A concept of an electron transfer stopped-flow method

被引:23
|
作者
Oyama, M [1 ]
Higuchi, T [1 ]
Okazaki, S [1 ]
机构
[1] Kyoto Univ, Grad Sch Engn, Dept Chem Mat, Sakyo Ku, Kyoto 6068501, Japan
关键词
electron transfer; stopped-flow method; tris(p-bromophenyl)amine cation radical; dimerization reaction; aromatic amine cation radical;
D O I
10.1016/S1388-2481(00)00106-5
中图分类号
O646 [电化学、电解、磁化学];
学科分类号
081704 ;
摘要
A new concept of an electron transfer stopped-flow (ETSF) method is proposed. In the ETSF method, unstable cation radicals are formed via the electron transfer reactions with stable cation radicals whose oxidation potential is positive to that of unstable ones. Thus, by combining the electron transfer reactions with the stopped-flow operation, absorption spectroscopic measurements for short-lived cation radical become possible with a reduced dead time from the point of the generation of cation radicals to the point of the optical measurement, in particular, with a help of a rapid-scan spectrophotometer. By mixing an acetonitrile (AN) solution of tris(p-bromophenyl)amine cation radical (TBPA(.+)) with an AN solution of methyldiphenylamine (MDPA) or diphenylamine (HDPA), the kinetic analyses could be carried out successfully for the dimerization reactions of MDPA(.+) and HDPA(.+) as well as the observations of their absorption spectra. It was found that fast reactions whose second order rate constants are around 10(7) M(-1)s(-1) can be analyzed using the ETSF method. (C) 2000 Published by Elsevier Science S.A.
引用
下载
收藏
页码:675 / 678
页数:4
相关论文
共 50 条
  • [41] FAST STOPPED-FLOW THERMAL APPARATUS
    BERGER, RL
    BALKO, B
    BIOPHYSICAL JOURNAL, 1977, 17 (02) : A103 - A103
  • [42] Stopped-flow techniques in Ziegler catalysis
    Liu, BP
    Matsuoka, H
    Terano, M
    MACROMOLECULAR RAPID COMMUNICATIONS, 2001, 22 (01) : 1 - 24
  • [43] A millisecond infrared stopped-flow apparatus
    Jia Tang
    Feng Gai
    APPLIED SPECTROSCOPY, 2006, 60 (12) : 1477 - 1481
  • [44] Stopped-flow fluorescence polarization immunoassay
    Gómez-Hens, A
    Aguilar-Caballos, MP
    COMBINATORIAL CHEMISTRY & HIGH THROUGHPUT SCREENING, 2003, 6 (03) : 177 - 182
  • [45] COMPUTER APPLICATION TO STOPPED-FLOW KINETICS
    DAHLGREN, G
    HENDERSO.GH
    ABSTRACTS OF PAPERS OF THE AMERICAN CHEMICAL SOCIETY, 1971, (MAR-A): : 38 - &
  • [46] STOPPED-FLOW CIRCULAR-DICHROISM
    PFLUMM, M
    LUCHINS, J
    BEYCHOK, S
    METHODS IN ENZYMOLOGY, 1986, 130 : 519 - 534
  • [47] STOPPED-FLOW STUDIES ON THYMIDYLATE SYNTHASE
    MITTELSTAEDT, D
    SCHIMERLIK, M
    FEDERATION PROCEEDINGS, 1985, 44 (05) : 1812 - 1812
  • [48] PANEL DISCUSSION ON STOPPED-FLOW TECHNIQUES
    WILKINS, RG
    MARGERUM, DW
    KUSTIN, K
    DESA, RJ
    ABSTRACTS OF PAPERS OF THE AMERICAN CHEMICAL SOCIETY, 1988, 196 : 6 - INOR
  • [49] STOPPED-FLOW FLUOROMETRY WITH AVAILABLE INSTRUMENTATION
    CHEN, RF
    SCHECHTER, AN
    BERGER, RL
    ANALYTICAL BIOCHEMISTRY, 1969, 29 (01) : 68 - +
  • [50] STOPPED-FLOW CALORIMETRY FOR BIOCHEMICAL REACTIONS
    BALKO, B
    BERGER, RL
    FRIAUF, W