In Situ Mg/Ca Measurements on Foraminifera: Comparison Between Laser Ablation Inductively Coupled Plasma Mass Spectrometry and Wavelength-Dispersive X-Ray Spectroscopy by Electron Probe Microanalyzer
被引:6
|作者:
Balestra, B.
[1
,2
,3
]
Rose, T.
[4
]
Fehrenbacher, J.
[5
]
Knobelspiesse, K. D.
[6
]
Huber, B. T.
[2
]
Gooding, T.
[4
]
Paytan, A.
[3
]
机构:
[1] Amer Univ, Environm Sci Dept, Washington, DC 20016 USA
[2] Natl Museum Nat Hist, Dept Paleobiol, MRC 121, Washington, DC 20560 USA
[3] Univ Calif Santa Cruz, Inst Marine Sci, Santa Cruz, CA 95064 USA
[4] Natl Museum Nat Hist, Dept Mineral Sci, Washington, DC 20560 USA
[5] Oregon State Univ, Coll Earth Ocean & Atmospher Sci, Corvallis, OR 97331 USA
[6] NASA, Goddard Space Flight Ctr, Ocean Ecol Lab, Greenbelt, MD USA
关键词:
EPMA;
foraminifera;
LA‐
ICP‐
MS;
Mg;
Ca;
paired analysis;
NEOGLOBOQUADRINA-DUTERTREI;
PLANKTONIC-FORAMINIFERA;
BENTHIC FORAMINIFERA;
TRACE-ELEMENTS;
VARIABILITY;
CALCITE;
TEMPERATURE;
DELTA-O-18;
DELTA-C-13;
CHEMISTRY;
D O I:
10.1029/2020GC009449
中图分类号:
P3 [地球物理学];
P59 [地球化学];
学科分类号:
0708 ;
070902 ;
摘要:
We present a comparison of two different techniques: Laser Ablation Inductively Coupled Plasma Mass Spectrometry (LA-ICP-MS) and wavelength-dispersive X-Ray Spectroscopy by electron probe microanalyzer (EPMA) for obtaining Mg/Ca ratios in individual foraminifera shells. The goal is to assess the use of EPMA as an alternative technique for Mg/Ca analyses of single foraminiferal calcite shells. Foraminifera obtained from sediments (benthic, Uvigerina spp.) and from plankton tows (planktonic, Orbulina universa) were analyzed. All specimens were prepared in epoxy mounts and exposed in cross-section such that multiple high-resolution analyses could be completed on the shells using both techniques. We examined our data using statistical methods designed for the assessment and comparison of measurement techniques. In the case of Uvigerina, the mean difference for ratios obtained using EPMA and LA-ICP-MS is very small (-0.046 mmol mol(-1)) and scale independent. The Limits of Agreement (LoA, the standard deviation of the bias plus the mean bias) is [-0.315, 0.223] mmol mol(-1). For samples with ratios lower than 13 mmol mol(-1), we found a mean EPMA-LA-ICP-MS bias of -2.44 mmol mol(-1) and a corresponding LoA of [-3.85, -1.04] mmol mol(-1). For ratios higher than 13 mmol mol(-1), there appears to be a scale dependent bias, meaning that the EPMA measured ratios become progressively larger than those of LA-ICP-MS as the Mg/Ca ratio increases, so the mean bias and LoA metrics are not meaningful. Results indicate that it is possible to use EPMA to collect Mg/Ca data, if the ratios are lower than similar to 13 mmol mol(-1).