Patient-Specific Oscillometric Blood Pressure Measurement

被引:38
|
作者
Liu, Jiankun [1 ]
Cheng, Hao-Min [2 ]
Chen, Chen-Huan [2 ]
Sung, Shih-Hsien [2 ]
Moslehpour, Mohsen [1 ]
Hahn, Jin-Oh [3 ]
Mukkamala, Ramakrishna [1 ]
机构
[1] Michigan State Univ, Dept Elect & Comp Engn, E Lansing, MI 48824 USA
[2] Natl Yang Ming Univ, Dept Med, Sch Med, Taipei 112, Taiwan
[3] Univ Maryland, Dept Mech Engn, College Pk, MD USA
基金
美国国家卫生研究院; 美国国家科学基金会;
关键词
Arterial stiffness; blood pressure measurement; cuff; mathematical model; oscillometry; PULSE PRESSURE; VALIDATION; RELEVANCE;
D O I
10.1109/TBME.2015.2491270
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
Objective: Most automatic cuff blood pressure (BP) measurement devices are based on oscillometry. These devices estimate BP from the envelopes of the cuff pressure oscillations using fixed ratios. The values of the fixed ratios represent population averages, so the devices may only be accurate in subjects with normal BP levels. The objective was to develop and demonstrate the validity of a patient-specific oscillometric BP measurement method. Methods: The idea of the developed method was to represent the cuff pressure oscillation envelopes with a physiologic model, and then estimate the patient-specific parameters of the model, which includes BP levels, by optimally fitting it to the envelopes. The method was investigated against gold standard reference BP measurements from 57 patients with widely varying pulse pressures. A portion of the data was used to optimize the patient-specific method and a fixed-ratio method, while the remaining data were used to test these methods and a current office device. Results: The patient-specific method yielded BP root-mean-square-errors ranging from 6.0 to 9.3 mmHg. On an average, these errors were nearly 40% lower than the errors of each existing method. Conclusion: The patient-specific method may improve automatic cuff BP measurement accuracy. Significance: A patient-specific oscillometric BP measurement method was proposed and shown to be more accurate than the conventional method and a current device.
引用
收藏
页码:1220 / 1228
页数:9
相关论文
共 50 条
  • [1] Patient-Specific Oscillometric Blood Pressure Measurement: Validation for Accuracy and Repeatability
    Liu, Jiankun
    Cheng, Hao-Min
    Chen, Chen-Huan
    Sung, Shih-Hsien
    Hahn, Jin-Oh
    Mukkamala, Ramakrishna
    [J]. IEEE JOURNAL OF TRANSLATIONAL ENGINEERING IN HEALTH AND MEDICINE, 2017, 5
  • [2] Rapid oscillometric blood pressure measurement compared to conventional oscillometric measurement
    Yarows, SA
    Patel, K
    Brook, R
    [J]. BLOOD PRESSURE MONITORING, 2001, 6 (03) : 145 - 147
  • [3] Rapid oscillometric blood pressure measurement compared to conventional oscillometric measurement
    Yarows, SA
    Patek, K
    Brook, R
    [J]. JOURNAL OF HYPERTENSION, 2000, 18 : S55 - S56
  • [4] OSCILLOMETRIC MEASUREMENT OF ARTERIAL BLOOD PRESSURE
    BENSON, H
    HERD, JA
    [J]. CIRCULATION, 1969, 40 (4S3) : II43 - &
  • [5] Assessment of Algorithms for Oscillometric Blood Pressure Measurement
    Chen, S.
    Groza, V. Z.
    Bolic, M.
    Dajani, H. R.
    [J]. I2MTC: 2009 IEEE INSTRUMENTATION & MEASUREMENT TECHNOLOGY CONFERENCE, VOLS 1-3, 2009, : 1711 - 1715
  • [6] A Developed Algorithm for Oscillometric Blood Pressure Measurement
    Chen, Min
    Wang, Qun
    Mei, Zhe
    Liu, Zhiwen
    [J]. 2013 IEEE INTERNATIONAL CONFERENCE OF IEEE REGION 10 (TENCON), 2013,
  • [7] Increasing the Accuracy of Oscillometric Blood Pressure Measurement
    Sutyagina, Anastasiya D.
    Shlyakhotka, Anna V.
    Lyakhova, Ekaterina K.
    [J]. PROCEEDINGS OF THE 2018 IEEE CONFERENCE OF RUSSIAN YOUNG RESEARCHERS IN ELECTRICAL AND ELECTRONIC ENGINEERING (EICONRUS), 2018, : 1231 - 1234
  • [8] Personalization of the Oscillometric Blood-Pressure Measurement
    Nagy, P.
    Jobbagy, A.
    [J]. WORLD CONGRESS ON MEDICAL PHYSICS AND BIOMEDICAL ENGINEERING 2018, VOL 2, 2019, 68 (02): : 885 - 888
  • [9] Oscillometric blood pressure measurement: Time for a rethink?
    Runcie, CJ
    [J]. ANESTHESIA AND ANALGESIA, 1996, 83 (04): : 886 - 887
  • [10] The oscillometric blood pressure measurement - mathematical modeling
    Juroszek, Barbara
    [J]. PRZEGLAD ELEKTROTECHNICZNY, 2008, 84 (12): : 304 - 307