Improved time response for large area microchannel plate photomultiplier tubes in fusion diagnostics

被引:6
|
作者
Milnes, J. S. [1 ]
Horsfield, C. J. [2 ]
Conneely, T. M. [1 ]
Howorth, J. [1 ]
机构
[1] Photek Ltd, St Leonards On Sea TN38 9NS, E Sussex, England
[2] AWE, Reading RG7 4PR, Berks, England
来源
REVIEW OF SCIENTIFIC INSTRUMENTS | 2014年 / 85卷 / 11期
关键词
D O I
10.1063/1.4886759
中图分类号
TH7 [仪器、仪表];
学科分类号
0804 ; 080401 ; 081102 ;
摘要
Fusion diagnostics that utilise high speed scintillators often need to capture a large area of light with a high degree of time accuracy. Microchannel plate (MCP) photomultiplier tubes (PMTs) are recognised as the leading device for capturing fast optical signals. However, when manufactured in their traditional proximity focused construction, the time response performance is reduced as the active area increases. This is due to two main factors: the capacitance of a large anode and the difficulty of obtaining small pore MCPs with a large area. Collaboration between Photek and AWE has produced prototype devices that combine the excellent time response of small area MCP-PMTs with a large active area by replacing the traditional proximity-gap front section with an electro-optically focused photocathode to MCP. We present results from both single and double MCP devices with a 40 mm diameter active area and show simulations for the 100 mm device being built this year. (C) 2014 AIP Publishing LLC.
引用
下载
收藏
页数:3
相关论文
共 50 条
  • [1] PROXIMITY FOCUSED MICROCHANNEL PLATE PHOTOMULTIPLIER TUBES
    CECKOWSKI, DH
    EBERHARDT, E
    CARNEY, E
    IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 1981, 28 (01) : 677 - 682
  • [2] SHORT RECOVERY-TIME, MULTIANODE, MICROCHANNEL PLATE PHOTOMULTIPLIER FOR PLASMA DIAGNOSTICS
    GIUDICOTTI, L
    BASSAN, M
    SARDELLA, A
    PERDON, E
    ULTRAVIOLET TECHNOLOGY III, 1989, 1158 : 173 - 182
  • [3] Area and extraction field analysis of the analogue saturation of 40 mm microchannel plate photomultiplier tubes
    Milnes, J. S.
    Conneely, T. M.
    Horsfield, C. J.
    Lapington, J.
    REVIEW OF SCIENTIFIC INSTRUMENTS, 2018, 89 (10):
  • [4] Study on the stability of 20 inch photomultiplier tubes based on microchannel plate
    Li, H.
    Qian, S.
    Ning, Z.
    Zhang, Y.
    Wu, Q.
    Ma, L.
    Zhu, Y.
    Peng, S.
    Yan, M.
    JOURNAL OF INSTRUMENTATION, 2021, 16 (06)
  • [5] Contribution of microchannel plate luminescence to the noise of 20-inch photomultiplier tubes
    Ren, Ling
    Wang, Xingchao
    Huang, Guorui
    Si, Shuguang
    Jin, Muchun
    Sun, Jianning
    Qiu, Xiangbiao
    Gu, Yan
    Jin, Zhen
    Qian, Sen
    Liu, Shulin
    Wang, Qilong
    NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 2022, 1022
  • [6] Development of a magnetic shield for 20-inch microchannel plate photomultiplier tubes
    Chen, Mingjun
    Gao, Bo
    Guo, JingYuan
    He, Huihai
    Liu, Cheng
    Li, Huicai
    Li, Kai
    Ren, Ling
    Si, Shuguang
    Wu, Jiewei
    Wang, Wei
    You, Xiaohao
    NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 2022, 1039
  • [8] Analogue saturation limit of single and double 10 mm microchannel plate photomultiplier tubes
    Milnes, J. S.
    Conneely, T. M.
    Horsfield, C. J.
    REVIEW OF SCIENTIFIC INSTRUMENTS, 2016, 87 (11):
  • [9] Study on the improvement of the 20-inch microchannel plate photomultiplier tubes for neutrino detector
    Ren, Ling
    Sun, Jianning
    Si, Shuguang
    Wang, Xingchao
    Jin, Muchun
    Huang, Guorui
    Xu, Haiyang
    Wang, Congjie
    Hou, Wei
    Wang, Liang
    Jin, Zhen
    Qiao, Fangjian
    Qiu, Xiangbiao
    Shi, Mengyao
    Zhang, Haoda
    Cao, Yiqi
    Gu, Yan
    Li, Shen
    Han, Xiaoming
    Qian, Sen
    Liu, Shulin
    NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 2020, 977
  • [10] Ultra-high speed photomultiplier tubes with nanosecond gating for fusion diagnostics
    Milnes, J. S.
    Horsfield, C. J.
    Rubery, M. S.
    Glebov, V. Yu.
    Herrmann, H. W.
    REVIEW OF SCIENTIFIC INSTRUMENTS, 2012, 83 (10):