Slice-based plan evaluation methods for three dimensional conformal radiotherapy treatment planning

被引:4
|
作者
Prabhakar, R. [1 ]
Rath, G. K. [1 ]
机构
[1] All India Inst Med Sci, Dept Radiat Oncol, Inst Rotary Canc Hosp, New Delhi 110029, India
关键词
radiotherapy; 3D-CRT; dose volume histogram; conformality index; equivalent uniform dose; VOLUME HISTOGRAM ANALYSIS; COMPLICATION PROBABILITIES; DISTRIBUTIONS; THERAPY; BLADDER; RECTUM;
D O I
10.1007/BF03179244
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
Dose volume histograms (DVHs) play a vital role in determining the optimal plan for radiotherapy treatment delivery. The current concepts of conformality index (CI), equivalent uniform dose (EUD) derived from dose volume histogram (DVH) does not provide any spatial information. In this study, slice-based evaluation methods have been proposed for spatially analyzing the radiotherapy treatment plans. A case of prostate cancer has been selected for demonstrating the proposed tools for evaluating the dose distribution. Three dimensional conformal radiotherapy treatment planning (3D-CRT) was performed to a dose of 27 Gy/15# with three fields (6 MV anteroposterior and two 15 MV lateral fields) employing multileaf collimator after delivering 45 Gy/25#. The dose was normalized to isocenter and the treatment plan was evaluated with DVH. The dose maximum point, conformality index, planning target volume coverage index (PCI), planning target volume overdose index (POI) and equivalent uniform dose (EUD) were evaluated for every single slice along the cranio-caudal direction for all the planning target volume (PTV) contours and plotted against the slice location. The dose maximum point plotted against the slice position helps in identifying the slices where the dose maximum point is outside the target volume. The plot of conformality index gives the information about the location of those slices where excess of surrounding normal tissues is encompassed inside the prescription isodose. POI quantifies the high dose regions inside the PTV slices that receive doses above 107% of the prescription dose. Similarly, the plot of PCI and EUD with slice position gives the information about those slices where the tumor is not covered adequately. The proposed methods in this study forms as a simpler way to assess the spatial distribution of the dose inside the target volume. It could be used in combination with the current plan evaluation tools and will be very helpful in analyzing the treatment plans.
引用
下载
收藏
页码:233 / 239
页数:7
相关论文
共 50 条
  • [31] Formulaic Approach to Treatment Planning of Non-small Cell Lung Cancer with Three-Dimensional Conformal Radiotherapy
    Zusag, Thomas
    Chen, Sea
    Kirk, Mike
    AMERICAN JOURNAL OF CLINICAL ONCOLOGY-CANCER CLINICAL TRIALS, 2008, 31 (06): : 620 - 620
  • [32] Treatment planning for the layer-stacking irradiation system for three-dimensional conformal heavy-ion radiotherapy
    Kanematsu, N
    Endo, M
    Futami, Y
    Kanai, T
    Asakura, H
    Oka, H
    Yusa, K
    MEDICAL PHYSICS, 2002, 29 (12) : 2823 - 2829
  • [33] Supporting paper: Treatment plan optimisation in conformal radiotherapy
    Rowbottom, C
    Khoo, V
    Webb, S
    PROCEEDINGS OF THE 22ND ANNUAL INTERNATIONAL CONFERENCE OF THE IEEE ENGINEERING IN MEDICINE AND BIOLOGY SOCIETY, VOLS 1-4, 2000, 22 : 127 - 132
  • [34] Time requirements in conformal radiotherapy treatment planning
    Panten, T
    Höss, A
    Bohsung, J
    Becker, G
    Sroka-Pérez, G
    RADIOTHERAPY AND ONCOLOGY, 1999, 51 (03) : 211 - 214
  • [35] TOMOTHERAPY - INVERSE TREATMENT PLANNING FOR CONFORMAL RADIOTHERAPY
    HOLMES, TW
    MACKIE, TR
    SWERDLOFF, S
    RECKWERDT, PJ
    KINSELLA, TJ
    INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 1993, 27 : 139 - 139
  • [36] Superiority of intensity modulated radiotherapy over best three-dimensional conformal radiotherapy combined with brachytherapy in treatment of nasopharyngeal carcinoma: A planning study
    Taheri-Kadkhoda, Z.
    Pettersson, N.
    Johansson, K-A.
    RADIOTHERAPY AND ONCOLOGY, 2006, 81 : S380 - S381
  • [37] Methods for transferring patient and plan data between radiotherapy treatment planning systems
    Bedford, JL
    Oldham, M
    Hoess, A
    Evans, PM
    Shentall, GS
    Webb, S
    BRITISH JOURNAL OF RADIOLOGY, 1997, 70 (835): : 740 - 749
  • [38] Treatment planning evaluation of non-coplanar techniques for conformal radiotherapy of the prostate
    Bedford, JL
    Henrys, AJ
    Dearnaley, DP
    Khoo, VS
    RADIOTHERAPY AND ONCOLOGY, 2005, 75 (03) : 287 - 292
  • [39] Postoperative radiotherapy for prostate cancer - evaluation of target motion and treatment technique (intensity modulated versus three-dimensional conformal radiotherapy)
    Siluschek, J.
    Pinkawa, M.
    Piroth, M. D.
    Gagel, B.
    Eble, M. J.
    EJC SUPPLEMENTS, 2005, 3 (02): : 242 - 242
  • [40] A dosimetric evaluation of intensity modulated radiotherapy and three-dimensional conformal radiotherapy for prostate cancer in Ghana
    Kyei, Kofi Adesi
    Daniels, Joseph
    Adom, Ameyaw Kwame
    Odonkor, Philip
    Nyantakyi, Andrew Yaw
    Adjabu, Dorothy Ekua
    ECANCERMEDICALSCIENCE, 2024, 18