Development of a bio-magnetic measurement system and sensor configuration analysis for rats

被引:1
|
作者
Kim, Ji-Eun [1 ,2 ]
Kim, In-Seon [3 ]
Kim, Kiwoong [1 ,2 ]
Lim, Sanghyun [1 ,2 ]
Kwon, Hyukchan [1 ]
Kang, Chan Seok [1 ,4 ]
Ahn, San [1 ]
Yu, Kwon Kyu [1 ]
Lee, Yong-Ho [1 ]
机构
[1] KRISS, Ctr Biosignals, Daejeon, South Korea
[2] UST, Dept Med Phys, Daejeon, South Korea
[3] KRISS, Ctr Mass & Related Quant, Daejeon, South Korea
[4] Inst for Basic Sci Korea, Ctr Underground Phys, Daejeon, South Korea
来源
REVIEW OF SCIENTIFIC INSTRUMENTS | 2017年 / 88卷 / 04期
关键词
SQUID; CORTEX; MAGNETOMETER; DEVICE; MEG;
D O I
10.1063/1.4979071
中图分类号
TH7 [仪器、仪表];
学科分类号
0804 ; 080401 ; 081102 ;
摘要
Magnetoencephalography (MEG) based on superconducting quantum interference devices enables the measurement of very weak magnetic fields (10-1000 fT) generated from the human or animal brain. In this article, we introduce a small MEG system that we developed specifically for use with rats. Our system has the following characteristics: (1) variable distance between the pick-up coil and outer Dewar bottom (similar to 5 mm), (2) small pick-up coil (4 mm) for high spatial resolution, (3) good field sensitivity (45 similar to 80 fT/cm/root Hz), (4) the sensor interval satisfies the Nyquist spatial sampling theorem, and (5) small source localization error for the region to be investigated. To reduce source localization error, it is necessary to establish an optimal sensor layout. To this end, we simulated confidence volumes at each point on a grid on the surface of a virtual rat head. In this simulation, we used locally fitted spheres as model rat heads. This enabled us to consider more realistic volume currents. We constrained the model such that the dipoles could have only four possible orientations: the x- and y-axes from the original coordinates, and two tangentially layered dipoles (local x- and y-axes) in the locally fitted spheres. We considered the confidence volumes according to the sensor layout and dipole orientation and positions. We then conducted a preliminary test with a 4-channel MEG system prior to manufacturing the multi-channel system. Using the 4-channel MEG system, we measured rat magnetocardiograms. We obtained well defined P-, QRS-, and T-waves in rats with a maximum value of 15 pT/cm. Finally, we measured auditory evoked fields and steady state auditory evoked fields with maximum values 400 fT/cm and 250 fT/cm, respectively. Published by AIP Publishing.
引用
收藏
页数:11
相关论文
共 50 条
  • [1] Development of integrated magnetic tunnel junctions for detection of bio-magnetic field
    Nishikawa, T.
    Oogane, M.
    Fujiwara, K.
    Kato, D.
    Naganuma, H.
    Ando, Y.
    [J]. Transactions of Japanese Society for Medical and Biological Engineering, 2014, 52
  • [2] Nested Magnetic Field Compensation With Regulated Coefficients for Bio-Magnetic Field Measurement
    Wei, Yutong
    Cheng, Longsheng
    Niu, Yaqiong
    Chen, Sitong
    Ye, Chaofeng
    [J]. IEEE TRANSACTIONS ON INSTRUMENTATION AND MEASUREMENT, 2023, 72
  • [3] Crosstalk analysis and suppression of optically pumped magnetometer array for bio-magnetic field measurement systems
    Yuan, Ziqi
    Liu, Ying
    Lin, Shudong
    Cao, Li
    Tang, Junjian
    Lei, Gaoyi
    Zhai, Yueyang
    [J]. MEASUREMENT, 2024, 237
  • [4] Fabrication of magnetic tunnel junctions with amorphous CoFeSiB for the bio-magnetic field sensor devices
    Kato, Daiki
    Oogane, Mikihiko
    Fujiwara, Kosuke
    Nishikawa, Takuo
    Naganuma, Hiroshi
    Ando, Yasuo
    [J]. Transactions of Japanese Society for Medical and Biological Engineering, 2014, 52
  • [5] AN INTEGRATED CHIP COIL SENSOR AND INSTRUMENTATION AMPLIFIER FOR BIO-MAGNETIC SIGNAL ACQUISITION
    Nair, Vijith Vijayakumaran
    Youn, Jin Hwan
    Choi, Jun Rim
    [J]. 2015 TRANSDUCERS - 2015 18TH INTERNATIONAL CONFERENCE ON SOLID-STATE SENSORS, ACTUATORS AND MICROSYSTEMS (TRANSDUCERS), 2015, : 1581 - 1584
  • [6] Spatial-temporal analysis of fetal bio-magnetic signals
    Soni, P.
    Chan, Y.
    Eswaran, H.
    Wilson, J. D.
    Murphy, P.
    Lowery, C. L.
    [J]. JOURNAL OF NEUROSCIENCE METHODS, 2007, 162 (1-2) : 333 - 345
  • [7] Bio-Magnetic/Eddy-Current Sensor Design for Biological Object Detection
    Lin, Chun-Yeon
    Wu, Yi-Chin
    Chen, Yuan-Liang
    Huang, Shih-Cheng
    [J]. 2020 IEEE/ASME INTERNATIONAL CONFERENCE ON ADVANCED INTELLIGENT MECHATRONICS (AIM), 2020, : 18 - 23
  • [8] Development of pico tesla resolution amorphous wire magneto-impedance sensor for bio-magnetic field measurements
    Uchiyama, Tsuyoshi
    Ma, Jiaju
    [J]. JOURNAL OF MAGNETISM AND MAGNETIC MATERIALS, 2020, 514
  • [9] Development of high resolution programmable oversampling MI sensor system with 32-bit ADC for multi-channel bio-magnetic measurements
    Ma, J.
    Yang, Z.
    Uchiyama, T.
    [J]. AIP ADVANCES, 2019, 9 (12)
  • [10] Numerical simulation of bio-magnetic nanofluid flow in the human circulatory system
    Jakeer, Shaik
    Shanmugapriyan, N.
    Reddisekhar Reddy, Seethi Reddy
    [J]. NUMERICAL HEAT TRANSFER PART A-APPLICATIONS, 2024,