Homogeneous vs. patient specific breast models for Monte Carlo evaluation of mean glandular dose in mammography

被引:31
|
作者
Sarno, A. [1 ,2 ]
Mettivier, G. [1 ,2 ]
Di Lillo, F. [1 ,2 ]
Bliznakova, K. [3 ]
Sechopoulos, I. [4 ,5 ]
Russo, P. [1 ,2 ]
机构
[1] Univ Napoli Federico II, Dipartimento Fis Ettore Pancini, I-80126 Naples, Italy
[2] INFN, Sez Napoli, I-80126 Naples, Italy
[3] Tech Univ Varna, Lab Comp Simulat Med, Varna, Bulgaria
[4] Radboud Univ Nijmegen, Dept Radiol & Nucl, Nijmegen, Netherlands
[5] Dutch Expert Ctr Screening LRCB, Nijmegen, Netherlands
基金
欧盟地平线“2020”;
关键词
Breast models; Mean glandular dose; Mammography; Monte Carlo simulations; COMPUTED-TOMOGRAPHY; DOSIMETRY; TISSUE; COEFFICIENTS; COMPRESSION; RADIATION; CT; SIMULATION; ALGORITHM; PHANTOMS;
D O I
10.1016/j.ejmp.2018.04.392
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Purpose: To compare, via Monte Carlo simulations, homogeneous and non-homogenous breast models adopted for mean glandular dose (MGD) estimates in mammography vs. patient specific digital breast phantoms. Methods: We developed a GEANT4 Monte Carlo code simulating four homogenous cylindrical breast models featured as follows: (1) semi-cylindrical section enveloped in a 5-mm adipose layer; (2) semi-elliptical section with a 4-mm thick skin; (3) semi-cylindrical section with a 1.45-mm skin layer; (4) semi-cylindrical section in a 1.45-mm skin layer and 2-mm subcutaneous adipose layer. Twenty patient specific digital breast phantoms produced from a dedicated CT scanner were assumed as reference in the comparison. We simulated two spectra produced from two anode/filter combinations. An additional digital breast phantom was produced via BreastSimulator software. Results: With reference to the results for patient-specific breast phantoms and for W/Al spectra, models #1 and #3 showed higher MGD values by about 1% (ranges [-33%; + 28%] and [-31%; + 30%], respectively), while for model #4 it was 2% lower (range [-34%; + 26%]) and for model #2-11% (range [-39%; + 14%]), on average. On the other hand, for W/Rh spectra, models # 1 and # 4 showed lower MGD values by 2% and 1%, while for model # 2 and # 3 it was 14% and 8% lower, respectively (ranges [-43%; + 13%] and [-41%; + 21%]). The simulation with the digital breast phantom produced with BreastSimulator showed a MGD overestimation of + 33%. Conclusions: The homogeneous breast models led to maximum MGD underestimation and overestimation of 43% and 28%, respectively, when compared to patient specific breast phantoms derived from clinical CT scans.
引用
收藏
页码:56 / 63
页数:8
相关论文
共 50 条
  • [1] Monte Carlo Methods for Assessment of the Mean Glandular Dose in Mammography: Simulations in Homogeneous Phantoms
    Tucciariello, R. M.
    Barca, P.
    Caramella, D.
    Lamastra, R.
    Traino, C.
    Fantacci, M. E.
    [J]. PROCEEDINGS OF THE 12TH INTERNATIONAL JOINT CONFERENCE ON BIOMEDICAL ENGINEERING SYSTEMS AND TECHNOLOGIES, VOL 3 (BIOINFORMATICS), 2019, : 242 - 249
  • [2] A Monte Carlo model for mean glandular dose evaluation in spot compression mammography
    Sarno, Antonio
    Dance, David R.
    van Engen, Ruben E.
    Young, Kenneth C.
    Russo, Paolo
    Di Lillo, Francesca
    Mettivier, Giovanni
    Bliznakova, Kristina
    Fei, Baowei
    Sechopoulos, Ioannis
    [J]. MEDICAL PHYSICS, 2017, 44 (07) : 3848 - 3860
  • [3] Monte Carlo calculation of monoenergetic and polyenergetic DgN coefficients for mean glandular dose estimates in mammography using a homogeneous breast model
    Sarno, A.
    Tucciariello, R. M.
    Mettivier, G.
    di Franco, F.
    Russo, P.
    [J]. PHYSICS IN MEDICINE AND BIOLOGY, 2019, 64 (12):
  • [4] Monte Carlo Evaluation of Normalized Glandular Dose Coefficients in Mammography
    Sarno, Antonio
    Mettivier, Giovanni
    Di Lillo, Francesca
    Russo, Paolo
    [J]. BREAST IMAGING, IWDM 2016, 2016, 9699 : 190 - 196
  • [5] Evaluation of glandular dose in mammography in presence of breast cysts using Monte Carlo simulation
    Deevband, Mohammad Reza
    Kaveh, Zeinab
    Ghorbani, Mahdi
    Khajetash, Benyamin
    [J]. POLISH JOURNAL OF MEDICAL PHYSICS AND ENGINEERING, 2021, 27 (01): : 41 - 50
  • [6] Monte Carlo Simulation of Glandular Breast Dose in Mammography Using Breast CT-Derived Glandular Distributions
    Hernandez, A.
    Boone, J.
    Seibert, J.
    [J]. MEDICAL PHYSICS, 2015, 42 (06) : 3548 - 3548
  • [7] Patient Specific Average Glandular Dose in Mammography
    Patel, V.
    Highnam, R.
    Tromans, C.
    Pizzutiello, R. J.
    Destounis, S.
    [J]. MEDICAL PHYSICS, 2013, 40 (06)
  • [8] Skin models and their impact on mean glandular dose in mammography
    Massera, Rodrigo Trevisan
    Tomal, Alessandra
    [J]. PHYSICA MEDICA-EUROPEAN JOURNAL OF MEDICAL PHYSICS, 2018, 51 : 38 - 47
  • [9] Monte Carlo evaluation of glandular dose in cone-beam X-ray computed tomography dedicated to the breast: Homogeneous and heterogeneous breast models
    Sarno, Antonio
    Mettivier, Giovanni
    Tucciariello, Raffaele M.
    Bliznakova, Kristina
    Boone, John M.
    Sechopoulos, Ioannis
    Di Lillo, Francesca
    Russo, Paolo
    [J]. PHYSICA MEDICA-EUROPEAN JOURNAL OF MEDICAL PHYSICS, 2018, 51 : 99 - 107
  • [10] MONTE-CARLO CALCULATION OF CONVERSION FACTORS FOR THE ESTIMATION OF MEAN GLANDULAR BREAST DOSE
    DANCE, DR
    [J]. PHYSICS IN MEDICINE AND BIOLOGY, 1990, 35 (09): : 1211 - 1219