Quantifying the accuracy of deformable image registration for cone-beam computed tomography with a physical phantom

被引:17
|
作者
Wu, Richard Y. [1 ]
Liu, Amy Y. [1 ]
Williamson, Tyler D. [1 ]
Yang, Jinzhong [1 ]
Wisdom, Paul G. [1 ]
Zhu, Xiaorong R. [1 ]
Frank, Steven J. [2 ]
Fuller, Clifton D. [2 ]
Gunn, Gary B. [2 ]
Gao, Song [1 ]
机构
[1] Univ Texas MD Anderson Canc Ctr, Dept Radiat Phys, Houston, TX 77030 USA
[2] Univ Texas MD Anderson Canc Ctr, Dept Radiat Oncol, Houston, TX 77030 USA
来源
JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS | 2019年 / 20卷 / 10期
基金
美国国家卫生研究院;
关键词
adaptive plan; CBCT image registration; deformable image registration; deformable phantom; QUALITY-ASSURANCE; RADIATION-THERAPY; ALGORITHMS; IMPLEMENTATION; VALIDATION; DESIGN; HEAD;
D O I
10.1002/acm2.12717
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Purpose Kilo-voltage cone-beam computed tomography (CBCT) is widely used for patient alignment, contour propagation, and adaptive treatment planning in radiation therapy. In this study, we evaluated the accuracy of deformable image registration (DIR) for CBCT under various imaging protocols with different noise and patient dose levels. Methods A physical phantom previously developed to facilitate end-to-end testing of the DIR accuracy was used with Varian Velocity v4.0 software to evaluate the performance of image registration from CT to CT, CBCT to CT, and CBCT to CBCT. The phantom is acrylic and includes several inserts that simulate different tissue shapes and properties. Deformations and anatomic changes were simulated by changing the rotations of both the phantom and the inserts. CT images (from a head and neck protocol) and CBCT images (from pelvis, head and "Image Gently" protocols) were obtained with different image noise and dose levels. Large inserts were filled with Mobil DTE oil to simulate soft tissue, and small inserts were filled with bone materials. All inserts were contoured before the DIR process to provide a ground truth contour size and shape for comparison. After the DIR process, all deformed contours were compared with the originals using Dice similarity coefficient (DSC) and mean distance to agreement (MDA). Both large and small volume of interests (VOIs) for DIR volume selection were tested by simulating a DIR process that included whole patient image volume and clinical target volumes (CTV) only (for CTVs propagation). Results For cross-modality DIR registration (CT to CBCT), the DSC were >0.8 and the MDA were <3 mm for CBCT pelvis, and CBCT head protocols. For CBCT to CBCT and CT to CT, the DIR accuracy was improved relative to the cross-modality tests. For smaller VOIs, the DSC were >0.8 and MDA <2 mm for all modalities. Conclusions The accuracy of DIR depends on the quality of the CBCT image at different dose and noise levels.
引用
收藏
页码:92 / 100
页数:9
相关论文
共 50 条
  • [1] Deformable registration of lateral cephalogram and cone-beam computed tomography image
    Zhang, Yungeng
    Qin, Haifang
    Li, Peixin
    Pei, Yuru
    Guo, Yuke
    Xu, Tianmin
    Zha, Hongbin
    MEDICAL PHYSICS, 2021, 48 (11) : 6901 - 6915
  • [2] The Impact of Cone-Beam Computed Tomography (CBCT) Artifacts on Deformable Image Registration Algorithms
    Ueda, U.
    Hu, W.
    Pouliot, J.
    Yom, S.
    Quivey, J.
    Aubin, M.
    Chen, J.
    MEDICAL PHYSICS, 2010, 37 (06) : 3091 - +
  • [3] Estimation of Rectal Dose Using Daily Megavoltage Cone-Beam Computed Tomography and Deformable Image Registration
    Akino, Yuichi
    Yoshioka, Yasuo
    Fukuda, Shoichi
    Maruoka, Shintaroh
    Takahashi, Yutaka
    Yagi, Masashi
    Mizuno, Hirokazu
    Isohashi, Fumiaki
    Ogawa, Kazuhiko
    INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2013, 87 (03): : 602 - 608
  • [4] A Phantom Study on Target Localization Accuracy Using Cone-Beam Computed Tomography
    Yan, Hui
    Zhang, Liwei
    Yin, Fang-Fang
    CLINICAL MEDICINE INSIGHTS-ONCOLOGY, 2008, 2 : 501 - 510
  • [5] Quantitative evaluation of a cone-beam computed tomography-planning computed tomography deformable image registration method for adaptive radiation therapy
    Lawson, Joshua D.
    Schreibmann, Eduard
    Jani, Ashesh B.
    Fox, Tim
    JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, 2007, 8 (04): : 96 - 113
  • [6] Stereotactic registration using cone-beam computed tomography
    Carlson, Jonathan Dennis
    CLINICAL NEUROLOGY AND NEUROSURGERY, 2019, 182 : 107 - 111
  • [7] Unsupervised computed tomography and cone-beam computed tomography image registration using a dual attention network
    Hu, Rui
    Yan, Hui
    Nian, Fudong
    Mao, Ronghu
    Li, Teng
    QUANTITATIVE IMAGING IN MEDICINE AND SURGERY, 2022, 12 (07) : 3705 - 3716
  • [8] A study on the image registration accuracy of intrafraction cone beam computed tomography images
    Arumugam, S.
    RADIOTHERAPY AND ONCOLOGY, 2019, 133 : S566 - S566
  • [9] A NEWLY DEVELOPED PHANTOM FOR PHYSICAL QUALITY ASSURANCE OF DENTAL CONE-BEAM COMPUTED TOMOGRAPHY
    Bamba, Junko
    Ishida, Hideki
    Araki, Kazuyuki
    RADIATION PROTECTION DOSIMETRY, 2019, 183 (03) : 303 - 311
  • [10] Evaluation of the accuracy of deformable image registration on MRI with a physical phantom
    Wu, Richard Y.
    Liu, Amy Y.
    Yang, Jinzhong
    Williamson, Tyler D.
    Wisdom, Paul G.
    Bronk, Lawrence
    Gao, Song
    Grosshan, David R.
    Fuller, David C.
    Gunn, Gary B.
    Ronald Zhu, X.
    Frank, Steven J.
    JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, 2020, 21 (01): : 166 - 173