Comparison of visual results and optical quality of two presbyopia-correcting intraocular lenses: TECNIS symfony versus TECNIS synergy

被引:15
|
作者
Shin, Da Eun [1 ]
Lee, Hun [2 ]
Kim, Tae-Im [3 ]
Koh, Kyungmin [1 ]
机构
[1] Konyang Univ, Kims Eye Hosp, Dept Opthalmol, Coll Med, 136 Youngshinro, Seoul 07301, South Korea
[2] Univ Ulsan, Asan Med Ctr, Dept Ophthalmol, Coll Med, Seoul, South Korea
[3] Yonsei Univ, Severance Eye Hosp, Dept Opthalmol, Coll Med, Seoul, South Korea
关键词
Defocus curve; extended depth of focus; extended range of vision; presbyopia; presbyopia-correcting intraocular lens; DEPTH-OF-FOCUS; EXTENDED DEPTH; PATIENT SATISFACTION; CLINICAL-OUTCOMES; IMPLANTATION; RANGE; PERFORMANCE; IOL;
D O I
10.1177/11206721221093024
中图分类号
R77 [眼科学];
学科分类号
100212 ;
摘要
Purpose: This study compared the extended depth of focus (EDOF) intraocular lens (IOL) (ZXR00; Tecnis (R) Symfony (R), Johnson & Johnson Vision, Santa Ana, CA, USA) and a new presbyopia-correcting IOL that combines EDOF and multifocal profiles (ZFR00; Tecnis (R) Synergy (R), Johnson & Johnson Vision, Santa Ana, CA, USA) with the same material. Methods: The medical records of patients undergoing cataract surgery with ZXR00 or ZFR00 implantation between March 2021 and July 2021 and with the data available for the 3-month visit were reviewed. The uncorrected near, intermediate, and distance visual acuity (VA), corrected distance VA, defocus curves, refractive errors (RE), optical quality parameters, and patient-reported visual performance parameters were the main outcome measures. Results: Among the twenty-three enrolled patients, twelve patients (21 eyes) received ZXR00 and nine patients (17 eyes) received ZFR00. The mean age of the patients was 59.6 +/- 10.6 (range: 49 to 70) and 65.2 +/- 8.2 (range: 45 to 82) years in ZXR00 and ZFR00 groups, respectively, without significant difference. There was not a significant difference between the two groups in terms of baseline parameters and postoperative RE. The ZFR00 group showed markedly superior intermediate and near VA (p < 0.001 for all) at 3 months postoperatively. In terms of optical quality, ZXR00 was significantly better than ZFR00 (p < 0.001). Conclusions: Both IOLs had comparable distance VA. ZFR00 IOL was superior for intermediate and near visual acuity, but ZXR00 IOL was better for optical quality. Research with more participants is needed to support this finding.
引用
收藏
页码:3461 / 3469
页数:9
相关论文
共 48 条
  • [1] Comparison of refractive and visual outcomes of 3 presbyopia-correcting intraocular lenses
    Ferreira, Tiago B.
    Ribeiro, Filomena J.
    Silva, Diana
    Matos, Ana Claudia
    Gaspar, Sylvia
    Almeida, Soraia
    JOURNAL OF CATARACT AND REFRACTIVE SURGERY, 2022, 48 (03): : 280 - 287
  • [2] Comparison of Five Presbyopia-Correcting Intraocular Lenses: Optical-Bench Assessment with Visual-Quality Simulation
    Labuz, Grzegorz
    Yan, Weijia
    Baur, Isabella D. D.
    Khoramnia, Ramin
    Auffarth, Gerd U. U.
    JOURNAL OF CLINICAL MEDICINE, 2023, 12 (07)
  • [3] Visual quality and performance following bilateral implantation of TECNIS Symfony intraocular lenses with or without micro-monovision
    Tan, Jilin
    Qin, Yan
    Wang, Changpeng
    Yuan, Shiman
    Ye, Jian
    CLINICAL OPHTHALMOLOGY, 2019, 13 : 1071 - 1077
  • [4] Comparison of Bilateral Reading Performance Among Two Presbyopia-Correcting Intraocular Lenses
    Benyoussef, Anas-Alexis
    Reboux, Noemi
    Cochener, Beatrice
    JOURNAL OF REFRACTIVE SURGERY, 2022, 38 (07) : 428 - +
  • [5] Performance of presbyopia-correcting intraocular lenses in distance optical bench tests
    Maxwell, W. Andrew
    Lane, Stephen S.
    Zhou, Fan
    JOURNAL OF CATARACT AND REFRACTIVE SURGERY, 2009, 35 (01): : 166 - 171
  • [6] In vitro optical quality comparison between the Mini WELL Ready progressive multifocal and the TECNIS Symfony
    Alberto Domínguez-Vicent
    Jose Juan Esteve-Taboada
    Antonio J. Del Águila-Carrasco
    Teresa Ferrer-Blasco
    Robert Montés-Micó
    Graefe's Archive for Clinical and Experimental Ophthalmology, 2016, 254 : 1387 - 1397
  • [7] Comparison of Clinical Outcomes, Visual Quality and Visual Function of Two Presbyopia-Correcting Intraocular Lenses Made from the Same Material, but with Different Design and Optics
    Novacek, Ladislav Viktor
    Nemcova, Marie
    Tyx, Katerina
    Lahodova, Kristyna
    Rejmont, Leos
    Rozsival, Pavel
    Studeny, Pavel
    JOURNAL OF CLINICAL MEDICINE, 2021, 10 (15)
  • [8] Comparative Analysis of the Clinical Outcomes of Two Toric Presbyopia-Correcting Intraocular Lenses
    Gabric, Kresimir
    Gabric, Nikica
    Pinero, David P.
    Gabric, Ivan
    OPHTHALMOLOGY AND THERAPY, 2024, 13 (03) : 775 - 790
  • [9] Comparative Analysis of the Clinical Outcomes of Two Toric Presbyopia-Correcting Intraocular Lenses
    Krešimir Gabrić
    Nikica Gabrić
    David P. Piñero
    Ivan Gabrić
    Ophthalmology and Therapy, 2024, 13 : 775 - 790
  • [10] Imaging quality of bifocal piggyback intraocular lens versus ReSTOR and TECNIS multifocal lenses
    Artigas, Jose M.
    Felipe, Adelina
    Diaz-Llopis, Manuel
    Garcia-Delpech, Salvador
    Navea, Amparo
    EUROPEAN JOURNAL OF OPHTHALMOLOGY, 2010, 20 (01) : 71 - 75