Performance evaluation of the MOLECUBES β-CUBE-a high spatial resolution and high sensitivity small animal PET scanner utilizing monolithic LYSO scintillation detectors

被引:104
|
作者
Krishnamoorthy, Srilalan [1 ]
Blankemeyer, Eric [1 ]
Mollet, Pieter [3 ]
Surti, Suleman [1 ]
Van Holen, Roel [3 ,4 ]
Karp, Joel S. [1 ,2 ]
机构
[1] Univ Penn, Dept Radiol, Philadelphia, PA 19104 USA
[2] Univ Penn, Dept Phys & Astron, Philadelphia, PA 19103 USA
[3] MOLECUBES NV, Ottergemsesteenweg Zuid 808-325, Ghent, Belgium
[4] Univ Ghent, Dept Elect & Informat Syst, Med Image & Signal Proc MEDISIP, Ghent, Belgium
来源
PHYSICS IN MEDICINE AND BIOLOGY | 2018年 / 63卷 / 15期
基金
美国国家卫生研究院;
关键词
PET; monolithic scintillation detectors; small animal PET scanner; preclinical PET; spatial resolution; MICROPET; DESIGN; DEPTH;
D O I
10.1088/1361-6560/aacec3
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
The MOLECUBES beta-CUBE scanner is the newest amongst commercially available predinical PET scanners for dedicated small animal imaging. The scanner is compact, lightweight and utilizes a small footprint to facilitate bench-top imaging. It can be used individually, or in combination with the X-CUBE CT scanner, which provides the ability to perform all necessary PET data corrections and provide fully quantitative PET images. The PET detector comprises of an 8 mm thick monolithic LYSO scintillator read-out by an array of 3 mm x 3 mm Hamamatsu silicon photomultipliers. The monolithic scintillator provides the ability to measure depth-of-interaction which aids in the development of such a compact scanner. With a scanner diameter of 7.6 cm and axial length of 13 cm it is suitable for imaging both whole-body mice and rats. This paper presents the design and imaging performance of the beta-CUBE scanner. NEMA NU4-2008 characterization and a variety of phantom and animal imaging studies to demonstrate the quantitative imaging performance of the PET scanner are presented. Spatial resolution of 1 mm is measured with a filtered-back projection reconstruction algorithm at the center of the scanner and DOI measurement helps maintain the excellent spatial resolution over the entire imaging FOV. An absolute peak sensitivity of 12.4% is measured with a 255-765 keV energy window. The scanner demonstrates good count-rate performance, with a peak NEC of 300 kcps and 160 kcps measured with similar to 900 mu Ci in the NEMA mouse and rat phantoms, respectively. Imaging data with the NEMA image quality phantom and Micro Derenzo phantoms demonstrate the ability to achieve good image quality and accurate quantitative data. Image uniformity of 7.4% and spill-over ratio of 8% were measured. The superior spatial resolution, excellent energy resolution and sensitivity also provide superior contrast recovery, with similar to 70% recovery for the 2 mm rods. While current commercial predinical PET scanners have spatial resolution in the 1-2 mm range, the 1 mm 3 volumetric resolution presents significant improvement over current commercially available predinical PET scanners. In combination with the X-CUBE scanner it provides the ability to perform fully quantitative imaging with spatially co-registered high-resolution 3D PET-CT images.
引用
收藏
页数:12
相关论文
共 50 条
  • [1] Imaging Performance of the MOLECUBES β-CUBE - a dedicated small animal PET scanner using monolithic LYSO scintillator
    Krishnamoorthy, S.
    Blankemeyer, E.
    Mollet, P.
    Surti, S.
    Holen, R. V.
    Karp, J. S.
    [J]. 2017 IEEE NUCLEAR SCIENCE SYMPOSIUM AND MEDICAL IMAGING CONFERENCE (NSS/MIC), 2017,
  • [2] Small animal PET scanner based on monolithic LYSO crystals: Performance evaluation
    Sanchez, F.
    Moliner, L.
    Correcher, C.
    Gonzalez, A.
    Orero, A.
    Carles, M.
    Soriano, A.
    Rodriguez-Alvarez, M. J.
    Medina, L. A.
    Mora, F.
    Benlloch, J. M.
    [J]. MEDICAL PHYSICS, 2012, 39 (02) : 643 - 653
  • [3] Simulated performance of a small-animal PET scanner based on monolithic scintillation detectors
    van der Laan, D. J.
    Maas, M. C.
    de Jong, H. W. A. M.
    Schaart, D. R.
    Bruyndonckx, P.
    Lemaitre, C.
    van Eijk, C. W. E.
    [J]. NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 2007, 571 (1-2): : 227 - 230
  • [4] Performance evaluation of a small field of view (FOV) high sensitivity high spatial resolution PET scanner.
    Hutchins, GD
    Simon, AJ
    Winkle, W
    Carlson, K
    [J]. JOURNAL OF NUCLEAR MEDICINE, 1996, 37 (05) : 333 - 333
  • [5] Design and performance evaluation of high resolution small animal PET scanner based on monolithic crystal: a simulation study
    Sanaat, A.
    Zafarghandi, M. S.
    Ay, M. R.
    [J]. JOURNAL OF INSTRUMENTATION, 2019, 14 (01)
  • [6] Preliminary performance evaluation of a high resolution small animal PET scanner with monolithic crystals and depth-of-interaction encoding
    Balcerzyk, Marcin
    Kontaxakis, George
    Delgado, Mercedes
    Garcia, Luis
    Benlloch, Jose M.
    Pozo, Miguel A.
    [J]. 8TH IEEE INTERNATIONAL CONFERENCE ON BIOINFORMATICS AND BIOENGINEERING, VOLS 1 AND 2, 2008, : 944 - +
  • [7] A comparison of BGO, GSO, MLS, LGSO, LYSO and LSO scintillation materials for high-spatial-resolution animal PET detectors
    Ramirez, Rocio A.
    Wong, Wai-Hoi
    Kim, Soonseok
    Baghaei, Hossain
    Li, Hongdi
    Wang, Yu
    Zhang, Yuxuan
    Liu, Shitao
    Liu, Jiguo
    [J]. 2005 IEEE NUCLEAR SCIENCE SYMPOSIUM CONFERENCE RECORD, VOLS 1-5, 2005, : 2835 - 2839
  • [8] Spatial Resolution and Sensitivity of the Inveon Small-Animal PET Scanner
    Visser, Eric P.
    Disselhorst, Jonathan A.
    Brom, Maarten
    Laverman, Peter
    Gotthardt, Martin
    Oyen, Wim J. G.
    Boerman, Otto C.
    [J]. JOURNAL OF NUCLEAR MEDICINE, 2009, 50 (01) : 139 - 147
  • [9] The 100μPET project: A small-animal PET scanner for ultra-high resolution molecular imaging with monolithic silicon pixel detectors
    Cadoux, Franck
    Cardella, Roberto
    Iacobucci, Giuseppe
    Iodice, Luca
    Ferrere, Didier
    Gonzalez-Sevilla, Sergio
    Kugathasan, Thanushan
    Martinelli, Fulvio
    Paolozzi, Lorenzo
    Picardi, Antonio
    Saidi, Jihad
    Vicente, Mateus
    Zambito, Stefano
    [J]. NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 2023, 1048
  • [10] Performance characterization of a high-sensitivity small-animal PET scanner
    Kao, Chien-Min
    Xie, Qingguo
    Dong, Yun
    Wan, Lu
    Chen, Chin-Tu
    [J]. 2007 IEEE NUCLEAR SCIENCE SYMPOSIUM CONFERENCE RECORD, VOLS 1-11, 2007, : 4176 - +