Development of a transportable apparatus for vacuum gauge calibration

被引:1
|
作者
Xi, Zhenhua [1 ]
Li, Detian [1 ]
Cheng, Yongjun [1 ]
Wang, Yongjun [1 ]
Zhang, Huzhong [1 ]
Liu, Jiatong [1 ]
Guan, Baoguo [1 ]
Ji, Kang [1 ]
机构
[1] Lanzhou Inst Phys, Lanzhou 730000, Peoples R China
基金
中国国家自然科学基金;
关键词
Transportable apparatus; Pressure attenuation method; Direct comparison; Metrological characteristic; PRESSURE; STANDARD; RANGE; PA;
D O I
10.1016/j.vacuum.2016.01.025
中图分类号
T [工业技术];
学科分类号
08 ;
摘要
A new ultrahigh vacuum transportable apparatus for vacuum gauge calibration in the pressure range from 10(-5) Pa-10(5) Pa is developed. On the basis of modularization design, it can be easily extended and assembled. A standard pressure p(std) with various gas species can be generated by means of pressure attenuation module which is easy to introduce the corresponding calibration gas through an orifice. The standard pressure p(std) in the range from 10(-5) Pa to 10(-2) Pa could be obtained by changing the upstream pressure from 1 Pa to 10(3) Pa with an uncertainty of approximately 2.4%. At the same time, calibration can also be performed by direct comparison with a reference gauge either in static or stationary equilibrium. The ultimate pressure could be in the level of 10(-6) Pa typically. The pumpdown characteristic and outgassing rate of the apparatus after exposing to atmospheric pressure are presented. The main metrological characteristics of the apparatus are given. The size and weight of this apparatus are 475 mm x 420 mm x 800 mm and 37.8 kg, respectively. Its calibration range is from 1.96 x 10(-5) Pa to 1.03 x 10(5) Pa with the relative combined standard uncertainty from 2.8% to 0.40%. (C) 2016 Elsevier Ltd. All rights reserved.
引用
收藏
页码:30 / 37
页数:8
相关论文
共 50 条
  • [11] CALIBRATION OF A KNUDSEN TYPE VACUUM GAUGE
    LASSETTRE, EN
    GLASER, FM
    REVIEW OF SCIENTIFIC INSTRUMENTS, 1965, 36 (11): : 1658 - +
  • [12] VACUUM GAUGE CALIBRATION BY THE STATIC METHOD
    BERMAN, A
    VACUUM, 1979, 29 (11-1) : 417 - 425
  • [13] A NEW METHOD FOR VACUUM GAUGE CALIBRATION
    KUZMIN, VV
    MEASUREMENT TECHNIQUES USSR, 1990, 33 (11): : 1156 - 1158
  • [14] Calibration of a pirani vacuum gauge by using a deadweight piston gauge
    Woo, Sam-Yong
    Song, Han-Wook
    Choi, In-Mook
    JOURNAL OF THE KOREAN PHYSICAL SOCIETY, 2007, 51 (03) : 916 - 919
  • [15] DEVELOPMENT OF A TRANSPORTABLE, EXPERIMENTAL, LIVER-PRESERVATION APPARATUS
    HINCHLIFFE, A
    IMMELMAN, EJ
    WHITE, HJO
    JOHNSON, MG
    BOWES, JB
    GOLBY, MG
    PEACOCK, JH
    RIDDELL, AG
    BRITISH JOURNAL OF SURGERY, 1970, 57 (11) : 853 - +
  • [16] A REVIEW OF VACUUM GAUGES AND METHODS FOR HIGH VACUUM GAUGE CALIBRATION
    SELLENGER, FR
    VACUUM, 1968, 18 (12) : 645 - +
  • [17] REFERENCE TRANSFER METHOD OF VACUUM GAUGE CALIBRATION
    MORRISON, CF
    JOURNAL OF VACUUM SCIENCE & TECHNOLOGY, 1966, 3 (05): : 303 - &
  • [18] THE SERIES EXPANSION METHOD FOR VACUUM GAUGE CALIBRATION
    REDGRAVE, FJ
    VACUUM, 1988, 38 (8-10) : 951 - 951
  • [19] REFERENCE TRANSFER METHOD OF VACUUM GAUGE CALIBRATION
    MORRISON, CF
    VACUUM, 1967, 17 (03) : 164 - &
  • [20] Calibration apparatus for precise barometers and vacuum gauges
    Woo, SY
    Choi, IM
    Lee, YJ
    Hong, SS
    Chung, KH
    JOURNAL OF VACUUM SCIENCE & TECHNOLOGY A, 2004, 22 (04): : 1816 - 1819