Comparison of SPECT/CT, SPECT, and planar imaging with single- and dual-phase 99mTc-Sestamibi parathyroid scintigraphy

被引:214
|
作者
Lavely, William C.
Goetze, Sibyll
Friedman, Kent P.
Leal, Jeffrey P.
Zhang, Zhe
Garret-Mayer, Elizabeth
Dackiw, Alan P.
Tufano, Ralph P.
Zeiger, Martha A.
Ziessman, Harvey A.
机构
[1] Johns Hopkins Univ, Russell H Morgan Dept Radiol & Radiol Sci, Div Nucl Med, Baltimore, MD 21218 USA
[2] Johns Hopkins Univ, Dept Oncol & Biostat, Baltimore, MD 21218 USA
[3] Johns Hopkins Univ, Sch Med, Dept Otolaryngol Head & Neck Surg, Baltimore, MD 21205 USA
关键词
hyperparathyroidism; parathyroid adenoma; parathyroid scintigraphy; SPECT/CT; minimally invasive parathyroidectomy;
D O I
10.2967/jnumed.107.040428
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Various methodologies for Tc-99m-sestamibi parathyroid scintigraphy are in clinical use. There are few direct comparisons between the different methods and even less evidence supporting the superiority of one over another. Some reports suggest that SPECT is superior to planar imaging. The addition of CT to SPECT may further improve parathyroid adenoma localization. The purpose of our investigation was to compare hybrid SPECT/CT, SPECT, and planar imaging and to determine whether dual-phase imaging is advantageous for the 3 methodologies. Methods: Scintigraphy was performed on 110 patients with primary hyperparathyroidism and no prior neck surgery. Of these, 98 had single adenomas and are the subject of this review. Planar imaging and SPECT/CT were performed at 15 min and 2 h after injection. Six image sets (early and delayed planar imaging, SPECT, and SPECT/CT) and combinations of the 2 image sets were reviewed for adenoma localization at 13 possible sites. Each review was scored for location and certainty of focus by 2 reviewer groups. Surgical location served as the standard. Sensitivity, specificity, area under the curve, positive predictive value, negative predictive value, and kappa-values were determined for each method. Results: The overall kappa-coefficient (certainty of adenoma focus) between reading groups was 0.68 (95% confidence interval, 0.66-0.70). The highest values were for dual-phase studies that included SPECT/CT. Dual-phase planar imaging, SPECT, and SPECT/CT were statistically significantly superior to single-phase early or delayed imaging in sensitivity, area under the curve, and positive predictive value. Neither single-phase nor dual-phase SPECT was statistically superior to dual-phase planar imaging. Early-phase SPECT/CT in combination with any delayed imaging method was superior to dual-phase planar imaging or SPECT for sensitivity, area under the curve, and positive predictive value. Conclusion: Early SPECT/ CT in combination with any delayed imaging method was statistically significantly superior to any single- or dual-phase planar or SPECT study for parathyroid adenoma localization. Localization with dual-phase acquisition was more accurate than with single-phase Tc-99m-sestamibi scintigraphy for planar imaging, SPECT, and SPECT/CT.
引用
收藏
页码:1084 / 1089
页数:6
相关论文
共 50 条
  • [31] Parathyroid adenoma localization with 99mTc-sestamibi SPECT/CT: a meta-analysis
    Wong, Ka Kit
    Fig, Lorraine M.
    Gross, Milton D.
    Dwamena, Ben A.
    [J]. NUCLEAR MEDICINE COMMUNICATIONS, 2015, 36 (04) : 363 - 375
  • [32] A rare case of double parathyroid adenoma evidenced by 99mTc-sestaMIBI SPECT/CT
    Martinez-Valle Torres, M. D.
    Ramos Moreno, E.
    Amrani Raissouni, T.
    Sanz Viedma, S.
    Delgado Garcia, A.
    Jimenez-Hoyuela Garcia, J. M.
    [J]. REVISTA ESPANOLA DE MEDICINA NUCLEAR E IMAGEN MOLECULAR, 2017, 36 (01): : 57 - 58
  • [33] The diagnostic value of 99mTc-Sestamibi SPECT/CT in the evaluation of parathyroid adenomas or hyperplasia
    Shi, H.
    Zhang, X.
    Chen, S.
    [J]. EUROPEAN JOURNAL OF NUCLEAR MEDICINE AND MOLECULAR IMAGING, 2005, 32 : S232 - S232
  • [34] Increasing the radiochemical purity of 99mTc-sestamibi commercial preparations results in improved sensitivity of dual-phase planar parathyroid scintigraphy
    Ballinger, DR
    Cooper, MS
    [J]. NUCLEAR MEDICINE COMMUNICATIONS, 2006, 27 (06) : 543 - 544
  • [35] Parathyroid imaging with Tc-99m sestamibi planar and SPECT scintigraphy
    Nguyen, BD
    [J]. RADIOGRAPHICS, 1999, 19 (03) : 601 - 614
  • [36] The value of 99mTc-sestamibi SPECT/CT over.: Conventional SPECT in the evaluation of parathyroid adenomas or hyperplasia
    Gayed, IW
    Kim, EE
    Broussard, WF
    Evans, D
    Lee, J
    Broemeling, LD
    Ochoa, BB
    Moxley, DM
    Erwin, WD
    Podoloff, DA
    [J]. JOURNAL OF NUCLEAR MEDICINE, 2005, 46 (02) : 248 - 252
  • [37] Dual-Phase 99mTc Sestamibi Scintigraphy With Neck and Thorax SPECT/CT in Primary Hyperparathyroidism A Single-Institution Experience
    Ciappuccini, Renaud
    Morera, Julia
    Pascal, Pierre
    Rame, Jean-Pierre
    Heutte, Natacha
    Aide, Nicolas
    Babin, Emmanuel
    Reznik, Yves
    Bardet, Stephane
    [J]. CLINICAL NUCLEAR MEDICINE, 2012, 37 (03) : 223 - 228
  • [38] 99mTC-Sestamibi SPECT/CT Imaging for the Risk Stratification of Renal Masses
    Sistani, Golmehr
    Bjazevic, Jennifer
    Zahra, Kassam
    Stephen, Pautler
    Romsa, Jonathan
    [J]. JOURNAL OF NUCLEAR MEDICINE, 2020, 61
  • [39] Comparison of 99mTc-sestamibi/123I subtraction SPECT/CT and combined 99mTc-sestamibi/123I sub-traction SPECT/CT with planar pinhole image in hyperparathyroidism
    Tunninen, V.
    Varjo, P.
    Seppanen, M.
    [J]. EUROPEAN JOURNAL OF NUCLEAR MEDICINE AND MOLECULAR IMAGING, 2015, 42 : S65 - S66
  • [40] Quantitative 99mTc-Sestamibi SPECT/CT in patients with hyperparathyroidism
    Matsunaga, Keiko
    Kamiya, Takashi
    Fujino, Kouichi
    Watabe, Tadashi
    Isohashi, Kayako
    Tatsumi, Mitsuaki
    Shimosegawa, Eku
    Hatazawa, Jun
    [J]. JOURNAL OF NUCLEAR MEDICINE, 2017, 58