QTR-FRET: Efficient background reduction technology in time-resolved forster resonance energy transfer assays

被引:4
|
作者
Syrjanpaa, Markku [1 ,4 ]
Vuorinen, Emmiliisa [2 ]
Kulmala, Sakari [3 ]
Wang, Qi [1 ]
Harma, Harri [2 ]
Kopra, Kari [2 ]
机构
[1] Univ Turku, Dept Cell Biol & Anat, Biophys Lab, Inst Biomed, Tykistokatu 6A, FI-20520 Turku, Finland
[2] Univ Turku, Dept Chem, Mat Chem & Chem Anal, Vatselankatu 2, FI-20500 Turku, Finland
[3] Aalto Univ, Dept Chem, Analyt Chem Lab, POB 16100, FI-00076 Aalto, Finland
[4] Abacus Diagnost Ltd, Tykistokatu 4D, FI-20520 Turku, Finland
基金
芬兰科学院;
关键词
Lanthanide chelate; Quenching resonance energy transfer (QRET); Streptavidin; Time-resolved forster resonance energy transfer (TR-FRET); KRAS; CONFORMATIONAL-CHANGES; BIOTIN; LABEL; STREPTAVIDIN; LUMINESCENCE; QRET;
D O I
10.1016/j.aca.2019.09.045
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
A novel homogeneous assay system QTR-FRET (Quencher modulated Time-Resolved Forster Resonance Energy Transfer) combining quenching resonance energy transfer (QRET) and time-resolved Forster resonance energy transfer (TR-FRET) was developed to reduce background signal in the conventional energy transfer applications. The TR-FRET functionality is often limited by the lanthanide donor background signal leading to the use of low donor concentration. QTR-FRET reduces this background by introducing soluble quencher molecule, and in this work the concept functionality was proven and compared to previously introduced QRET and TR-FRET technologies. Comparison was performed with three different Eu3+-chelates exhibiting different luminescent lifetime and stability. The side-by-side comparison of the three signaling systems and Eu3+ -chelates was demonstrated in a model assay with Eu3+-chelate conjugated biotin and streptavidin (SA) or Cy5-SA conjugate. Comparison of the methodologies showed increased signal-to-background ratios when comparing QTR-FRET to TR-FRET, especially at high Eu3+ -biotin concentrations. Quenching the non-bound Eu3+-biotin improved the assay performance, which suggests that an improved assay performance can be attained with the QTR-FRET method. QTR-FRET is expected to be especially useful for Eu3+-labeled ligands with low affinity or assays requiring high Eu3+-ligand concentration. The QTR-FRET indicated potential for multi-analyte approaches separately utilizing the direct QRET-type Eu3+-chelate signal and energy transfer signal readout in a single- well. This potential was hypothesized with Avi-KRAS nucleotide exchange assay as a second biologically relevant model system. (C) 2019 Elsevier B.V. All rights reserved.
引用
收藏
页码:93 / 101
页数:9
相关论文
共 50 条
  • [21] G Protein- Coupled Receptor Signaling Analysis Using Homogenous Time-Resolved Forster Resonance Energy Transfer ( HTRF ) Technology
    Norskov-Lauritsen, Lenea
    Thomsen, Alex Rojas Bie
    Brauner-Osborne, Hans
    INTERNATIONAL JOURNAL OF MOLECULAR SCIENCES, 2014, 15 (02) : 2554 - 2572
  • [22] Versatility of Homogeneous Time-Resolved Fluorescence Resonance Energy Transfer Assays for Biologics Drug Discovery
    Rossant, Christine J.
    Matthews, Carl
    Neal, Frances
    Colley, Caroline
    Gardener, Matthew J.
    Vaughan, Tristan
    JOURNAL OF BIOMOLECULAR SCREENING, 2015, 20 (04) : 508 - 518
  • [23] A 15-min non-competitive homogeneous assay for microcystin and nodularin based on time-resolved Forster resonance energy transfer (TR-FRET)
    Akter, Sultana
    Lamminmaki, Urpo
    ANALYTICAL AND BIOANALYTICAL CHEMISTRY, 2021, 413 (24) : 6159 - 6170
  • [24] Feasibility of single nucleotide polymorphism genotyping with a single-probe by time-resolved Forster resonance energy transfer
    Andreoni, Alessandra
    Bondani, Maria
    Nardo, Luca
    MOLECULAR AND CELLULAR PROBES, 2009, 23 (02) : 119 - 121
  • [25] Multicolor time-resolved Forster resonance energy transfer microscopy reveals the impact of GPCR oligomerization on internalization processes
    Faklaris, Orestis
    Cottet, Martin
    Falco, Amandine
    Villier, Brice
    Laget, Michel
    Zwier, Jurriaan M.
    Trinquet, Eric
    Mouillac, Bernard
    Pin, Jean-Philippe
    Durroux, Thierry
    FASEB JOURNAL, 2015, 29 (06): : 2235 - 2246
  • [26] Rapid Homogeneous Immunoassay Based on Time-Resolved Forster Resonance Energy Transfer for Serodiagnosis of Acute Hantavirus Infection
    Hepojoki, Satu
    Hepojoki, Jussi
    Hedman, Klaus
    Vapalahti, Olli
    Vaheri, Antti
    JOURNAL OF CLINICAL MICROBIOLOGY, 2015, 53 (02) : 636 - 640
  • [27] Quantification of HER Expression and Dimerization in Patients' Tumor Samples Using Time-Resolved Forster Resonance Energy Transfer
    Ho-Pun-Cheung, Alexandre
    Bazin, Herve
    Gaborit, Nadege
    Larbouret, Christel
    Garnero, Patrick
    Assenat, Eric
    Castan, Florence
    Bascoul-Mollevi, Caroline
    Ramos, Jeanne
    Ychou, Marc
    Pelegrin, Andre
    Mathis, Gerard
    Lopez-Crapez, Evelyne
    PLOS ONE, 2012, 7 (07):
  • [28] Time-Resolved Fluorescence Resonance Energy Transfer as a Versatile Tool in the Development of Homogeneous Cellular Kinase Assays
    Saville, Lisa
    Spais, Chrysanthe
    Mason, Jennifer L.
    Albom, Mark S.
    Murthy, Seetha
    Meyer, Sheryl L.
    Ator, Mark A.
    Angeles, Thelma S.
    Husten, Jean
    ASSAY AND DRUG DEVELOPMENT TECHNOLOGIES, 2012, 10 (06) : 551 - 557
  • [29] A time-resolved Forster resonance energy transfer assay to measure activity of the deamidase of the prokaryotic ubiquitin-like protein
    Eustis, Ian C.
    Huang, Jessica
    Pilkerton, Meagan E.
    Whedon, Samuel D.
    Chatterjee, Champak
    ANALYTICAL BIOCHEMISTRY, 2015, 487 : 27 - 29
  • [30] Time-resolved fluorescence and absence of Forster resonance energy transfer in ferroelectric liquid crystal-quantum dots composites
    Singh, D. P.
    Pandey, S.
    Manohar, R.
    Kumar, S.
    Pujar, G. H.
    Inamdar, S. R.
    JOURNAL OF LUMINESCENCE, 2017, 190 : 161 - 170