A study and optimization of lumbar spine X-ray imaging systems

被引:15
|
作者
McVey, G
Sandborg, M
Dance, DR
Carlsson, GA
机构
[1] Royal Marsden NHS Trust, Joint Dept Phys, London SW3 6JJ, England
[2] Linkoping Univ, Fac Hlth Sci, Dept Radiat Phys, SE-58185 Linkoping, Sweden
来源
BRITISH JOURNAL OF RADIOLOGY | 2003年 / 76卷 / 903期
关键词
D O I
10.1259/bjr/52734084
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
A Monte Carlo program has been developed that incorporates a voxel phantom of an adult patient in a model of the complete X-ray imaging system, including the anti-scatter grid and screen-film receptor. This allows the realistic estimation of patient dose and the corresponding image (optical density map) for a wide range of equipment configurations. This paper focuses on the application of the program to lumbar spine anteroposterior and lateral screen-film examinations. The program has been applied to study the variation of physical image quality measures and effective dose for changing system parameters such as tube voltage, grid design and screen-film system speed. These variations form the basis for optimization of these system parameters. In our approach to optimization, the best systems are those that can match (or come close to) the calculated image quality measure of systems preferred in a recent European clinical trial, but with lower patient dose. The largest dose savings found were 21% for a 400 speed class system with a grid having a strip density of 40 cm(-1) and a grid ratio of 16. A further dose saving of 13% was possible when. a 600 speed class system was employed. The best systems found from the optimization correspond to those recommended by the European Commission guidelines on image quality criteria for diagnostic radiographic images.
引用
收藏
页码:177 / 188
页数:12
相关论文
共 50 条
  • [31] FLYING SPOT X-RAY IMAGING SYSTEMS
    STEIN, JA
    SWIFT, RD
    MATERIALS EVALUATION, 1972, 30 (07) : 137 - &
  • [32] Improved imaging for X-ray inspection systems
    Eilbert, RF
    Shi, SH
    IEEE AEROSPACE AND ELECTRONIC SYSTEMS MAGAZINE, 2005, 20 (03) : 23 - 28
  • [33] A simulation tool for X-ray imaging systems
    Bernardini, A
    Fanti, V
    Marzeddu, R
    Massazza, G
    Randaccio, P
    2004 IEEE NUCLEAR SCIENCE SYMPOSIUM CONFERENCE RECORD, VOLS 1-7, 2004, : 3071 - 3074
  • [34] X-ray imaging on the base of Kumakhov polycapillary systems for hard X-ray region
    Nikitin, AN
    HARD X-RAY AND GAMMA-RAY DETECTOR PHYSICS, OPTICS, AND APPLICATIONS, 1997, 3115 : 143 - 146
  • [35] Evaluation and optimization of multimodal X-ray imaging techniques for inverse Compton X-ray sources
    Guenther, Benedikt
    Gradl, Regine
    Jud, Christoph
    Eggl, Elena
    Kulpe, Stephanie
    Braig, Eva-Maria
    Heck, Lisa
    Brantl, Johannes
    Achterhold, Klaus
    Gleich, Bernhard
    Dierolf, Martin
    Pfeiffer, Franz
    ADVANCES IN LABORATORY-BASED X-RAY SOURCES, OPTICS, AND APPLICATIONS VII, 2019, 11110
  • [36] An imaging X-ray polarimeter for the study of galactic and extragalactic X-ray sources
    Weisskopf, Martin C.
    Bellazzini, Ronaldo
    Costa, Enrico
    Ramsey, Brian D.
    O'Dell, Stephen L.
    Elsner, Ronald F.
    Tennant, Allyn F.
    Pavlov, George G.
    Matt, Giorgio
    Kaspi, Victoria M.
    Coppi, Paolo S.
    Wu, Kinwah
    Siegmund, Oswald H. W.
    Zavlin, Vyacheslav E.
    SPACE TELESCOPES AND INSTRUMENTATION 2008: ULTRAVIOLET TO GAMMA RAY, PTS 1 AND 2, 2008, 7011
  • [37] A voxel phantom based Monte Carlo computer program for optimisation of chest and lumbar spine X ray imaging systems
    Sandborg, M
    McVey, G
    Dance, DR
    Persliden, J
    Carlsson, GA
    RADIATION PROTECTION DOSIMETRY, 2000, 90 (1-2) : 105 - 108
  • [38] Dose optimization in pediatric cardiac x-ray imaging
    Gislason, Amber J.
    Davies, Andrew G.
    Cowen, Arnold R.
    MEDICAL PHYSICS, 2010, 37 (10) : 5258 - 5269
  • [39] OPTIMIZATION OF THE SPATIAL-RESOLUTION OF MOVING OBJECT IMAGING WITH MEDICAL X-RAY SYSTEMS
    THOEN, PJ
    PROCEEDINGS OF THE SOCIETY OF PHOTO-OPTICAL INSTRUMENTATION ENGINEERS, 1982, 375 : 107 - 111
  • [40] The female spine: gender-specific aspects in x-ray imaging