Quantification of left ventricular volumes and ejection fraction using freehand transthoracic three-dimensional echocardiography: Comparison with magnetic resonance imaging

被引:60
|
作者
Mannaerts, HFJ
van der Heide, JA
Kamp, O
Papavassiliu, T
Marcus, JT
Beek, A
van Rossum, AC
Twisk, J
Visser, CA
机构
[1] VU Univ Med Ctr, Dept Cardiol, NL-1007 MB Amsterdam, Netherlands
[2] VU Univ Med Ctr, Inst Cardiovasc Res, NL-1007 MB Amsterdam, Netherlands
关键词
D O I
10.1067/mje.2003.7
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Objectives: Our aim was to validate 3-dimensional echocardiography (3DE) for assessment of left ventricular (LV) end-diastolic volume, end-systolic volume (ESV), stroke volume, and ejection fraction (EF) using the freehand-acquisition method. Furthermore, LV volumes by breath hold-versus free breathing-3DE acquisition were assessed and compared with magnetic resonance imaging (MRI). Methods: From the apical position, a fan-like 3DE image was acquired during free breathing and another, thereafter, during breath hold. In 27 patients, 28 breath hold-and 24 free breathing-3DE images were acquired. A total of 17 patients underwent both MRI and 3DE. MRI contours were traced along the outer endocardial contour, including trabeculae, and along the inner endocardial contour, excluding trabeculae, from the LV volume. Results: All 28 (100%) breath hold- and 86% of free breathing-3DE acquisitions could be analyzed. Intraobserver variation (percentual bias +/- 2 SD) of end-diastolic volume, ESV, stroke volume, and EF for breath-hold 3DE was, respectively, 0.3 +/- 10.2%, 0.3 +/- 14.6%, 0.1 +/- 18.4%, and -0.1 +/- 5.8%. For free-breathing 3DE, findings were similar. A significantly better interobserver variability, however, was observed for breath-hold 3DE for ESV and EF. Comparison of breath-hold 3DE with MRI inner contour showed for end-diastolic volume, ESV, stroke volume, and EF, a percentual bias ( 2 SD) of, respectively, -13.5 +/- 26.9%, -17.7 +/- 47.8%, -10.6 +/- 43.6%, and -1.8 +/- 11.6%. Compared with the MRI outer contour, a significantly greater difference was observed, except for EF. Conclusions: 3DE using the freehand method is fast and highly reproducible for (serial) LV volume and EF measurement, and, hence, ideally suited for clinical decision making and trials. Breath-hold 3DE is superior to free-breathing 3DE regarding image quality and reproducibility. Compared with MRI, 3DE underestimates LV volumes, but not EF, which is mainly explained by differences in endocardial contour tracing by MRI (outer contour) and 3DE (inner contour) of the trabecularized endocardium. Underestimation is reduced when breath-hold 3DE is compared with inner contour analysis of the MRI dataset.
引用
收藏
页码:101 / 109
页数:9
相关论文
共 50 条
  • [41] LEFT VENTRICULAR VOLUMES AND EJECTION FRACTION BY ECHOCARDIOGRAPHY
    POMBO, JF
    TROY, BL
    RUSSELL, RO
    CIRCULATION, 1971, 43 (04) : 480 - +
  • [42] COMPARISON OF LEFT VENTRICULAR VOLUMES AND EJECTION FRACTION VALUES BETWEEN ECHOCARDIOGRAPHY AND CARDIAC MAGNETIC RESONANCE: A PILOT STUDY
    Zydzik, Lukasz
    Dadanski, Emil
    Tyjas, Kinga
    Wojciechowska, Wiktoria
    Chyrchel, Bernadeta
    Popiela, Tadeusz
    Rajzer, Marek
    JOURNAL OF HYPERTENSION, 2024, 42
  • [43] Three-dimensional echocardiographic acquisition and validity of left ventricular volumes and ejection fraction
    Jenkins, Carly
    Tsang, Wendy
    ECHOCARDIOGRAPHY-A JOURNAL OF CARDIOVASCULAR ULTRASOUND AND ALLIED TECHNIQUES, 2020, 37 (10): : 1646 - 1653
  • [44] Comparison of left ventricular volumes and ejection fraction measured by gated SPECT and by cine magnetic resonance imaging
    He, ZX
    Vick, GW
    Vaduganathan, P
    Verani, MS
    JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 1998, 31 (02) : 44A - 44A
  • [45] A Meta-Analysis and Investigation for the Source of Bias of Left Ventricular Volumes and Function by Three-Dimensional Echocardiography in Comparison With Magnetic Resonance Imaging
    Shimada, Yuichi J.
    Shiota, Takahiro
    AMERICAN JOURNAL OF CARDIOLOGY, 2011, 107 (01): : 126 - 138
  • [46] Comparison Between 3D Echocardiography and Cardiac Magnetic Resonance Imaging (CMRI) in the Measurement of Left Ventricular Volumes and Ejection Fraction
    Benameur, Narjes
    Arous, Younes
    Ben Abdallah, Nejmeddine
    Kraiem, Tarek
    CURRENT MEDICAL IMAGING, 2019, 15 (07) : 654 - 660
  • [47] The role of three-dimensional echocardiography in the left ventricular mass evaluation. A comparison with magnetic resonance imaging
    Saidova, M
    Botvina, J
    Stukalova, O
    Sinitsin, V
    JOURNAL OF HYPERTENSION, 2005, 23 : S169 - S169
  • [48] Left ventricular ejection fraction by real-time three-dimensional echocardiography
    van Dalen, B. M.
    NETHERLANDS HEART JOURNAL, 2014, 22 (09) : 380 - 382
  • [49] Accuracy of gated SPECT thallium left ventricular volumes and ejection fractions: Comparison with three-dimensional echocardiography
    Akinboboye, OO
    Coffin, LEK
    Sciacca, RR
    Bergmann, SR
    Blood, DK
    King, DL
    JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 1998, 31 (02) : 85A - 85A
  • [50] Two-dimensional strain echocardiography-derived left ventricular ejection fraction, volumes, and global systolic dyssynchrony index: Comparison with three-dimensional echocardiography
    Vaidya, Gaurang Nandkishor
    Salgado, Benjamin C.
    Badar, Faraz
    John, Anub
    Stoddard, Marcus F.
    ECHOCARDIOGRAPHY-A JOURNAL OF CARDIOVASCULAR ULTRASOUND AND ALLIED TECHNIQUES, 2019, 36 (06): : 1054 - 1065