An Evaluation of Commercial Pedometers for Monitoring Slow Walking Speed Populations

被引:44
|
作者
Beevi, Femina H. A. [1 ]
Miranda, Jorge [2 ]
Pedersen, Christian F. [1 ]
Wagner, Stefan [1 ]
机构
[1] Aarhus Univ, Dept Engn, Sect ECE, E301,Finlandsgade 22, DK-8200 Aarhus N, Denmark
[2] Univ Minho, Algorithm Ctr, Guimaraes, Portugal
关键词
rehabilitation; home health monitoring; telehealth; sensor technology; e-health; PHYSICAL-ACTIVITY; GAIT SPEED; TREADMILL WALKING; STEP COUNTS; ACCURACY; VALIDITY; ACCELEROMETERS; VALIDATION; PLACEMENT; ALGORITHM;
D O I
10.1089/tmj.2015.0120
中图分类号
R19 [保健组织与事业(卫生事业管理)];
学科分类号
摘要
Background: Pedometers are considered desirable devices for monitoring physical activity. Two population groups of interest include patients having undergone surgery in the lower extremities or who are otherwise weakened through disease, medical treatment, or surgery procedures, as well as the slow walking senior population. For these population groups, pedometers must be able to perform reliably and accurately at slow walking speeds. The objectives of this study were to evaluate the step count accuracy of three commercially available pedometers, the Yamax (Tokyo, Japan) Digi-Walker (R) SW-200 (YM), the Omron (Kyoto, Japan) HJ-720 (OM), and the Fitbit (San Francisco, CA) Zip (FB), at slow walking speeds, specifically at 1, 2, and 3 km/h, and to raise awareness of the necessity of focusing research on step-counting devices and algorithms for slow walking populations. Materials and Methods: Fourteen participants 29.93 +/- 4.93 years of age were requested to walk on a treadmill at the three specified speeds, in four trials of 100 steps each. The devices were worn by the participants on the waist belt. The pedometer counts were recorded, and the error percentage was calculated. Results: The error rate of all three evaluated pedometers decreased with the increase of speed: at 1 km/h the error rates varied from 87.11% (YM) to 95.98% (FB), at 2 km/h the error rates varied from 17.27% (FB) to 46.46% (YM), and at 3 km/h the error rates varied from 22.46% (YM) to a slight overcount of 0.70% (FB). Conclusions: It was observed that all the evaluated devices have high error rates at 1 km/h and mixed error rates at 2 km/h, and at 3 km/h the error rates are the smallest of the three assessed speeds, with the OM and the FB having a slight overcount. These results show that research on pedometers' software and hardware should focus more on accurate step detection at slow walking speeds.
引用
收藏
页码:441 / 449
页数:9
相关论文
共 50 条
  • [21] Development and Evaluation of a Device to Measure Walking Speed
    Caves, Kevin
    Davis, Leighanne
    DeRuyter, Frank
    [J]. COMPUTERS HELPING PEOPLE WITH SPECIAL NEEDS, ICCHP 2018, PT II, 2018, 10897 : 468 - 471
  • [22] Walking characteristics including mild motor paralysis and slow walking speed in post-stroke patients
    Mizuta, Naomichi
    Hasui, Naruhito
    Nakatani, Tomoki
    Takamura, Yusaku
    Fujii, Shintaro
    Tsutsumi, Masako
    Taguchi, Junji
    Morioka, Shu
    [J]. SCIENTIFIC REPORTS, 2020, 10 (01)
  • [23] Walking characteristics including mild motor paralysis and slow walking speed in post-stroke patients
    Naomichi Mizuta
    Naruhito Hasui
    Tomoki Nakatani
    Yusaku Takamura
    Shintaro Fujii
    Masako Tsutsumi
    Junji Taguchi
    Shu Morioka
    [J]. Scientific Reports, 10
  • [24] Comparing Self-Selected Speed Walking of the Elderly With Self-Selected Slow, Moderate, and Fast Speed Walking of Young Adults
    Kim, Woo Sub
    Kim, Eun Young
    [J]. ANNALS OF REHABILITATION MEDICINE-ARM, 2014, 38 (01): : 101 - 108
  • [25] DEMODULATION ANALYSIS FOR MONITORING SLOW-SPEED MACHINERY
    BARCLAY, J
    [J]. SOUND AND VIBRATION, 1994, 28 (02): : 8 - &
  • [26] Evaluation of a new sensor system for ambulatory monitoring of human posture and walking speed using accelerometers and gyroscope
    Motoi, K
    Tanaka, S
    Nogawa, M
    Yamakoshi, K
    [J]. SICE 2003 ANNUAL CONFERENCE, VOLS 1-3, 2003, : 1232 - 1235
  • [27] Promoting walking among office employees - evaluation of a randomized controlled intervention with pedometers and e-mail messages
    Aittasalo, Minna
    Rinne, Marjo
    Pasanen, Matti
    Kukkonen-Harjula, Katriina
    Vasankari, Tommi
    [J]. BMC PUBLIC HEALTH, 2012, 12
  • [28] Carried Weight Affects Walking Speed Monitoring with the IngVaL System
    Hellstrom, Per Anders Rickard
    Folke, Mia
    [J]. PHEALTH 2019, 2019, 261 (317-320): : 317 - 320
  • [29] Promoting walking among office employees ― evaluation of a randomized controlled intervention with pedometers and e-mail messages
    Minna Aittasalo
    Marjo Rinne
    Matti Pasanen
    Katriina Kukkonen-Harjula
    Tommi Vasankari
    [J]. BMC Public Health, 12
  • [30] Fast and slow processes underlie the selection of both step frequency and walking speed
    Pagliara, Renato
    Snaterse, Mark
    Donelan, J. Maxwell
    [J]. JOURNAL OF EXPERIMENTAL BIOLOGY, 2014, 217 (16): : 2939 - 2946