Laser-induced fluorescence of seeded nitric oxide as a flame thermometer

被引:44
|
作者
Tamura, M [1 ]
Luque, J
Harrington, JE
Berg, PA
Smith, GP
Jeffries, JB
Crosley, DR
机构
[1] SRI Int, Mol Phys Lab, Menlo Park, CA 94025 USA
[2] Tokyo Gas Co Ltd, Tokyo, Japan
来源
APPLIED PHYSICS B-LASERS AND OPTICS | 1998年 / 66卷 / 04期
关键词
D O I
10.1007/s003400050425
中图分类号
O43 [光学];
学科分类号
070207 ; 0803 ;
摘要
Gas temperatures deduced from one-line and two-line laser-induced fluorescence (LIF) measurements of NO seeded into low-pressure methane flames are compared with those determined from LIF measurements of OH rotational distribution. The purpose is to make a quantitative assessment of the accuracy of one-and two-line LIF temperature strategies. Temperature determination from NO LIF intensity from a single transition must accurately account for any variation in the mole fraction of seeded NO. In the hot post-flame region. this added NO is diluted by the thermal equilibrium of radical species and incomplete combustion; this dilution depends on gas temperature and fuel/air stoichiometry. Chemical reactions of the NO produce an even larger variation in the NO mole fraction. Reburning of NO in the flame front (via reaction with CH) reduces the added NO by as much as 16% in the burnt gases of a slightly fuel-rich (Phi = 1.07) methane/air flame. The size of the reduction depends on both the amount of NO addition and the fuel/air stoichiometry. This effect can alter the apparent temperature determined from one-line NO LIF by as much as 25%. Two-line LIF strategies for gas temperature avoid these complications. Temperatures determined from all three LIF methods show good agreement in low-pressure flames, when a detailed model of the chemistry is used to correct the one-line NO temperature. However, in reactive systems where a modal of the chemistry cannot be applied, one-line LIF strategies for gas temperature from seeded NO are not reliable.
引用
收藏
页码:503 / 510
页数:8
相关论文
共 50 条
  • [21] Mechanism for laser-induced fluorescence signal generation in a nanoparticle-seeded flow for planar flame thermometry
    Gu, D. H.
    Sun, Z. W.
    Medwell, P. R.
    Alwahabi, Z. T.
    Dally, B. B.
    Nathan, G. J.
    APPLIED PHYSICS B-LASERS AND OPTICS, 2015, 118 (02): : 209 - 218
  • [22] Mechanism for laser-induced fluorescence signal generation in a nanoparticle-seeded flow for planar flame thermometry
    D. H. Gu
    Z. W. Sun
    P. R. Medwell
    Z. T. Alwahabi
    B. B. Dally
    G. J. Nathan
    Applied Physics B, 2015, 118 : 209 - 218
  • [23] Simultaneous Nitric Oxide/Atomic Oxygen Laser-Induced Fluorescence in an Arcjet Facility
    Johansen, Craig T.
    Lincoln, Daniel A.
    Bathel, Brett F.
    Inman, Jennifer A.
    Danehy, Paul M.
    JOURNAL OF THERMOPHYSICS AND HEAT TRANSFER, 2016, 30 (04) : 912 - 918
  • [24] Nonstationary collisional dynamics in determining nitric oxide laser-induced fluorescence spectra
    Daily, JW
    Bessler, WG
    Schulz, C
    Sick, V
    Settersten, TB
    AIAA JOURNAL, 2005, 43 (03) : 458 - 464
  • [25] COLLISIONAL EFFECTS ON LASER-INDUCED FLUORESCENCE FLAME MEASUREMENTS
    CROSLEY, DR
    OPTICAL ENGINEERING, 1981, 20 (04) : 511 - 521
  • [26] Evaluation of nitric oxide laser-induced fluorescence thermometry techniques in a hypersonic boundary layer
    Connor C. McDougall
    W. Schuyler Hinman
    Craig T. Johansen
    Brett F. Bathel
    Jennifer A. Inman
    Paul M. Danehy
    Experiments in Fluids, 2020, 61
  • [27] Ultrahigh-frame-rate nitric oxide planar laser-induced fluorescence imaging
    Jiang, Naibo
    Lempert, Walter R.
    OPTICS LETTERS, 2008, 33 (19) : 2236 - 2238
  • [28] Confocal, two-photon laser-induced fluorescence technique for the detection of nitric oxide
    Reeves, M
    Musculus, M
    Farrell, P
    APPLIED OPTICS, 1998, 37 (28): : 6627 - 6635
  • [29] Evaluation of nitric oxide laser-induced fluorescence thermometry techniques in a hypersonic boundary layer
    McDougall, Connor C.
    Hinman, W. Schuyler
    Johansen, Craig T.
    Bathel, Brett F.
    Inman, Jennifer A.
    Danehy, Paul M.
    EXPERIMENTS IN FLUIDS, 2020, 61 (04)
  • [30] Confocal, two-photon laser-induced fluorescence technique for the detection of nitric oxide
    Reeves, Mark
    Musculus, Mark
    Farrell, Patrick
    Applied Optics, 1998, 37 (28): : 6627 - 6635