Transdermal Delivery of Fentanyl from Matrix and Reservoir Systems: Effect of Heat and Compromised Skin

被引:60
|
作者
Prodduturi, Suneela [2 ]
Sadrieh, Nakissa [1 ]
Wokovich, Anna M. [1 ]
Doub, William H. [1 ]
Westenberger, Benjamin J. [1 ]
Buhse, Lucinda [1 ]
机构
[1] US FDA, Div Pharmaceut Anal, St Louis, MO 63101 USA
[2] Aizant Drug Res Solut Private Ltd, Hyderabad 50014, Andhra Pradesh, India
关键词
fentanyl; transdermal drug delivery systems; matrix; reservoir; compromised skin; skin permeation;
D O I
10.1002/jps.22004
中图分类号
R914 [药物化学];
学科分类号
100701 ;
摘要
The United States Food and Drug Administration (FDA) has received numerous reports of serious adverse events, including death, in patients using fentanyl transdermal systems (FTS). To gain a better understanding of these problems, the current research focuses on the in vitro characterization of fentanyl reservoir (Duragesic (R)) and matrix (Mylan) systems with respect to drug release and skin permeation under conditions of elevated temperature and compromised skin. In addition, different synthetic membrane barriers were evaluated to identify the one that best simulates fentanyl skin transport, and thus may be useful as a model for these systems in future studies. The results indicate that reservoir and matrix FTS are comparable when applied to intact skin at normal skin temperature but the kinetics of drug delivery are different in the two systems. At 40 degrees C, the permeation rate of fentanyl was twice that seen at 32 degrees C over the first 24h in both systems; however, the total drug permeation in 72 h is significantly higher in the reservoir FTS. When applied to partially compromised skin, matrix FTS has a greater permeation enhancement effect than reservoir FTS. The intrinsic rate limiting membrane of the reservoir system served to limit drug permeation when the skin (barrier) permeability was compromised. Different ethylene vinyl acetate membranes were shown to have fentanyl permeability values encompassing the variability in human skin. Results using the in vitro model developed using synthetic membranes suggest that they mimic the effect of compromised skin on fentanyl permeability. Especially for highly potent drugs such as fentanyl, it is important that patients follow instructions regarding application of heat and use of the product on compromised skin. (C) 2009 Wiley-Liss, Inc. and the American Pharmacists Association J Pharm Sci 99:2357-2366, 2010
引用
收藏
页码:2357 / 2366
页数:10
相关论文
共 50 条
  • [41] Pharmacokinetics of a New Fentanyl Tape with a Novel Delivery System of Transdermal Matrix Patches in Patients with Cancer Pain
    Saito, Takefumi
    Fujii, Masato
    Saito, Shiro
    Shimada, Ken
    Fujiwara, Keiichi
    Kobayashi, Kunihiko
    ONCOLOGY, 2014, 86 (01) : 10 - 15
  • [42] On the Road to Development of an in Vitro Permeation Test (IVPT) Model to Compare Heat Effects on Transdermal Delivery Systems: Exploratory Studies with Nicotine and Fentanyl
    Shin, Soo Hyeon
    Ghosh, Priyanka
    Newman, Bryan
    Hammell, Dana C.
    Raney, Sam G.
    Hassan, Hazem E.
    Stinchcomb, Audra L.
    PHARMACEUTICAL RESEARCH, 2017, 34 (09) : 1817 - 1830
  • [43] On the Road to Development of an in Vitro Permeation Test (IVPT) Model to Compare Heat Effects on Transdermal Delivery Systems: Exploratory Studies with Nicotine and Fentanyl
    Soo Hyeon Shin
    Priyanka Ghosh
    Bryan Newman
    Dana C. Hammell
    Sam G. Raney
    Hazem E. Hassan
    Audra L. Stinchcomb
    Pharmaceutical Research, 2017, 34 : 1817 - 1830
  • [44] MORE THAN SKIN DEEP - TRANSDERMAL DRUG-DELIVERY SYSTEMS
    LASAGNA, L
    GREENBLATT, DJ
    NEW ENGLAND JOURNAL OF MEDICINE, 1986, 314 (25): : 1638 - 1639
  • [45] Skin irritation in transdermal drug delivery systems: A strategy for its reduction
    Kawahara, Koji
    Tojo, Kakuji
    PHARMACEUTICAL RESEARCH, 2007, 24 (02) : 399 - 408
  • [46] Functional nano-systems for transdermal drug delivery and skin therapy
    Liu, Lijun
    Zhao, Wenbin
    Ma, Qingming
    Gao, Yang
    Wang, Weijiang
    Zhang, Xuan
    Dong, Yunxia
    Zhang, Tingting
    Liang, Yan
    Han, Shangcong
    Cao, Jie
    Wang, Xinyu
    Sun, Wentao
    Ma, Haifeng
    Sun, Yong
    NANOSCALE ADVANCES, 2023, 5 (06): : 1527 - 1558
  • [47] Transdermal drug delivery systems and skin sensitivity reactions: Incidence and management
    Murphy M.
    Carmichael A.J.
    American Journal of Clinical Dermatology, 2000, 1 (6) : 361 - 368
  • [48] Transdermal Delivery Systems of Natural Products Applied to Skin Therapy and Care
    Cheng, Ying-Chen
    Li, Tzong Shiun
    Su, Hong Lin
    Lee, Po Chun
    Wang, Hui-Min David
    MOLECULES, 2020, 25 (21):
  • [49] Skin Irritation in Transdermal Drug Delivery Systems: A Strategy for its Reduction
    Koji Kawahara
    Kakuji Tojo
    Pharmaceutical Research, 2007, 24
  • [50] Accidental Occupational Exposure to a Large Volume of Liquid Fentanyl on a Compromised Skin Barrier with No Resultant Effect
    Feldman, Ryan
    Weston, Benjamin W.
    PREHOSPITAL AND DISASTER MEDICINE, 2022, 37 (04) : 550 - 552