High-resolution handheld rigid endomicroscope based on full-field optical coherence tomography

被引:15
|
作者
la Guillaume, Emilie Benoit A. [1 ]
Martins, Franck [1 ]
Boccara, Claude [1 ,2 ]
Harms, Fabrice [1 ]
机构
[1] LLTech, Pepiniere Paris Sante Cochin, 29 Rue Faubourg St Jacques, F-75014 Paris, France
[2] PSL Res Univ, ESPCI ParisTech, UMR CNRS 7587, INSERM U979,Inst Langevin, 1 Rue Jussieu, F-75005 Paris, France
关键词
endoscopy; endomicroscopy; full-field optical coherence tomography; tandem interferometry; medical and biomedical imaging; optical biopsy; VOLUMETRIC LASER ENDOMICROSCOPY; IN-VIVO; MICROSCOPY; PROBE;
D O I
10.1117/1.JBO.21.2.026005
中图分类号
Q5 [生物化学];
学科分类号
071010 ; 081704 ;
摘要
Full-field optical coherence tomography (FF-OCT) is a powerful tool for nondestructive assessment of biological tissue, i.e., for the structural examination of tissue in depth at a cellular resolution. Mostly known as a microscopy device for ex vivo analysis, FF-OCT has also been adapted to endoscopy setups since it shows good potential for in situ cancer diagnosis and biopsy guidance. Nevertheless, all the attempts to perform endoscopic FF-OCT imaging did not go beyond lab setups. We describe here, to the best of our knowledge, the first handheld FF-OCT endoscope based on a tandem interferometry assembly using incoherent illumination. A common-path passive imaging interferometer at the tip of an optical probe makes it robust and insensitive to environmental perturbations, and a low finesse Fabry-Perot processing interferometer guarantees a compact system. A good resolution (2.7 mu m transverse and 6 mu m axial) is maintained through the long distance, small diameter relay optics of the probe, and a good signal-to-noise ratio is achieved in a limited 100 ms acquisition time. High-resolution images and a movie of a rat brain slice have been recorded by moving the contact endoscope over the surface of the sample, allowing for tissue microscopic exploration at 20 mu m under the surface. These promising ex vivo results open new perspectives for in vivo imaging of biological tissue, in particular, in the field of cancer and surgical margin assessment. (C) 2016 Society of Photo-Optical Instrumentation Engineers (SPIE)
引用
收藏
页数:6
相关论文
共 50 条
  • [1] High-resolution full-field optical coherence tomography with a Linnik microscope
    Dubois, A
    Vabre, L
    Boccara, AC
    Beaurepaire, E
    [J]. APPLIED OPTICS, 2002, 41 (04) : 805 - 812
  • [2] High-resolution imaging of biological tissue with full-field optical coherence tomography
    Zhu, Yue
    Gao, Wanrong
    [J]. THREE-DIMENSIONAL AND MULTIDIMENSIONAL MICROSCOPY: IMAGE ACQUISITION AND PROCESSING XXII, 2015, 9330
  • [3] Rapid and high-resolution imaging of human liver specimens by full-field optical coherence tomography
    Zhu, Yue
    Gao, Wanrong
    Zhou, Yuan
    Guo, Yingcheng
    Guo, Feng
    He, Yong
    [J]. JOURNAL OF BIOMEDICAL OPTICS, 2015, 20 (11)
  • [4] Ultrahigh-resolution full-field optical coherence tomography
    Dubois, A
    Grieve, K
    Moneron, G
    Lecaque, R
    Vabre, L
    Boccara, C
    [J]. APPLIED OPTICS, 2004, 43 (14) : 2874 - 2883
  • [5] High-resolution full-field spatial coherence gated optical tomography using monochromatic light source
    Srivastava, Vishal
    Nandy, Sreyankar
    Mehta, Dalip Singh
    [J]. APPLIED PHYSICS LETTERS, 2013, 103 (10)
  • [6] Effects of temporal and spatial coherence on resolution in full-field optical coherence tomography
    Gao, Wanrong
    [J]. JOURNAL OF MODERN OPTICS, 2015, 62 (21) : 1764 - 1774
  • [7] Full-Field Optical Coherence Tomography: A New Technology for 3D High-Resolution Skin Imaging
    Dalimier, Eugenie
    Salomon, Denis
    [J]. DERMATOLOGY, 2012, 224 (01) : 84 - 92
  • [8] Stroboscopic ultrahigh-resolution full-field optical coherence tomography
    Moneron, G
    Boccara, AC
    Dubois, A
    [J]. OPTICS LETTERS, 2005, 30 (11) : 1351 - 1353
  • [9] Full-field optical coherence tomography system with controllable longitudinal resolution
    Nitta, Kouichi
    Li, Tingyu
    Motoyama, Toshiki
    Matoba, Osamu
    Yoshimura, Takeaki
    [J]. JAPANESE JOURNAL OF APPLIED PHYSICS PART 1-REGULAR PAPERS BRIEF COMMUNICATIONS & REVIEW PAPERS, 2006, 45 (11): : 8897 - 8903
  • [10] Full-field optical coherence tomography system with controllable longitudinal resolution
    Nitta, Kouichi
    Li, Tingyu
    Motoyama, Toshiki
    Matoba, Osamu
    Yoshimura, Takeaki
    [J]. Japanese Journal of Applied Physics, Part 1: Regular Papers and Short Notes and Review Papers, 2006, 45 (11): : 8897 - 8903