Accuracy of the Ventilator Automated Displayed Respiratory Mechanics in Passive and Active Breathing Conditions: A Bench Study

被引:8
|
作者
Daoud, Ehab G. [1 ,2 ]
Katigbak, Reynaldo [3 ]
Ottochian, Marcus [4 ]
机构
[1] Kapiolani Community Coll, 4303 Diamond Head Rd, Honolulu, HI 96816 USA
[2] Univ Hawaii, John A Burns Sch Med, Honolulu, HI 96822 USA
[3] Kaiser Permanente, Kaiser Permanente Resp Dept, Honolulu, HI USA
[4] Univ Maryland, Shock Trauma Ctr, Baltimore, MD 21201 USA
关键词
least squares fitting method; lung model; muscle pressure; respiratory mechanics; simulation models; PRESSURE; SIMULATION;
D O I
10.4187/respcare.06422
中图分类号
R4 [临床医学];
学科分类号
1002 ; 100602 ;
摘要
BACKGROUND: New-generation ventilators display dynamic measures of respiratory mechanics, such as compliance, resistance, and auto-PEEP. Knowledge of the respiratory mechanics is paramount to clinicians at the bedside. These calculations are obtained automatically by using the least squares fitting method of the equation of motion. The accuracy of these calculations in static and dynamic conditions have not been fully validated or examined in different clinical conditions or various ventilator modes. METHODS: A bench study was performed by using a lung simulator to compare the ventilator automated calculations during passive and active conditions. Three clinical scenarios (normal, COPD, and ARDS) were simulated with known compliances and resistance set per respective condition: normal (compliance 50 ml/cm H2O, resistance 10 cm H2O/L/s), COPD (compliance 60 mL/cm H2O, resistance 22 cm H2O/L/s), and ARDS (compliance 30 mlJcm H2O, and resistance 13 cm H2O/L/s). Each scenario was subjected to 4 different muscle pressures (P-mus): 0, -5, -10, and -15 cm H2O. All the experiments were done using adaptive support ventilation. The resulting automated dynamic calculations of compliance and resistance were then compared based on the clinical scenarios. RESULTS: There was a small bias (average error) and level of agreement in the passive conditions in all the experiments; however, these errors and levels of agreement got progressively higher proportional to the increased P-mus. There was a strong positive correlation between P-mus and compliance measured as well as a strong negative correlation between P-mus and resistance measured. CONCLUSIONS: Automated displayed calculations of respiratory mechanics were not dependable or accurate in active breathing conditions. The calculations were clinically more reliable in passive conditions. We propose different methods of calculating P-mus, which, if incorporated into the calculations, would improve the accuracy of respiratory mechanics made via the least squares fitting method in actively breathing conditions. (C) 2019 Daedalus Enterprises
引用
收藏
页码:1555 / 1560
页数:6
相关论文
共 50 条
  • [41] Inspiratory Effort and Lung Mechanics in Spontaneously Breathing Patients with Acute Respiratory Failure due to COVID-19: A Matched Control Study
    Tonelli, Roberto
    Busani, Stefano
    Tabb, Luca
    Fantini, Riccardo
    Castaniere, Ivana
    Biagioni, Emanuela
    Mussini, Cristina
    Girardis, Massimo
    Clini, Enrico
    Marchioni, Alessandro
    AMERICAN JOURNAL OF RESPIRATORY AND CRITICAL CARE MEDICINE, 2021, 204 (06) : 725 - 728
  • [42] Early vs Late Intubation in COVID-19 Acute Respiratory Distress Syndrome: A Retrospective Study of Ventilator Mechanics, Computed Tomography Findings, and Outcomes
    Bavishi, A. A.
    Mylvaganam, R.
    Agrawal, R.
    Avery, R.
    Cuttica, M. J.
    AMERICAN JOURNAL OF RESPIRATORY AND CRITICAL CARE MEDICINE, 2021, 203 (09)
  • [43] Accuracy of different registration areas using active and passive dynamic navigation systems in dental implant surgery: An in vitro study
    Wu, Bin-Zhang
    Ma, Fei-Fei
    Yan, Xiao-Yan
    Sun, Feng
    CLINICAL ORAL IMPLANTS RESEARCH, 2024, 35 (08) : 888 - 897
  • [44] Accuracy of a noninvasive estimated continuous cardiac output measurement under different respiratory conditions: a prospective observational study
    Masashi Takakura
    Tasuku Fujii
    Tomoya Taniguchi
    Shogo Suzuki
    Kimitoshi Nishiwaki
    Journal of Anesthesia, 2023, 37 : 394 - 400
  • [45] Accuracy of a noninvasive estimated continuous cardiac output measurement under different respiratory conditions: a prospective observational study
    Takakura, Masashi
    Fujii, Tasuku
    Taniguchi, Tomoya
    Suzuki, Shogo
    Nishiwaki, Kimitoshi
    JOURNAL OF ANESTHESIA, 2023, 37 (03) : 394 - 400
  • [46] Comparison of ventilatory responses of different respiratory profiles between humidified high-flow, pressure support and T-piece spontaneous breathing trials: A bench study
    Fossat, Guillaume
    Fresnel, Emeline
    Nay, Mai-Anh
    Alejos, Roberto Martinez
    Medrinal, Clement
    Lebret, Marius
    EUROPEAN RESPIRATORY JOURNAL, 2024, 64
  • [47] Comparative study on heat transfer enhancement by turbulent impinging jet under conditions of swirl, active excitations and passive excitations
    Uddin, Naseem
    Weigand, Bernhard
    Younis, Bassam A.
    INTERNATIONAL COMMUNICATIONS IN HEAT AND MASS TRANSFER, 2019, 100 : 35 - 41
  • [48] A prospective multicentre study testing the diagnostic accuracy of an automated cough sound centred analytic system for the identification of common respiratory disorders in children
    Paul Porter
    Udantha Abeyratne
    Vinayak Swarnkar
    Jamie Tan
    Ti-wan Ng
    Joanna M. Brisbane
    Deirdre Speldewinde
    Jennifer Choveaux
    Roneel Sharan
    Keegan Kosasih
    Phillip Della
    Respiratory Research, 20
  • [49] A prospective multicentre study testing the diagnostic accuracy of an automated cough sound centred analytic system for the identification of common respiratory disorders in children
    Porter, Paul
    Abeyratne, Udantha
    Swarnkar, Vinayak
    Tan, Jamie
    Ng, Ti-wan
    Brisbane, Joanna M.
    Speldewinde, Deirdre
    Choveaux, Jennifer
    Sharan, Roneel
    Kosasih, Keegan
    Della, Phillip
    RESPIRATORY RESEARCH, 2019, 20 (1)
  • [50] Comparison of Ventilator Mechanics and Mortality Between COVID-19 Versus Non-COVID-19 Acute Respiratory Distress Syndrome Patients: A Cross-Sectional Study
    Hanif, Sadaf
    Iqbal, Madiha
    Sethi, Sher Muhammad
    Sabeen, Amber
    JOURNAL OF ACUTE MEDICINE, 2023, 13 (04) : 144 - 149