Investigation of the effects of a carbon-fiber tabletop on the surface dose and attenuation dose for megavoltage photon beams

被引:2
|
作者
Gursoy, G. [1 ]
Eser, E. [2 ]
Yigitoglu, I [3 ]
Koc, H. [4 ]
Kahraman, F. C. [5 ]
Yamcicier, S. [1 ]
机构
[1] Ahi Evran Univ, Dept Med Serv & Tech, Vocat Sch Hlth Serv, Kirsehir, Turkey
[2] Gazi Univ, Polath Fac Arts & Sci, Dept Phys, Ankara, Turkey
[3] Gaziosmanpasa Univ, Fac Arts & Sci, Dept Phys, Tokat, Turkey
[4] Mus Alparslan Univ, Dept Elect & Elect Engn, Fac Engn, Mus, Turkey
[5] Prov Hlth Directorate, Isparta, Turkey
来源
关键词
Radiation; radiotherapy; carbon fiber tabletop; surface dose; buildup region; TREATMENT PLANNING SYSTEM; COUCH; RADIOTHERAPY; WEDGES;
D O I
10.18869/acadpub.ijrr.16.2.235
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Backgroung: Multiple beams are generally used with an increased possibility that the beam axis intersects the treatment table. Treatment tabletops are commonly made of carbon fiber due to its high mechanical strength and rigidity, low specific density, extremely light and low radiation beam attenuation properties. Purpose of this paper is investigated the dose changes in the buildup region and beam attenuation by a carbon fiber tabletop for high energy 6- and 18-MV photon beams. Materials and Methods: Measurements were performed for 10 cm x 10 cm and 20 cm x 20 cm field sizes. The surface dose and percentage depth doses (%DD) were measured by a Markus parallel plate chamber at a source-surface distance (SSD) of 100 cm for 6 MV and 18 MV photon beams. Attenuation measurements were made at the solid-water phantom for gantry angles of 0 degrees and 180 degrees rotation of the beam. Results: A carbon fiber tabletop increases the surface dose from 12.87% to 86.65% for 10 cm x 10 cm and from 8.72% to 71.16% for 20 cm x 20 cm field at 6 and 18 MV, respectively. The surface dose with the carbon fiber tabletop in an open field (0 degrees) increases with field size. Conclusion: The carbon fiber tabletop causes a substantially increased surface dose, and also significantly decreases the skin-sparing effect, which is clinically important. The dosimetric effect of the tabletop may be higher, especially for the intensity-modulated radiation therapy depending on the beam orientation.
引用
收藏
页码:235 / 241
页数:7
相关论文
共 50 条
  • [41] Investigation of Kodak extended dose range (EDR) film for megavoltage photon beam dosimetry
    Chetty, IJ
    Charland, PM
    PHYSICS IN MEDICINE AND BIOLOGY, 2002, 47 (20): : 3629 - 3641
  • [42] Surface dose measurements and comparison of unflattened and flattened photon beams
    Sigamani, Ashokkumar
    Nambiraj, Arunai
    Yadav, Girigesh
    Giribabu, Ananda
    Srinivasan, Karthikeyan
    Gurusamy, Venkadamanickam
    Raman, Kothanda
    Karunakaran, Kaviarasu
    Thiyagarajan, Rajesh
    JOURNAL OF MEDICAL PHYSICS, 2016, 41 (02) : 85 - 91
  • [43] AN ANALYTICAL EXPRESSION FOR SURFACE DOSE FOR OBLIQUELY INCIDENT PHOTON BEAMS
    GERBI, BJ
    KHAN, FM
    MEDICAL PHYSICS, 1987, 14 (03) : 490 - 490
  • [44] DOSE REDUCTION POTENTIAL OF CARBON-FIBER MATERIAL IN DIAGNOSTIC-RADIOLOGY
    SCHUPPING, RE
    FEWELL, TR
    PHILLIPS, RA
    GROSS, RE
    SHOWALTER, CK
    HAUS, RW
    DODD, GD
    ROSSETTE, W
    HUMPHRIES, LJ
    AMERICAN JOURNAL OF ROENTGENOLOGY, 1980, 134 (04) : 857 - 857
  • [45] REDUCTION OF RADIATION-DOSE FROM THE USE OF CARBON-FIBER CASSETTES
    HAJEK, P
    NOWOTNY, R
    FORTSCHRITTE AUF DEM GEBIETE DER RONTGENSTRAHLEN UND DER NUKLEARMEDIZIN, 1984, 140 (03): : 334 - 336
  • [46] The Effect of a Carbon-Fiber Couch on the Depth-Dose Curves and Transmission Properties for Megavoltage Photon BeamsDer Einfluss eines Carbon-Tisches auf die Tiefendosisverteilung und Transmissionseigenschaften hochenergetischer Photonenstrahlung
    Björn Poppe
    Ndimofor Chofor
    Antje Rühmann
    Wolfgang Kunth
    Armand Djouguela
    Ralf Kollhoff
    Kay C. Willborn
    Strahlentherapie und Onkologie, 2007, 183 (1) : 43 - 48
  • [47] Dose response of BaFBrI: Eu2+ storage phosphor plates exposed to megavoltage photon beams
    Li, H. Harold
    Gonzalez, Albin L.
    Ji, Huaying
    Duggan, Dennis M.
    MEDICAL PHYSICS, 2007, 34 (01) : 103 - 111
  • [48] Design of a Solid Nine-Voxel Calorimeter Array for Absolute Dose Measurements of Megavoltage Photon Beams
    Lambeck, J. B.
    Radtke, J.
    DeWerd, L. A.
    MEDICAL PHYSICS, 2024, 51 (10) : 7796 - 7796
  • [49] Absorbed dose measurements in the build-up region of flattened versus unflattened megavoltage photon beams
    De Puysseleyr, Annemieke
    Lechner, Wolfgang
    De Neve, Wilfried
    Georg, Dietmar
    De Wagter, Carlos
    ZEITSCHRIFT FUR MEDIZINISCHE PHYSIK, 2016, 26 (02): : 177 - 183
  • [50] Investigating discrepancies between Monte Carlo dose calculations and measurements in the buildup region for megavoltage photon beams
    Chibani, O