Tibial component fixation in porous- and hydroxyapatite-coated total knee arthroplasty -: A radiostereometric evaluation of migration and inducible displacement after 5 years

被引:53
|
作者
Regnér, L [1 ]
Carlsson, L [1 ]
Kärrholm, J [1 ]
Herberts, P [1 ]
机构
[1] Gothenburg Univ, Sahlgrens Hosp, Dept Orthopaed, Inst Surg Sci, S-41345 Gothenburg, Sweden
来源
JOURNAL OF ARTHROPLASTY | 2000年 / 15卷 / 06期
关键词
knee prosthesis; cementless; hydroxyapatite; radiostereometry; inducible displacement;
D O I
10.1054/arth.2000.8103
中图分类号
R826.8 [整形外科学]; R782.2 [口腔颌面部整形外科学]; R726.2 [小儿整形外科学]; R62 [整形外科学(修复外科学)];
学科分类号
摘要
A total of 51 knees (45 patients) with osteoarthrosis were stratified in a randomized study to receive a hydroxyapatite-coated Freeman-Samuelson (FS HA) or a porous-coated Miller-Galante II (MG II) uncemented total knee arthroplasty. Repeated clinical, radiographic, and radiostereometric analysis (RSA) evaluations of the tibial components were done for 5 years. The clinical outcome was equal at the 5-year follow-up (mean Hospital for Special Surgery score, 93). Standard radiographs displayed more zones around the tibial stem in the MG II group. RSA revealed that the FS HA components migrated less (smaller maximum total point motion and maximum subsidence) after 5 years and showed less inducible displacements at the 1-year follow-up. The stability of the implants obtained is equal to or better than cemented implants after 5 years.
引用
收藏
页码:681 / 689
页数:9
相关论文
共 44 条
  • [31] Long lasting outcome of hydroxyapatite-coated implants in primary knee arthroplasty: a continuous series of two hundred and seventy total knee arthroplasties at fifteen to twenty two years of clinical follow-up
    Jean-Alain Epinette
    International Orthopaedics, 2014, 38 : 305 - 311
  • [32] Long lasting outcome of hydroxyapatite-coated implants in primary knee arthroplasty: a continuous series of two hundred and seventy total knee arthroplasties at fifteen to twenty two years of clinical follow-up
    Epinette, Jean-Alain
    INTERNATIONAL ORTHOPAEDICS, 2014, 38 (02) : 305 - 311
  • [33] Fixation of a hydroxyapatite-tricalcium phosphate-coated cementless knee prosthesis - Clinical and radiographic evaluation seven years after surgery
    Akizuki, S
    Takizawa, T
    Horiuchi, H
    JOURNAL OF BONE AND JOINT SURGERY-BRITISH VOLUME, 2003, 85B (08): : 1123 - 1127
  • [34] The Influence of Postoperative Coronal Alignment on Tibial Migration After Total Knee Arthroplasty in Preoperative Varus and Valgus Knees A Secondary Analysis of 10 Randomized Controlled Trials Using Radiostereometric Analysis
    Hasan, Shaho
    Kaptein, Bart L.
    Nelissen, Rob G. H. H.
    van Hamersveld, Koen T.
    Toksvig-Larsen, Soren
    Marang-van de Mheen, Perla J.
    JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME, 2021, 103 (24): : 2281 - 2290
  • [35] Tibial Component Undersizing Is Related to High Degrees of Implant Migration Following Cementless Total Knee Arthroplasty A Study of Radiostereometric Analysis Data for 111 Patients with 2-Year Follow-up
    Andersen, Mikkel Rathsach
    Winther, Nikolaj
    Lind, Thomas
    Schroder, Henrik
    Flivik, Gunnar
    Petersen, Michael Mork
    JBJS OPEN ACCESS, 2023, 8 (03)
  • [36] Influence of marker-selection method in radiostereometric analysis of total knee arthroplasty on tibial baseplate migration patterns: a secondary analysis of a randomized controlled trial with 5-year follow-up
    van der Lelij, Thies J. N.
    Koster, Lennard A.
    van de Mheen, Perla J.
    Toksvig-larsen, Soeren
    Nelissen, Rob G. H. H.
    Kaptein, Bart L.
    ACTA ORTHOPAEDICA, 2024, 95 : 157 - 165
  • [37] Porous Tantalum Metaphyseal Cones for Severe Femoral and Tibial Bone Defects in Revision Total Knee Arthroplasty are Reliable for Fixation at Mean 5-Year Follow-Up
    Kayani, Babar
    Howard, Lisa C.
    Neufeld, Michael E.
    Greidanus, Nelson V.
    Masri, Bassam A.
    Garbuz, Donald S.
    JOURNAL OF ARTHROPLASTY, 2024, 39 (09): : S374 - S379
  • [38] Survival rate of cemented versus cementless tibial component in primary total knee arthroplasty over 5 years of follow-up: comparative study of 109 prostheses
    Pacoret, Victor
    Kalk, Etienne
    Labattut, Ludovic
    Girardot, Guillaume
    Baulot, Emmanuel
    Martz, Pierre
    SICOT-J, 2020, 6
  • [39] Evaluation of different coatings of the tibial tray in uncemented total knee arthroplasty. A randomized controlled trial with 5 years follow-up with RSA and DEXA
    Dyreborg, Karen
    Winther, Nikolaj
    Lind, Thomas
    Flivik, Gunnar
    Petersen, Michael Mork
    KNEE, 2021, 29 : 208 - 215
  • [40] Late stabilization after initial migration in patients undergoing cemented total knee arthroplasty: a 5-year follow-up of 2 randomized controlled trials using radiostereometric analysis
    Hasan, Shaho
    Kaptein, Bart L.
    Marang-van de Mheen, Perla J.
    van Hamersveld, Koen T.
    Nelissen, Rob G. H. H.
    Toksvig-Larsen, Soren
    ACTA ORTHOPAEDICA, 2022, 93 : 271 - 276