Performance evaluation of the MOLECUBES β-CUBE-a high spatial resolution and high sensitivity small animal PET scanner utilizing monolithic LYSO scintillation detectors

被引:104
|
作者
Krishnamoorthy, Srilalan [1 ]
Blankemeyer, Eric [1 ]
Mollet, Pieter [3 ]
Surti, Suleman [1 ]
Van Holen, Roel [3 ,4 ]
Karp, Joel S. [1 ,2 ]
机构
[1] Univ Penn, Dept Radiol, Philadelphia, PA 19104 USA
[2] Univ Penn, Dept Phys & Astron, Philadelphia, PA 19103 USA
[3] MOLECUBES NV, Ottergemsesteenweg Zuid 808-325, Ghent, Belgium
[4] Univ Ghent, Dept Elect & Informat Syst, Med Image & Signal Proc MEDISIP, Ghent, Belgium
来源
PHYSICS IN MEDICINE AND BIOLOGY | 2018年 / 63卷 / 15期
基金
美国国家卫生研究院;
关键词
PET; monolithic scintillation detectors; small animal PET scanner; preclinical PET; spatial resolution; MICROPET; DESIGN; DEPTH;
D O I
10.1088/1361-6560/aacec3
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
The MOLECUBES beta-CUBE scanner is the newest amongst commercially available predinical PET scanners for dedicated small animal imaging. The scanner is compact, lightweight and utilizes a small footprint to facilitate bench-top imaging. It can be used individually, or in combination with the X-CUBE CT scanner, which provides the ability to perform all necessary PET data corrections and provide fully quantitative PET images. The PET detector comprises of an 8 mm thick monolithic LYSO scintillator read-out by an array of 3 mm x 3 mm Hamamatsu silicon photomultipliers. The monolithic scintillator provides the ability to measure depth-of-interaction which aids in the development of such a compact scanner. With a scanner diameter of 7.6 cm and axial length of 13 cm it is suitable for imaging both whole-body mice and rats. This paper presents the design and imaging performance of the beta-CUBE scanner. NEMA NU4-2008 characterization and a variety of phantom and animal imaging studies to demonstrate the quantitative imaging performance of the PET scanner are presented. Spatial resolution of 1 mm is measured with a filtered-back projection reconstruction algorithm at the center of the scanner and DOI measurement helps maintain the excellent spatial resolution over the entire imaging FOV. An absolute peak sensitivity of 12.4% is measured with a 255-765 keV energy window. The scanner demonstrates good count-rate performance, with a peak NEC of 300 kcps and 160 kcps measured with similar to 900 mu Ci in the NEMA mouse and rat phantoms, respectively. Imaging data with the NEMA image quality phantom and Micro Derenzo phantoms demonstrate the ability to achieve good image quality and accurate quantitative data. Image uniformity of 7.4% and spill-over ratio of 8% were measured. The superior spatial resolution, excellent energy resolution and sensitivity also provide superior contrast recovery, with similar to 70% recovery for the 2 mm rods. While current commercial predinical PET scanners have spatial resolution in the 1-2 mm range, the 1 mm 3 volumetric resolution presents significant improvement over current commercially available predinical PET scanners. In combination with the X-CUBE scanner it provides the ability to perform fully quantitative imaging with spatially co-registered high-resolution 3D PET-CT images.
引用
收藏
页数:12
相关论文
共 50 条
  • [21] Performance of a high sensitivity PET scanner based on LSO panel detectors
    Conti, M.
    Bendriem, B.
    Casey, M.
    Eriksson, L.
    Jakoby, B.
    Jones, W. F.
    Jones, J.
    Michel, C.
    Nahmias, C.
    Panin, V.
    Rappoport, V.
    Sibomana, M.
    Townsend, D. W.
    [J]. 2005 IEEE NUCLEAR SCIENCE SYMPOSIUM CONFERENCE RECORD, VOLS 1-5, 2005, : 2501 - 2505
  • [22] A High-Sensitivity Small-Animal PET Scanner: Development and Initial Performance Measurements
    Kao, Chien-Min
    Xie, Qingguo
    Dong, Yun
    Wan, Lu
    Chen, Chin-Tu
    [J]. IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 2009, 56 (05) : 2678 - 2688
  • [23] Spatial resolution of a small cubic LYSO scintillator crystal detector with depth-of-interaction capabilities in a small animal PET scanner
    Perera, Lakshal
    Lerch, Michael
    Rosenfeld, Anatoly
    Meikle, Steve R.
    Chan, Jan
    [J]. 2007 IEEE NUCLEAR SCIENCE SYMPOSIUM CONFERENCE RECORD, VOLS 1-11, 2007, : 3754 - +
  • [24] SiliPET: Design of an ultra high resolution small animal PET scanner based on stacks of semiconductor detectors.
    Auricchio, Natalia
    Cesca, Nicola
    Di Domenico, Giovanni
    Moretti, Elena
    Sabba, Nicola
    Gambaccini, Mauro
    Zavattini, Guido
    Andritschke, Robert
    Kanbach, Gottfried
    Schopper, Florian
    [J]. 2005 IEEE NUCLEAR SCIENCE SYMPOSIUM CONFERENCE RECORD, VOLS 1-5, 2005, : 3010 - 3013
  • [25] Performance evaluation of Microinsert II - A submillimeter resolution small animal PET scanner
    Ravindranath, Bosky
    Wen, Jie
    Mathews, Aswin
    Tomov, Dimitre
    Catherall, David
    Li, Ke
    Wang, Qiang
    Komarov, Sergey
    O'Sullivan, Joseph
    Tai, Yuan-chuan
    [J]. JOURNAL OF NUCLEAR MEDICINE, 2014, 55
  • [26] Performance evaluation of Microinsert II - A submillimeter resolution small animal PET scanner
    Ravindranath, Bosky
    Wen, Jie
    Mathews, Aswin
    Tomov, Dimitre
    Catherall, David
    Li, Ke
    Wang, Qiang
    Komarov, Sergey
    O'Sullivan, Joseph
    Tai, Yuan-chuan
    [J]. JOURNAL OF NUCLEAR MEDICINE, 2014, 55
  • [27] Feasibility of a novel design of a high resolution/sensitivity animal PET scanner using thick and thin arrangement of monolithic scintillators
    Sanaat, Amirhossein
    Ay, Mohammad Reza
    Zaidi, Habib
    [J]. 2019 IEEE NUCLEAR SCIENCE SYMPOSIUM AND MEDICAL IMAGING CONFERENCE (NSS/MIC), 2019,
  • [28] A high resolution animal PET scanner using compact PS-PMT detectors
    Watanabe, M
    Okada, H
    Shimizu, K
    Omura, T
    Yoshikawa, E
    Kosugi, T
    Mori, S
    Yamashita, T
    [J]. IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 1997, 44 (03) : 1277 - 1282
  • [29] A high resolution animal PET scanner using compact PS-PMT detectors
    Watanabe, M
    Okada, H
    Shimizu, K
    Omura, T
    Yoshikawa, E
    Kosugi, T
    Mori, S
    Yamashita, T
    [J]. 1996 IEEE NUCLEAR SCIENCE SYMPOSIUM - CONFERENCE RECORD, VOLS 1-3, 1997, : 1330 - 1334
  • [30] Performance Evaluation of microPET: A High-Resolution Lutetium Oxyorthosilicate PET Scanner for Animal Imaging
    Chatziioannou, Anion F.
    Cherry, Simon R.
    Shao, Yiping
    Silverman, Robert W.
    Meadors, Ken
    Farquhar, Thomas H.
    Pedarsani, Marjan
    Phelps, Michael E.
    [J]. JOURNAL OF NUCLEAR MEDICINE, 2020, 61 : 155S - 165S