Eye lens dosimetry with BeOSL Hp(3) dosemeters in interventional angiography and neuroradiology

被引:0
|
作者
Strobel, I [2 ]
Bandalo, V [1 ]
Herz, K. [2 ]
Fehrenbacher, G. [2 ]
Grozinger, G. [2 ]
Artzner, C. [2 ]
Bronner, J. [1 ]
Kleinau, P. [1 ]
Hoedlmoser, H. [1 ]
机构
[1] Mirion Technol AWST GmbH, Munich, Germany
[2] Univ Hosp Tubingen, Tubingen, Germany
关键词
Eye lens dosimetry; Interventional radiology; OSL; BeO; Radiation protection glasses; BeOSL; RADIOLOGY; CALIBRATION; EXPOSURES;
D O I
10.1016/j.radmeas.2021.106570
中图分类号
TL [原子能技术]; O571 [原子核物理学];
学科分类号
0827 ; 082701 ;
摘要
In a first clinical application study, recently introduced BeOSL Hp(3) dosemeters were used for cumulative workplace characterization measurements in interventional radiology and interventional neuroradiology. Eye lens doses were accumulated on dosemeters assigned to interventions in specific anatomical areas and shared by multiple physicians. Doses were measured inside and outside of radiation protection glasses in order to compare shielded and unshielded doses near the left eye of the operator. Dose values per intervention were in the range of 0.014 mSv?0.097 mSv with the highest doses in interventions performed on the anatomical areas of the spine, abdomen, and pelvis. Additionally, Hp(10) doses were recorded at the collar of the thyroid protector, which showed a limited correlation with eye lens doses. Dose reduction factors obtained for two types of radiation protection glasses were between 2.5 and 6 in accordance with previous static measurements in the laboratory. While no systematic extrapolation to annual doses based on actual annual numbers of interventions for individual physicians has been performed yet, a simple time-based extrapolation of the recorded doses suggests that unshielded eye lens doses can approach the annual limit of 20 mSv and underline the importance of the use of radiation protection glasses and dose monitoring.
引用
收藏
页数:7
相关论文
共 50 条
  • [1] Eye lens equivalent dose in interventional neuroradiology based on OSL dosimetry
    Patino, G.
    Ballester, F.
    Vijande, J.
    Gras, P.
    Villaescusa, J.
    Campayo, J.
    [J]. 2016 GLOBAL MEDICAL ENGINEERING PHYSICS EXCHANGES/PAN AMERICAN HEALTH CARE EXCHANGES (GMEPE/PAHCE), 2016,
  • [2] Eye Lens Dosimetry in Interventional Radiology: Assessment With Dedicated Hp(3) Dosimeters
    Merrachi, Nassim-Albert
    Bouchard-Bellavance, Roxanne
    Perreault, Pierre
    Gilbert, Patrick
    Soulez, Gilles
    Bouchard, Louis
    Oliva, Vincent L.
    Giroux, Marie-France
    Normandeau, Lysanne
    Therasse, Eric
    [J]. CANADIAN ASSOCIATION OF RADIOLOGISTS JOURNAL-JOURNAL DE L ASSOCIATION CANADIENNE DES RADIOLOGISTES, 2021, 72 (02): : 317 - 323
  • [3] Test of ring, eye lens and whole body dosemeters for the dose quantity Hp(3) to be used in interventional radiology
    Szumska, A.
    Budzanowski, M.
    Kopec, R.
    [J]. RADIATION PHYSICS AND CHEMISTRY, 2017, 140 : 92 - 97
  • [4] On the operational quantity Hp(3) for eye lens dosimetry
    Behrens, R.
    [J]. JOURNAL OF RADIOLOGICAL PROTECTION, 2012, 32 (04) : 455 - 464
  • [5] Hp(0.07) PHOTON DOSEMETERS FOR EYE LENS DOSIMETRY: CALIBRATION ON A ROD Vs. A SLAB PHANTOM
    Behrens, R.
    Engelhardt, J.
    Figel, M.
    Hupe, O.
    Jordan, M.
    Seifert, R.
    [J]. RADIATION PROTECTION DOSIMETRY, 2012, 148 (02) : 139 - 142
  • [6] Eye lens monitoring for interventional radiology personnel: dosemeters, calibration and practical aspects of Hp(3) monitoring. A 2015 review
    Carinou, Eleftheria
    Ferrari, Paolo
    Bjelac, Olivera Ciraj
    Gingaume, Merce
    Merce, Marta Sans
    O'Connor, Una
    [J]. JOURNAL OF RADIOLOGICAL PROTECTION, 2015, 35 (03) : R17 - R34
  • [7] Hp(3) vs TLD-100 for eye lens dosimetry in interventional radiology procedures: a preliminary study
    Manna, Francesco
    De Nardellis, Gianleonardo
    Carmosino, Patrizio Antonio
    Ambrosino, Fabrizio
    Caruso, Umberto
    Correra, Marco
    Fiore, Francesco
    La Verde, Giuseppe
    Tarotto, Luca
    Pugliese, Mariagabriella
    [J]. EUROPEAN PHYSICAL JOURNAL PLUS, 2023, 138 (09):
  • [8] Hp(3) vs TLD-100 for eye lens dosimetry in interventional radiology procedures: a preliminary study
    Francesco Manna
    Gianleonardo De Nardellis
    Patrizio Antonio Carmosino
    Fabrizio Ambrosino
    Umberto Caruso
    Marco Correra
    Francesco Fiore
    Giuseppe La Verde
    Luca Tarotto
    Mariagabriella Pugliese
    [J]. The European Physical Journal Plus, 138
  • [9] INFLUENCE OF DIFFERENT TYPES OF PHANTOMS ON THE CALIBRATION OF DOSEMETERS FOR EYE LENS DOSIMETRY
    Yoshitomi, H.
    Kowatari, M.
    [J]. RADIATION PROTECTION DOSIMETRY, 2016, 170 (1-4) : 199 - 203
  • [10] ENERGY AND ANGULAR DEPENDENCE OF RADIOPHOTOLUMINESCENT GLASS DOSEMETERS FOR EYE LENS DOSIMETRY
    Silva, E. H.
    Knezevic, Z.
    Struelens, L.
    Covens, P.
    Ueno, S.
    Vanhavere, F.
    Buls, N.
    [J]. RADIATION PROTECTION DOSIMETRY, 2016, 170 (1-4) : 208 - 212