERS IN BIOENGINEERING AND BIOTECHNOLOGY, 2024, 12
  • [42] Comparison of clinical outcomes with InterTan vs Gamma nail or PFNA in the treatment of intertrochanteric fractures: A meta-analysis
    Ma, Jian-xiong
    Kuang, Ming-jie
    Fan, Zheng-rui
    Xing, Fei
    Zhao, Yun-long
    Zhang, Lu-kai
    Chen, Heng-ting
    Han, Chao
    Ma, Xin-long
    SCIENTIFIC REPORTS, 2017, 7
  • [43] Treatment of unstable intertrochanteric femoral fractures with locking gamma nail (LGN): A retrospective cohort study
    Zhang, Lei
    Shen, Jie
    Chen, Sixu
    Wu, Zhendong
    Huang, Zaiyu
    He, Shuai
    Ying, Zhihao
    Liu, Dan
    Wang, Yong
    INTERNATIONAL JOURNAL OF SURGERY, 2016, 26 : 12 - 17
  • [44] A comparative study between gamma nail and percutaneous compression plating for the treatment of intertrochanteric hip fractures
    Crespo, E.
    Galvez, J.
    Tenias, J. M.
    Cano, I.
    Crespo, R.
    Palacios, V.
    EUROPEAN JOURNAL OF TRAUMA AND EMERGENCY SURGERY, 2012, 38 (04) : 443 - 449
  • [45] A comparative study between gamma nail and percutaneous compression plating for the treatment of intertrochanteric hip fractures
    E. Crespo
    J. Galvez
    J. M. Tenías
    I. Cano
    R. Crespo
    V. Palacios
    European Journal of Trauma and Emergency Surgery, 2012, 38 : 443 - 449
  • [46] Expandable Proximal Femoral Nail versus Gamma Proximal Femoral Nail for the treatment of hip reverse oblique fractures
    Warschawski, Yaniv
    Ankori, Ran
    Rutenberg, Tal Frenkel
    Steinberg, Ely L.
    Atzmon, Ran
    Drexler, Michael
    ARCHIVES OF ORTHOPAEDIC AND TRAUMA SURGERY, 2022, 142 (05) : 777 - 785
  • [47] Subtrochanteric femoral fractures treated with the Long Gamma3® nail: A historical control case study versus Long trochanteric Gamma nail®
    Georgiannos, D.
    Lampridis, V.
    Bisbinas, I.
    ORTHOPAEDICS & TRAUMATOLOGY-SURGERY & RESEARCH, 2015, 101 (06) : 675 - 680
  • [48] Expandable Proximal Femoral Nail versus Gamma Proximal Femoral Nail for the treatment of hip reverse oblique fractures
    Yaniv Warschawski
    Ran Ankori
    Tal Frenkel Rutenberg
    Ely L. Steinberg
    Ran Atzmon
    Michael Drexler
    Archives of Orthopaedic and Trauma Surgery, 2022, 142 : 777 - 785
  • [49] Proximal femoral nail antirotation versus InterTan nail for the treatment of intertrochanteric fractures: A systematic review and meta-analysis
    Liao, Chang-sheng
    He, Fang-zheng
    Li, Xi-yong
    Han, Peng-fei
    PLOS ONE, 2024, 19 (07):
  • [50] Finite element analysis of two cephalomedullary nails in treatment of elderly reverse obliquity intertrochanteric fractures: zimmer natural nail and proximal femoral nail antirotation-ΙΙ
    Jian Chen
    Jian-xiong Ma
    Ying Wang
    Hao-hao Bai
    Lei Sun
    Yan Wang
    Bin Lu
    Ben-chao Dong
    Ai-xian Tian
    Xin-long Ma
    Journal of Orthopaedic Surgery and Research, 14