A new G-M counter dead time model

被引:47
|
作者
Lee, SH [1 ]
Gardner, RP [1 ]
机构
[1] N Carolina State Univ, Ctr Engn Applicat Radioisotopes, Raleigh, NC 27695 USA
关键词
G-M counter model; G-M counter dead time; dead time model;
D O I
10.1016/S0969-8043(00)00261-X
中图分类号
O61 [无机化学];
学科分类号
070301 ; 081704 ;
摘要
A hybrid G-M counter dead time model was derived by combining the idealized paralyzable and non-paralyzable models. The new model involves two parameters, which are the paralyzable and non-paralyzable dead times. The dead times used in the model are very closely related to the physical dead time of the G-M tube and its resolving time. To check the validity of the model, the decaying source method with Mn-56 was used. The corrected counting rates by the new G-M dead time model were compared with the observed counting rates obtained from the measurement and gave very good agreement within 5% up to 7 x 10(4) counts/s for a G-M tube with a dead time of about 300 mu s. (C) 2000 Published by Elsevier Science Ltd.
引用
收藏
页码:731 / 737
页数:7
相关论文
共 50 条
  • [1] A new G-M counter hybrid dead-time correction model
    Hou, Guojing
    Gardner, Robin P.
    RADIATION PHYSICS AND CHEMISTRY, 2015, 116 : 125 - 129
  • [2] G-M COUNTER DEAD TIME AND THE 2-SOURCE METHOD
    MANI, KV
    WEHRING, BW
    HEALTH PHYSICS, 1987, 52 : S14 - S14
  • [3] Non-poisson counting statistics of a hybrid G-M counter dead time model
    Lee, Sang Hoon
    Jae, Moosung
    Gardner, Robin P.
    NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION B-BEAM INTERACTIONS WITH MATERIALS AND ATOMS, 2007, 263 (01): : 46 - 49
  • [4] G-M COUNTER FOR LIQUIDS
    GRAVES, A
    NUCLEONICS, 1952, 10 (01): : 78 - 78
  • [5] THE EFFICIENCY OF THE G-M COUNTER
    SONODA, M
    JOURNAL OF THE PHYSICAL SOCIETY OF JAPAN, 1950, 5 (02) : 53 - 56
  • [6] An improved formula for dead time correction of G-M detectors
    Arkani, Masoomeh
    Khalafi, Hossein
    Arkani, Mohammad
    NUKLEONIKA, 2013, 58 (04) : 533 - 536
  • [7] A NOMOGRAM FOR G-M COUNTER RESOLVING-TIME CORRECTIONS
    ANDREWS, HL
    REVIEW OF SCIENTIFIC INSTRUMENTS, 1950, 21 (02): : 191 - 192
  • [8] CORRECTION OF G-M COUNTER DATA
    KURBATOV, JD
    MANN, HB
    PHYSICAL REVIEW, 1945, 68 (1-2): : 40 - 43
  • [9] Reliability of G-M counter tubes
    Li, J.Y.
    Lai, Y.F.
    Zou, B.R.
    Zhang, H.
    Hedianzixue Yu Tance Jishu/Nuclear Electronics & Detection Technology, 2001, 21 (02):
  • [10] A TSEE G-M GASFLOW COUNTER
    OBERHOFER, M
    ROBINSON, EM
    KERNTECHNIK, 1970, 12 (01) : 34 - +