Vascular response to bioresorbable polymer sirolimus-eluting stent vs. permanent polymer everolimus-eluting stent at 9-month follow-up: an optical coherence tomography sub-study from the CENTURY II trial

被引:30
|
作者
Kuramitsu, Shoichi [1 ]
Kazuno, Yoshio [1 ]
Sonoda, Shinjo [2 ]
Domei, Takenori [1 ]
Jinnouchi, Hiroyuki [1 ]
Yamaji, Kyohei [1 ]
Soga, Yoshimitsu [1 ]
Shirai, Shinichi [1 ]
Ando, Kenji [1 ]
Saito, Shigeru [3 ]
机构
[1] Kokura Mem Hosp, Dept Cardiol, Kokurakita Ku, 3-2-1 Asano, Kitakyushu, Fukuoka 8028555, Japan
[2] Univ Occupat & Environm Hlth Japan, Sch Med, Dept Internal Med 2, Kitakyushu, Fukuoka, Japan
[3] Shonan Kamakura Gen Hosp, Cardiol & Catheterizat Labs, Kamakura, Kanagawa, Japan
关键词
Bioresorbable polymer; Sirolimus-eluting stent; Optical coherence tomography; Coronary artery disease; DURABLE-POLYMER; THROMBOSIS; COVERAGE; RISK;
D O I
10.1093/ehjci/jev203
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Aims The Ultimaster bioresorbable polymer sirolimus-eluting stent (BP-SES) is a newly developed drug-eluting stent (DES) that consists of a thin-strut, cobalt chromium with bioresorbable polymer coated only albuminally. We sought to compare tissue coverage in coronary lesions treated with BP-SES with the XIENCE permanent polymer everolimus-eluting stent (PP-EES) using optical coherence tomography (OCT). Methods and results A total of 36 patients participated in the CENTURY II trial in our institution and were randomly assigned to BP-SES (n = 15) and PP-EES (n = 21). Of these, 27 patients (13 BP-SES and 14 PP-EES) underwent OCT at 9-month follow-up. Tissue coverage and apposition were assessed on each strut, and the results in both groups were compared using multilevel logistic or linear regression models with randomeffects at three levels: patient, lesion, and struts. A total of 6450 struts (BP-SES, n = 2951; PP-EES, n = 3499) were analysed. Thirty and 79 uncovered struts (1.02 and 2.26%, P = 0.35), and 3 and 4 malapposed struts (0.10 and 0.11%, P = 0.94) were found in BP-SES and PP-EES groups, respectively. Mean neointimal thickness did not significantly differ between both groups (110 +/- 10 vs. 93 +/- 10 mu m, P = 0.22). No significant differences in per cent neointimal volume obstruction (13.2 +/- 4.6 vs. 10.5 +/- 4.9%, P = 0.14) or other areas-volumetric parameters were detected between both groups. Conclusion BP-SES shows an excellent vascular healing response at 9-month follow-up, which is similar to PP-EES.
引用
收藏
页码:34 / 40
页数:7
相关论文
共 50 条
  • [41] Effects of hybrid polymer-coated sirolimus-eluting stent on vascular healing: an optical coherence tomography imaging study
    Kretov, E.
    Naryshkin, I.
    Ponomarev, D.
    EUROPEAN HEART JOURNAL, 2016, 37 : 491 - 491
  • [42] Comparison of Late Catch-Up Phenomenon Between Sirolimus-Eluting Stent and Everolimus-Eluting Stent During 3 Years Follow-Up - a Propensity Matched Study -
    Kobayashi, Norihiro
    Muramatsu, Toshiya
    Tsukahara, Reiko
    Ito, Yoshiaki
    Ishimori, Hiroshi
    Hirano, Keisuke
    Nakano, Masatsugu
    CIRCULATION, 2014, 130
  • [43] INCIDENCE OF LATE CATCH-UP PHENOMENON BETWEEN SIROLIMUS-ELUTING STENT AND EVEROLIMUS-ELUTING STENT DURING 3 YEARS FOLLOW-UP: A PROPENSITY MATCHED STUDY
    Kobayashi, Norihiro
    Muramatsu, Toshiya
    Tsukahara, Reiko
    Ito, Yoshiaki
    Ishimori, Hiroshi
    Hirano, Keisuke
    Nakano, Masatsugu
    JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 2015, 65 (10) : A1612 - A1612
  • [44] Comparison of in-stent responses in 8 month between durable polymer and bioabsorbable polymer everolimus-eluting stent: serial observation with angioscopy and optical coherence tomography
    Kitano, D.
    Takayama, T.
    Fukamachi, D.
    Migita, S.
    Morikawa, T.
    Kogo, T.
    Kojima, K.
    Mineki, T.
    Murata, N.
    Akutsu, N.
    Oshima, T.
    Sudo, M.
    Haruta, H.
    Hiro, T.
    Hirayama, A.
    EUROPEAN HEART JOURNAL, 2018, 39 : 528 - 528
  • [45] Bioabsorbable polymer-coated sirolimus-eluting stent implantation preserves coronary vasomotion: A DESSOLVE II trial sub-study
    Rusinaru, Dan
    Vrolix, Mathias
    Verheye, Stefan
    Chowdhary, Saqib
    Schoors, Danny
    Di Mario, Carlo
    Desmet, Walter
    Donohoe, Dennis J.
    Ormiston, John A.
    Knape, Charlene
    Bezerra, Hiram
    Lansky, Alexandra
    Wijns, William
    CATHETERIZATION AND CARDIOVASCULAR INTERVENTIONS, 2015, 86 (07) : 1141 - 1150
  • [46] Comparative efficacy and safety of novel biodegradable polymer sirolimus-eluting stents versus permanent polymer everolimus-eluting stents at 5-year follow-up
    Colleran, R.
    Michel, J.
    Giacoppo, D.
    Cassese, S.
    Joner, M.
    Kastrati, A.
    Byrne, R. A.
    EUROPEAN HEART JOURNAL, 2017, 38 : 445 - 445
  • [47] Coverage of a hydrophilic polymer-coated zotarolimus-eluting stent vs. a fluoropolymer-coated everolimus-eluting stent at 13 months: an optical coherence tomography substudy from the RESOLUTE AC trial
    Gutierrez-Chico, J. L.
    Van Geuns, R. J.
    Regar, E.
    Kelbaek, H.
    Escaned-Barbosa, J.
    Di Mario, C.
    Garcia-Garcia, H. M.
    Silber, S.
    Windecker, S.
    Serruys, P. W.
    EUROPEAN HEART JOURNAL, 2011, 32 : 405 - 406
  • [48] Temporal Change in Coronary Evagination After Implantation of Bioresorbable Polymer Sirolimus-Eluting Stent: An Mechanism-Ultimaster Sub-Study
    Tsuji, Keiko
    Ishida, Masaru
    Itoh, Tomonori
    Otake, Hiromasa
    Tsukiyama, Yoshiro
    Kimura, Takumi
    Kikuchi, Tatsuo
    Okubo, Munenori
    Hayashi, Takatoshi
    Oda, Hideto
    Uchimura, Yohei
    Kaneko, Kyosuke
    Sakamoto, Tsubasa
    Taguchi, Yuya
    Goto, Iwao
    Shinke, Toshiro
    Morino, Yoshihiro
    CIRCULATION, 2019, 140
  • [49] The Differennt Vascular Response Between Sirolimus-Eluting Stent and Everolimus-Eluting Stent in Ostial Left Circumflex Artery After Uunprotected Left Main by Means of Frequency-Domain Optical Coherence Tomography
    Fujino, Yusuke
    Attizzani, Guilherme F.
    Tahara, Satoko
    Takagi, Kensuke
    Bezerra, Hiram G.
    Nakamura, Shotaro
    Nakamura, Sunao
    CIRCULATION, 2013, 128 (22)
  • [50] Serial Evaluation of Vascular Response After Implantation of a New Sirolimus-Eluting Stent With Bioabsorbable Polymer (MISTENT): An Optical Coherence Tomography and Histopathological Study
    Attizzani, Guilherme F.
    Bezerra, Hiram G.
    Chamie, Daniel
    Fujino, Yusuke
    Spognardi, Anna-Maria
    Stanley, James R. L.
    Yamamoto, Hirosada
    Mehanna, Emile
    Wang, Wei
    Carlyle, Wenda C.
    McClain, James B.
    Costa, Marco A.
    JOURNAL OF INVASIVE CARDIOLOGY, 2012, 24 (11): : 560 - 568