An ion-implanted InP receiver for polarization resolved terahertz spectroscopy

被引:46
|
作者
Castro-Camus, E.
Lloyd-Hughes, J.
Fu, L.
Tan, H. H.
Jagadish, C.
Johnston, M. B.
机构
[1] Univ Oxford, Dept Phys, Clarendon Lab, Oxford OX1 3PU, England
[2] Australian Natl Univ, Dept Elect Mat Engn, Res Sch Phys Sci & Engn, Inst Adv Studies, Canberra, ACT 0200, Australia
基金
英国工程与自然科学研究理事会;
关键词
D O I
10.1364/OE.15.007047
中图分类号
O43 [光学];
学科分类号
070207 ; 0803 ;
摘要
We report on the construction, optical alignment and performance of a receiver which is capable of recording the full polarization state of coherent terahertz radiation. The photoconductive detector was fabricated on InP which had been implanted with Fe+ ions. The device operated successfully when it was gated with near infrared femtosecond pulses from either a Ti: sapphire laser oscillator or a 1 kHz regenerative laser amplifier. When illuminated with terahertz radiation from a typical photoconductive source, the optimized device had a signal to noise figure of 100:1 with a usable spectral bandwidth of up to 4 THz. The device was shown to be very sensitive to terahertz polarization, being able to resolve changes in polarization of 0.34 degrees. Additionally, we have demonstrated the usefulness of this device for (i) polarization sensitive terahertz spectroscopy, by measuring the birefringence of quartz and (ii) terahertz emission experiments, by measuring the polarization dependence of radiation generated by optical rectification in (110)-ZnTe. (C) 2007 Optical Society of America.
引用
收藏
页码:7047 / 7057
页数:11
相关论文
共 50 条
  • [1] TIME-RESOLVED SPECTROSCOPY OF ION-IMPLANTED ZNO
    DANIELS, CR
    PIERCE, BJ
    HENGEHOLD, RL
    BULLETIN OF THE AMERICAN PHYSICAL SOCIETY, 1975, 20 (03): : 376 - 377
  • [2] RAMAN-SPECTROSCOPY FOR CHARACTERIZATION OF ANNEALING OF ION-IMPLANTED INP
    MYERS, DR
    GOURLEY, PL
    VAIDYANATHAN, KV
    DUNLAP, HL
    INSTITUTE OF PHYSICS CONFERENCE SERIES, 1983, (65): : 635 - 640
  • [3] DEFECT RECOVERY OF ION-IMPLANTED INP
    WEYER, G
    NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION B-BEAM INTERACTIONS WITH MATERIALS AND ATOMS, 1995, 96 (1-2): : 315 - 318
  • [4] ELECTROCHEMICAL PROFILING OF ION-IMPLANTED INP
    BAHIR, G
    MERZ, JL
    ABELSON, JR
    SIGMON, TW
    JOURNAL OF THE ELECTROCHEMICAL SOCIETY, 1987, 134 (09) : 2187 - 2193
  • [5] Mechanisms for the activation of ion-implanted Fe in InP
    Cesca, T.
    Verna, A.
    Mattei, G.
    Gasparotto, A.
    Fraboni, B.
    Impellizzeri, G.
    Priolo, F.
    JOURNAL OF APPLIED PHYSICS, 2006, 100 (02)
  • [6] FULLY ION-IMPLANTED INP JUNCTION FETS
    BOOS, JB
    DIETRICH, HB
    WENG, TH
    SLEGER, KJ
    BINARI, SC
    HENRY, RL
    ELECTRON DEVICE LETTERS, 1982, 3 (09): : 256 - 258
  • [7] Ion-implanted InP for ultrafast photodetector applications
    Carmody, C
    Boudinov, H
    Tan, HH
    Jagadish, C
    Dao, LV
    Gal, M
    COMMAD 2000 PROCEEDINGS, 2000, : 153 - 156
  • [8] ELECTROCHEMICAL PROFILING OF ION-IMPLANTED InP.
    Bahir, G.
    Merz, J.L.
    Abelson, J.R.
    Sigmon, T.W.
    1600, (l34):
  • [9] ANNEALING BEHAVIOR OF ION-IMPLANTED FE IN INP
    SCHWARZ, SA
    SCHWARTZ, B
    SHENG, TT
    SINGH, S
    TELL, B
    JOURNAL OF APPLIED PHYSICS, 1985, 58 (04) : 1698 - 1700
  • [10] RADIOGENIC SN DEFECTS IN ION-IMPLANTED INP
    DAMGAARD, S
    PETERSEN, JW
    WEYER, G
    JOURNAL OF PHYSICS C-SOLID STATE PHYSICS, 1981, 14 (07): : 993 - 1000