Technical Note: Improved CT number stability across patient size using dual-energy CT virtual monoenergetic imaging

被引:39
|
作者
Michalak, Gregory [1 ]
Grimes, Joshua [1 ]
Fletcher, Joel [1 ]
Halaweish, Ahmed [2 ]
Yu, Lifeng [1 ]
Leng, Shuai [1 ]
McCollough, Cynthia [1 ]
机构
[1] Mayo Clin, Dept Radiol, Rochester, MN 55905 USA
[2] Siemens Med Solut, Malvern, PA 19355 USA
关键词
dual-energy CT; monoenergetic; stability; EXPERIMENTAL-VERIFICATION; DIAGNOSTIC PERFORMANCE; MULTIDETECTOR CT; NOISE-REDUCTION; QUALITY; STONES; SINGLE; SUBTRACTION; SELECTION; ACCURACY;
D O I
10.1118/1.4939128
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Purpose: The purpose of this study was to evaluate, over a wide range of phantom sizes, CT number stability achieved using two techniques for generating dual-energy computed tomography (DECT) virtual monoenergetic images. Methods: Water phantoms ranging in lateral diameter from 15 to 50 cm and containing a CT number test object were scanned on a DSCT scanner using both single-energy (SE) and dual-energy (DE) techniques. The SE tube potentials were 70, 80, 90, 100, 110, 120, 130, 140, and 150 kV; the DE tube potential pairs were 80/140, 70/150Sn, 80/150Sn, 90/150Sn, and 100/150Sn kV (Sn denotes that the 150 kV beam was filtered with a 0.6 mm tin filter). Virtual monoenergetic images at energies ranging from 40 to 140 keV were produced from the DECT data using two algorithms, monoenergetic (mono) and monoenergetic plus (mono+). Particularly in large phantoms, water CT number errors and/or artifacts were observed; thus, datasets with water CT numbers outside +/- 10 HU or with noticeable artifacts were excluded from the study. CT numbers were measured to determine CT number stability across all phantom sizes. Results: Data exclusions were generally limited to cases when a SE or DE technique with a tube potential of less than 90 kV was used to scan a phantom larger than 30 cm. The 90/150Sn DE technique provided the most accurate water background over the large range of phantom sizes evaluated. Mono and mono+ provided equally improved CT number stability as a function of phantom size compared to SE; the average deviation in CT number was only 1.4% using 40 keV and 1.8% using 70 keV, while SE had an average deviation of 11.8%. Conclusions: The authors' report demonstrates, across all phantom sizes, the improvement in CT number stability achieved with mono and mono+ relative to SE. (C) 2016 American Association of Physicists in Medicine.
引用
收藏
页码:513 / 517
页数:5
相关论文
共 50 条
  • [31] Influence of beam hardening in dual-energy CT imaging: phantom study for iodine mapping, virtual monoenergetic imaging, and virtual non-contrast imaging
    Risa Kanatani
    Takashi Shirasaka
    Tsukasa Kojima
    Toyoyuki Kato
    Masateru Kawakubo
    European Radiology Experimental, 5
  • [32] Influence of beam hardening in dual-energy CT imaging: phantom study for iodine mapping, virtual monoenergetic imaging, and virtual non-contrast imaging
    Kanatani, Risa
    Shirasaka, Takashi
    Kojima, Tsukasa
    Kato, Toyoyuki
    Kawakubo, Masateru
    EUROPEAN RADIOLOGY EXPERIMENTAL, 2021, 5 (01)
  • [33] Dual-energy CT in musculoskeletal imaging: technical considerations and clinical applications
    Albano, Domenico
    Di Luca, Filippo
    D'Angelo, Tommaso
    Booz, Christian
    Midiri, Federico
    Gitto, Salvatore
    Fusco, Stefano
    Serpi, Francesca
    Messina, Carmelo
    Sconfienza, Luca Maria
    RADIOLOGIA MEDICA, 2024, 129 (07): : 1038 - 1047
  • [34] Spectral CT imaging: Technical principles of dual-energy CT and multi-energy photon-counting CT
    Greffier, Joel
    Villani, Nicolas
    Defez, Didier
    Dabli, Djamel
    Si-Mohamed, Salim
    DIAGNOSTIC AND INTERVENTIONAL IMAGING, 2023, 104 (04) : 167 - 177
  • [35] Dual-energy CT and coronary imaging
    Kay, Fernando Uliana
    CARDIOVASCULAR DIAGNOSIS AND THERAPY, 2020, 10 (04) : 1090 - 1107
  • [36] Imaging the Spine with Dual-Energy CT
    Murray N.
    Le M.
    Ebrahimzadeh O.
    Alharthy A.
    Mohammed M.F.
    Ouellette H.A.
    Khosa F.
    Current Radiology Reports, 5 (9)
  • [37] Dual-Energy CT in Oncologic Imaging
    Foti, Giovanni
    Ascenti, Giorgio
    Agostini, Andrea
    Longo, Chiara
    Lombardo, Fabio
    Inno, Alessandro
    Modena, Alessandra
    Gori, Stefania
    TOMOGRAPHY, 2024, 10 (03) : 299 - 319
  • [38] Virtual Monoenergetic Images of Dual-Energy CT-Impact on Repeatability, Reproducibility, and Classification in Radiomics
    Euler, Andre
    Laqua, Fabian Christopher
    Cester, Davide
    Lohaus, Niklas
    Sartoretti, Thomas
    dos Santos, Daniel Pinto
    Alkadhi, Hatem
    Baessler, Bettina
    CANCERS, 2021, 13 (18)
  • [39] Evaluation of Dual-Energy CT Reconstructed Virtual Monoenergetic Images for Radiation Therapy Treatment Planning
    Broekhoven, B.
    Godwin, W.
    McDonald, D.
    MEDICAL PHYSICS, 2020, 47 (06) : E534 - E534
  • [40] Optimal Virtual Monoenergetic Image in TwinBeam Dual-Energy CT for Head-And-Neck Radiotherapy
    Wang, T.
    Ghavidel, J.
    Beitler, J.
    Tang, X.
    Curran, W.
    Liu, T.
    Yang, X.
    MEDICAL PHYSICS, 2018, 45 (06) : E589 - E589