Evaluation of EBT-2 model GAFCHROMIC™ film performance in water

被引:29
|
作者
Aldelaijan, Saad [1 ,2 ]
Devic, Slobodan [1 ,2 ]
Mohammed, Huriyyah [1 ,2 ]
Tomic, Nada [1 ,2 ]
Liang, Li-Heng [1 ,2 ]
DeBlois, Francois [1 ,2 ]
Seuntjens, Jan [1 ]
机构
[1] McGill Univ, Med Phys Unit, Montreal, PQ H3G 1A4, Canada
[2] McGill Univ, SMBD Jewish Gen Hosp, Dept Radiat Oncol, Montreal, PQ H3T 1E2, Canada
关键词
radiochromic film dosimetry; water immersion; RADIOCHROMIC FILM; DOSIMETRY;
D O I
10.1118/1.3455713
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Purpose: The authors present results of the measurements on the impact of radiochromic film immersion in water. The impact of film piece size, initial optical density, postimmersion waiting time prior to scanning, and the time film was kept in water has been investigated. The authors also investigated the pathways of water penetration into the film during the film immersion in water. Methods: To study the impact of water immersion on change in optical density, the authors used various sizes of the latest EBT-2 model GAFCHROMIC (TM) film: 2 X 2, 4 X 4, and 8 X 8 in.(2). In addition, to test any existing dependence of the film's optical density on water diffusion, the authors used two sets of films: Unexposed (0 Gy) and film pieces exposed to a dose of 3 Gy. Times that film pieces were left in water ranged from 30 min to 24 h, and once the film was permanently removed from water, the authors also studied the impact of the scanning time (Delta t) that ranged from 0 (films scanned right after removal from water) to 72 h postimmersion. Results: While the penetration depth can reach as much as 9 mm around the edges of the EBT-2 GAFCHROMIC T film, the anticipated dose error due to the change in optical density due to the water immersion appears to be negligible for the short immersions of the order of 30 min. However, as the immersion time increases, the anticipated dose error may reach 22 cGy on a 2 X 2 in.(2) piece of film, which corresponds to 7% dose error at 3 Gy of measured dose. Conclusions: In this work, the authors report on an undoubted impact of radiochromic film immersion in water on the measured change in optical density, which may lead to systematic errors in dose measurements if the film is kept in water for longer periods of time. The magnitude of the impact depends on many parameters: Size of the film piece, initial optical density, postimmersion waiting time prior to scanning (defined by the current radiochromic film dosimetry protocol in place), and the time film was kept in water. The authors also suggested various approaches in correcting for the change in netOD due to water penetration into the film, but the authors believe that the use of the control film piece would be the most appropriate. (c) 2010 American Association of Physicists in Medicine. [DOI: 10.1118/1.3455713]
引用
收藏
页码:3687 / 3693
页数:7
相关论文
共 50 条
  • [1] Investigation of energy dependence of EBT and EBT-2 Gafchromic film
    Lindsay, Patricia
    Rink, Alexandra
    Ruschin, Mark
    Jaffray, David
    [J]. MEDICAL PHYSICS, 2010, 37 (02) : 571 - 576
  • [2] Absorption spectra time evolution of EBT-2 model GAFCHROMIC™ film
    Devic, Slobodan
    Aldelaijan, Saad
    Mohammed, Huriyyah
    Tomic, Nada
    Liang, Li-Heng
    DeBlois, Francois
    Seuntjens, Jan
    [J]. MEDICAL PHYSICS, 2010, 37 (05) : 2207 - 2214
  • [3] Dosimetric Verification of IMRT Fields Using GafChromic® EBT-2 Film
    McLawhorn, R.
    Ter-Antonyan, R.
    Benedict, S.
    Wijesooriya, K.
    [J]. MEDICAL PHYSICS, 2010, 37 (06)
  • [4] Reaction of three UV exposure to gafchromic EBT-2 and EBT-3
    Katsuda, T.
    Gotanda, T.
    Gotanda, R.
    Akagawa, T.
    Tanki, N.
    Kuwano, T.
    Yabunaka, K.
    [J]. WORLD CONGRESS ON MEDICAL PHYSICS AND BIOMEDICAL ENGINEERING, 2015, VOLS 1 AND 2, 2015, 51 : 759 - 762
  • [5] Impact of Post-Irradiation Time on the Accuracy of Dose Measurements for EBT-2 Model GAFCHROMIC(TM) Film
    Aldelaijan, S.
    Mohammed, H.
    Tomic, N.
    Seuntjens, J.
    DeBlois, F.
    Lning, L.
    Devic, S.
    [J]. MEDICAL PHYSICS, 2010, 37 (06)
  • [6] IMPACT OF WATER IMMERSION ON DOSE MEASUREMENT ACCURACY FOR THE EBT-2 MODEL RADIOCHROMIC FILM
    Aldelajjan, S.
    Mohhamed, H.
    Tomic, N.
    Seuntjens, J.
    DeBlois, F.
    Liang, L. H.
    Devic, S.
    [J]. RADIOTHERAPY AND ONCOLOGY, 2010, 96 : S25 - S25
  • [7] Evaluation of GAFCHROMIC® EBT film for CyberKnife® dosimetry
    Wilcox, Ellen E.
    Daskalov, George M.
    [J]. MEDICAL PHYSICS, 2007, 34 (06) : 1967 - 1974
  • [8] Absorption spectroscopy of EBT model GAFCHROMIC™ film
    Devic, Slobodan
    Tomic, Nada
    Pang, Zhiyu
    Seuntjens, Jan
    Podgorsak, Ervin B.
    Soares, Christopher G.
    [J]. MEDICAL PHYSICS, 2007, 34 (01) : 112 - 118
  • [9] Evaluation of the Gafchromic® EBT2 film for the dosimetry of radiosurgical beams
    Larraga-Gutierrez, Jose M.
    Garcia-Hernandez, Diana
    Garcia-Garduno, Olivia A.
    Galvan de la Cruz, Olga O.
    Ballesteros-Zebadua, Paola
    Esparza-Moreno, Karina P.
    [J]. MEDICAL PHYSICS, 2012, 39 (10) : 6111 - 6117
  • [10] Evaluation of gafchromic EBT film for prostate IMRT verification
    Aziz, T.
    Budgell, G. J.
    Mackay, R. I.
    [J]. CLINICAL ONCOLOGY, 2007, 19 (03) : S24 - S24