Evaluation of a Handheld Gluten Detection Device

被引:0
|
作者
Taylor, Steve L. [1 ]
Nordlee, Julie A. [1 ]
Jayasena, Shyamali [1 ]
Baumert, Joseph L. [1 ]
机构
[1] Univ Nebraska, Food Innovat Ctr, Food Allergy Res & Resource Program, 1901 North 21st St, Lincoln, NE 68588 USA
关键词
Celiac disease; Consumer; Detection; Gluten; Nima; Sensor; CELIAC-DISEASE; DOUBLE-BLIND; WHEAT; DERMATITIS; DIAGNOSIS; THRESHOLD; CHILDREN; FOODS;
D O I
10.4315/0362-028X.JFP-18-184
中图分类号
Q81 [生物工程学(生物技术)]; Q93 [微生物学];
学科分类号
071005 ; 0836 ; 090102 ; 100705 ;
摘要
A portable, handheld gluten detection device, the Nima sensor, is now available for consumers wishing to determine if gluten is present in food. By U.S. regulation, gluten-free foods should contain <20 ppm of gluten. Thirteen gluten-free foods (muffins, three different types of bread, three different types of pasta, puffed corn snack, ice cream, meatballs, vinegar and oil salad dressing, oatmeal, and dark chocolate) were prepared; each food was spiked on a weight to weight basis with gluten levels of 0, 5, 10, 20, 30, 40, and 100 ppm before processing or preparation. Unprocessed and processed foods were tested with the handheld gluten sensor and by two gluten-specific enzyme-linked immunosorbent assays (ELISAs) on the basis of the R5 and G12 monoclonal antibodies, respectively. The portable gluten detection device detected gluten in all food types at the 30-ppm addition level, failing to detect gluten in only 5 (6.4%) of 78 subsamples. At the 20-ppm addition level, the portable gluten detection device failed to detect gluten in one type of pasta but detected gluten residues in 63 (87.5%) of 72 other subsamples. The device was able to detect gluten at the 10-ppm addition level in 9 of the 13 food matrices (41 of 54 subsamples, 75.9%) but not in the three types of pasta and the puffed corn snack. The gluten-sensing device did not perform reliably at the 5-ppm addition level in 11 of 13 food matrices (exceptions: ice cream and muffins). In contrast, the ELISA methods were highly reliable at gluten addition levels of >= 10 ppm in all food matrices. The portable gluten detection device yielded a low percentage of false-positive results (4 of 111, 3.6%) in these food matrices. Thus, this handheld portable gluten sensor performed reliably in the detection of gluten in foods having >= 20 ppm of added gluten with only 18 (5.9%) of 306 failures, if results of the one type of pasta are excluded. The device worked with greater reliability as the gluten levels in the foods increased.
引用
收藏
页码:1723 / 1728
页数:6
相关论文
共 50 条
  • [41] Handheld Localization Device for Indoor Environments
    Bayer, Jan
    Faigl, Jan
    2020 4TH INTERNATIONAL CONFERENCE ON AUTOMATION, CONTROL AND ROBOTS (ICACR 2020), 2020, : 60 - 64
  • [42] Evaluation of a novel handheld point-of-care ultrasound device in an African emergency department
    Burleson, Samuel L.
    Swanson, Jonathan F.
    Shufflebarger, Erin F.
    Wallace, Douglas W.
    Heimann, Matthew A.
    Crosby, James C.
    Pigott, David C.
    Gullett, John P.
    Thompson, Maxwell A.
    Greene, Christopher J.
    ULTRASOUND JOURNAL, 2020, 12 (01):
  • [43] Data Filtering for Handheld Device Application
    Balitanas, Maricel O.
    INTERNATIONAL JOURNAL OF GRID AND DISTRIBUTED COMPUTING, 2009, 2 (02): : 35 - 41
  • [44] Personalized Pain Diary on a Handheld Device
    Hofmann, U. G.
    Michalski, T.
    Tronnier, V.
    Bonsanto, M. M.
    WORLD CONGRESS ON MEDICAL PHYSICS AND BIOMEDICAL ENGINEERING, VOL 25, PT 5, 2009, 25 : 133 - +
  • [45] How to Squeeze a Configurator into a Handheld Device
    Schneider, Homero M.
    Espindola, Marcos F.
    Iano, Yuzo
    PROCEEDINGS OF THE 7TH WORLD CONFERENCE ON MASS CUSTOMIZATION, PERSONALIZATION, AND CO-CREATION (MCPC 2014) - TWENTY YEARS OF MASS CUSTOMIZATION - TOWARDS NEW FRONTIERS, 2014, : 253 - 264
  • [46] Handheld measurement device for field-effect sensor structures:: Practical evaluation and limitations
    Wagner, Torsten
    Maris, Rob J.
    Ackermann, Hans-Josef
    Otto, Ralph
    Beging, Stefan
    Poghossian, Arshak
    Schoening, Michael J.
    SENSORS AND ACTUATORS B-CHEMICAL, 2007, 127 (01) : 217 - 223
  • [47] Handheld robotic device for endoscopic neurosurgery: system integration and pre-clinical evaluation
    Dimitrakakis, Emmanouil
    Dwyer, George
    Newall, Nicola
    Khan, Danyal Z.
    Marcus, Hani J.
    Stoyanov, Danail
    FRONTIERS IN ROBOTICS AND AI, 2024, 11
  • [48] A handheld device for leaf area measurement
    Gong, Aiping
    Wu, Xiang
    Qiu, Zhengjun
    He, Yong
    COMPUTERS AND ELECTRONICS IN AGRICULTURE, 2013, 98 : 74 - 80
  • [49] Handheld SFDI/polarimetric imaging device for objective evaluation of hypertrophic scars (Conference Presentation)
    Ramella-Roman, Jessica C.
    Montejo, Karla
    Sevilla, Nicole
    Stoff, Susan
    Gonzalez, Mariacarla
    Chue-Sang, Joseph
    MOLECULAR-GUIDED SURGERY: MOLECULES, DEVICES, AND APPLICATIONS III, 2017, 10049
  • [50] Clinical Validation of a Handheld Elastic Scattering Spectroscopic Device in the Evaluation of Lesions Suggestive of Melanoma
    Hartman, R.
    Chatha, K.
    Tepedino, K.
    Grant-Kels, J.
    Paasch, U.
    JOURNAL DER DEUTSCHEN DERMATOLOGISCHEN GESELLSCHAFT, 2022, 20 : 69 - 69