A novel low-profile external skeletal fixator for type IIIB open tibial fractures: A biomechanical and clinical pilot study

被引:1
|
作者
Tarng, Yih-Wen [1 ,2 ]
Lin, Kai-Cheng [1 ]
Lin, Kun-Jhih [3 ,4 ]
Yang, Yi-Ping [5 ]
Chien, Yeuh [5 ]
Wei, Hung-Wen [3 ,4 ]
机构
[1] Kaohsiung Vet Gen Hosp, Dept Orthopaed, 386 Da Chung 1st Rd, Kaohsiung 813, Taiwan
[2] Natl Def Med Ctr, Dept Orthopaed, Taipei, Taiwan
[3] Chung Yuan Christian Univ, Dept Elect Engn, Taoyuan, Taiwan
[4] Chung Yuan Christian Univ, Technol Translat Ctr Med Device, Taoyuan, Taiwan
[5] Taipei Vet Gen Hosp, Dept Med Res, Taipei, Taiwan
关键词
Bone union; Gustilo open type IIIb; Low-profile external skeletal fixator; Tibial fractures; LOCKING COMPRESSION PLATE; EXPERIENCE; MANAGEMENT; TEMPORARY; INJURIES; CLOSURE;
D O I
10.1097/JCMA.0000000000000506
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Background: Although external fixator is standard for managing staged treatment of open tibial fracture, the main disadvantage of this device is too bulky to be tolerated by most patients for longtime use. The purposes of this pilot study were to compare the biomechanical properties of a novel low-profile external fixator (LP-ESF) with a traditional ESF and also to evaluate its performance in patients with Gustilo type IIIb tibial open fractures. Methods: A prospective clinical pilot study started from January 2015 to December 2017, and 18 patients with Gustilo type IIIb open tibial fractures underwent the fixation with a novel LP-ESF system. The biomechanical properties of the LP-ESF were compared with the Synthes External Fixation System according to the standard ASTM F1541-02. These patients were divided into two groups according to the size of bony defect. The postoperative clinical outcomes were subsequently collected. Results: The biomechanical properties of the LP-ESF were comparable with those of Synthes External Fixation System and had an improved the axial/torsional stiffness and ultimate strength. In the clinical study, all patients with LP-ESF had fracture union. The duration of application of LP-ESF was 3.5 to 18 months until fracture union. In 10 of 18 patients, their fractures were immobilized with the LP-ESF until bone union, and no pin tract infection and no chronic osteomyelitis were recorded. The 36-Item Short Form Health Survey life quality and health survey were good to excellent in these patients. Notably, the LP-ESF allowed a patient with severe bone and soft-tissue defects to preserve the leg and joints function. Conclusion: In this study, we found that the novel LP-ESFs had improved clinical outcomes. The long-term LP-ESF application seems to be tolerable in our patients. This novel approach permits better controls in deep infection and faster healing of fractures, and thus may provide a viable alternative treatment for Gustilo type IIIb open tibial fractures.
引用
收藏
页码:528 / 535
页数:8
相关论文
共 48 条
  • [1] Investigating the biomechanical function of the plate-type external fixator in the treatment of tibial fractures: a biomechanical study
    Shi, Di
    Liu, Kaiyuan
    Zhang, Haomeng
    Wang, Xinli
    Li, Guochen
    Zheng, Lianhe
    BMC MUSCULOSKELETAL DISORDERS, 2020, 21 (01)
  • [2] Investigating the biomechanical function of the plate-type external fixator in the treatment of tibial fractures: a biomechanical study
    Di Shi
    Kaiyuan Liu
    Haomeng Zhang
    Xinli Wang
    Guochen Li
    Lianhe Zheng
    BMC Musculoskeletal Disorders, 21
  • [3] The treatment of open tibial fractures and of tibial non-union with a novel external fixator
    Ong, CT
    Choon, DSK
    Cabrera, NP
    Maffulli, N
    INJURY-INTERNATIONAL JOURNAL OF THE CARE OF THE INJURED, 2002, 33 (09): : 829 - 834
  • [4] Metaphyseal locking plate as an external fixator for open tibial fracture: Clinical outcomes and biomechanical assessment
    Ma, Ching-Hou
    Wu, Chin-Hsien
    Jiang, Jiun-Ru
    Tu, Yuan-Kun
    Lin, Ting-Sheng
    INJURY-INTERNATIONAL JOURNAL OF THE CARE OF THE INJURED, 2017, 48 (02): : 501 - 505
  • [5] Biomechanical study of the stiffness of the femoral locking compression plate of an external fixator for lower tibial fractures
    Huan Su
    Siyang Zhong
    Tianyong Ma
    Weidong Wu
    Yihong Lu
    Dewei Wang
    BMC Musculoskeletal Disorders, 24
  • [6] Biomechanical study of the stiffness of the femoral locking compression plate of an external fixator for lower tibial fractures
    Su, Huan
    Zhong, Siyang
    Ma, Tianyong
    Wu, Weidong
    Lu, Yihong
    Wang, Dewei
    BMC MUSCULOSKELETAL DISORDERS, 2023, 24 (01)
  • [7] Treatment of type IIIA open fractures of tibial shaft with Ilizarov external fixator versus unreamed tibial nailing
    Inan, Muharrem
    Halici, Mehmet
    Ayan, Irfan
    Tuncel, Mehmet
    Karaoglu, Sinan
    ARCHIVES OF ORTHOPAEDIC AND TRAUMA SURGERY, 2007, 127 (08) : 617 - 623
  • [8] Treatment of type IIIA open fractures of tibial shaft with Ilizarov external fixator versus unreamed tibial nailing
    Muharrem Inan
    Mehmet Halici
    Irfan Ayan
    Mehmet Tuncel
    Sinan Karaoglu
    Archives of Orthopaedic and Trauma Surgery, 2007, 127 : 617 - 623
  • [9] INTERFRAGMENTARY LAG SCREWS AND EXTERNAL FIXATION FOR SEVERE OPEN TIBIAL FRACTURES - A BIOMECHANICAL AND CLINICAL-STUDY
    ZIV, I
    ROSEN, E
    ELYASHUV, O
    SEGAL, D
    SIMKIN, A
    CLINICAL BIOMECHANICS, 1994, 9 (03) : 167 - 173
  • [10] Role of early Ilizarov ring fixator in the definitive management of type II, IIIA and IIIB open tibial shaft fractures
    Wani, Naveed
    Baba, Asif
    Kangoo, Khurshid
    Mir, Mohammad
    INTERNATIONAL ORTHOPAEDICS, 2011, 35 (06) : 915 - 923