Articular damage caused by metal plugs in a rabbit model for treatment of localized cartilage defects

被引:42
|
作者
Clusters, R. J. H. [1 ]
Dhert, W. J. A. [1 ]
van Rijen, M. H. P. [1 ]
Verbout, A. J. [1 ]
Creemers, L. B. [1 ]
Saris, D. B. F. [1 ]
机构
[1] Univ Utrecht, Med Ctr, Dept Orthopaed, NL-3508 TC Utrecht, Netherlands
关键词
cartilage defect; cartilage repair; knee; oxidized zirconium; cobalt-chromium; rabbit; small metal implant; Mankin score; osseointegration;
D O I
10.1016/j.joca.2007.02.007
中图分类号
R826.8 [整形外科学]; R782.2 [口腔颌面部整形外科学]; R726.2 [小儿整形外科学]; R62 [整形外科学(修复外科学)];
学科分类号
摘要
Objective: Currently, the surgical treatment of localized cartilage defects has limitations. Alternatively, localized cartilage defects may be treated with small biocompatible metal cartilage tacks. Our purpose was to investigate the applicability of defect-size femoral implants. Different bearing materials, cobalt-chromium (CoCr) and oxidized zirconium (OxZr), were tested to evaluate the effect on opposing cartilage quality and osseointegration at different insertion depths. Methods: In 18 adult female New Zealand White rabbits, a medial femoral condyle defect was filled with either an OxZr or a CoCr implant (0 articulating surface 3.5 mm; fixating pin of 9.1 mm length), placed flush, 1 mm deep or 1 mm protruding with respect to the level of the surrounding cartilage. Animals were sacrificed after 4 weeks. Tibial cartilage quality was scored microscopically and osseointegration measured by automated histomorphometry. Results: Considerable articulating cartilage erosion was found in all conditions. Tibial cartilage quality was least compromised when both implants were placed flush compared to deep (P=0.01) or protruding position (P=0.004) and was better for OxZr compared to CoCr (P=0.011) when left protruding, while no differences were found when placed deep of flush. Most bone formation around the fixating pin was observed in a protruding position (P=0.01). In deep position, more bone-implant contact was observed with CoCr compared to OxZr (P=0.02). Conclusions: OxZr and CoCr implants showed good osseointegration when used as a localized cartilage defect treatment in the rabbit knee; however, opposite cartilage damage was observed in all cases. Placement flush to the surrounding cartilage seems essential and when left protruding OxZr may be less erosive. In conclusion, caution is warranted using small metal implants for the treatment of localized cartilage in the human patient. (c) 2007 Osteoarthritis Research Society International. Published by Elsevier Ltd. All rights reserved.
引用
收藏
页码:937 / 945
页数:9
相关论文
共 50 条
  • [31] The treatment of articular cartilage defects using the microfracture technique
    Gill, Thomas J.
    Asnis, Peter D.
    Berkson, Eric M.
    JOURNAL OF ORTHOPAEDIC & SPORTS PHYSICAL THERAPY, 2006, 36 (10): : 728 - 738
  • [32] Focal articular cartilage defects of the knee: Surgical treatment
    Padua R.
    Bondì R.
    Journal of Orthopaedics and Traumatology, 2004, 5 (1) : 63 - 65
  • [33] Results of microfracture in the treatment of articular cartilage defects of the talus
    Becher, C
    Thermann, H
    FOOT & ANKLE INTERNATIONAL, 2005, 26 (08) : 583 - 589
  • [34] Treatment of Articular Cartilage Defects: Focus on Tissue Engineering
    Zylinska, Beata
    Silmanowicz, Piotr
    Sobczynska-Rak, Aleksandra
    Jarosz, Lukasz
    Szponder, Tomasz
    IN VIVO, 2018, 32 (06): : 1289 - 1300
  • [35] The role of biodegradable materials in the treatment of articular cartilage defects
    Bartha, Lajos
    Hangody, Laszlo
    Kaposi, Novak Pal
    Vajda, Andras
    EKLEM HASTALIKLARI VE CERRAHISI-JOINT DISEASES AND RELATED SURGERY, 2007, 18 (03): : 101 - 107
  • [36] TGF-β1 Improves Articular Cartilage Damage in Rabbit Knee
    Singh, N. K.
    Shiwani, S.
    Singh, G. R.
    Jeong, D. K.
    Kinjavdekar, P.
    Amarpal
    Lohakare, J. D.
    Lee, S. J.
    PAKISTAN VETERINARY JOURNAL, 2012, 32 (03) : 412 - 417
  • [37] Evaluation of Articular Cartilage Progenitor Cells for the Repair of Articular Defects in an Equine Model
    Frisbie, David D.
    McCarthy, Helen E.
    Archer, Charles W.
    Barrett, Myra F.
    McIlwraith, C. Wayne
    JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME, 2015, 97A (06): : 484 - 493
  • [38] Repair of rabbit focal articular cartilage defects with autologous chondrocytes embedded in alginate
    Masjudin, T.
    Chong, Pan-Pan
    Kamarul, T.
    Selvaratnam, L.
    Ab-Rahim, S.
    Sara, T.
    3RD KUALA LUMPUR INTERNATIONAL CONFERENCE ON BIOMEDICAL ENGINEERING 2006, 2007, 15 : 700 - +
  • [39] REPAIR OF SHEEP ARTICULAR-CARTILAGE DEFECTS WITH A RABBIT COSTAL PERICHONDRIAL GRAFT
    HOMMINGA, GN
    BULSTRA, SK
    KUIJER, R
    VANDERLINDEN, AJ
    ACTA ORTHOPAEDICA SCANDINAVICA, 1991, 62 (05): : 415 - 418
  • [40] Effect of low-energy shock waves in microfracture holes in the repair of articular cartilage defects in a rabbit model
    WANG QiLI ZhongliFU YangmuWANG ZhigangWEI MinZHAO BinZHANG Li and ZHU Juanli Department of OrthopedicsSports Medicine CenterPeoples Liberation Army General HospitalBeijing China
    中华医学杂志(英文版), 2011, (09) : 1386 - 1394