Ultra-fast quantitation of voriconazole in human plasma by coated blade spray mass spectrometry

被引:33
|
作者
Tascon, Marcos [1 ]
Gomez-Rios, German Augusto [1 ]
Reyes-Garces, Nathaly [1 ]
Poole, Justen [1 ]
Boyaci, Ezel [1 ]
Pawliszyn, Janusz [1 ]
机构
[1] Univ Waterloo, Dept Chem, Waterloo, ON, Canada
基金
加拿大自然科学与工程研究理事会;
关键词
Ambient mass spectrometry; Coated blade spray; SPME; Voriconazole; Small volumes; Biological fluid; SOLID-PHASE MICROEXTRACTION; OF-THE-ART; COMPLEX MATRICES; REAL-TIME; FUNGAL-INFECTIONS; CHROMATOGRAPHY; ANTIFUNGALS; VALIDATION; SAFETY; SERUM;
D O I
10.1016/j.jpba.2017.03.009
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
Voriconazole is a triazole broad-spectrum antifungal medication often used to treat fungal infections caused by Aspergillus and Fusarium species. One of the main challenges associated with the implementation of this medication is its narrow therapeutic concentration range, demonstrating toxicity at concentrations above 6 mu g/mL and limited efficacy at concentrations below 2 mu g/mL. As a result, methodologies which permit the rapid and accurate quantitation of voriconazole in patients are highly desirable. In this work two different approaches based on coated blade spray directly coupled to mass spectrometry (CBS-MS) are introduced; each enabling the quantitation of voriconazole in plasma samples with a simple and fast sample preparation and no chromatographic step. The first approach involves a rapid extraction (1 min) of the target analyte from 300 mu L of human plasma using conventional laboratory vessels (e.g. vial, 96-well plate). Alternatively, the second strategy consists of a 2 min extraction from a plasma droplet (10 mu L) placed on the coated area of the blade. Both procedures were successfully validated and good linearity (R-2 >= 0.998), accuracy (91-122%) and precision (<8%) were attained in the concentration range evaluated (0.1-50 mu g/mL). Moreover, very good results in terms of relative matrix effects were obtained given that the slopes of the calibration curves constructed in five different plasma lots exhibited relative standard deviation (RSD) values below 7%. Herein we demonstrated that CBS-MS is a technology suitable for the ultra-fast determination of voriconazole in human plasma samples. Indeed, the proposed methodology can be easily used either for routine drug monitoring or for in vitro pharmacokinetic studies in applications where very small sample volumes are available and great temporal resolution is needed. (C) 2017 Elsevier B.V. All rights reserved.
引用
收藏
页码:106 / 111
页数:6
相关论文
共 50 条
  • [41] Coulometry-assisted quantitation in spray ionization mass spectrometry
    Song, Xiaowei
    Chen, Hao
    Zare, Richard N.
    JOURNAL OF MASS SPECTROMETRY, 2021, 56 (04):
  • [42] A gas chromatography mass spectrometry method for the quantitation of total homocysteine in human plasma
    Pietzsch, J
    Julius, U
    Hanefeld, M
    ATHEROSCLEROSIS, 1997, 134 (1-2) : 165 - 165
  • [43] Quantitation of terameprocol in human plasma by liquid chromatography-tandem mass spectrometry
    Anders, Nicole M.
    Romo, Carlos G.
    Hemingway, Avelina
    Ahluwalia, Manmeet S.
    Rudek, Michelle A.
    JOURNAL OF PHARMACEUTICAL AND BIOMEDICAL ANALYSIS, 2022, 209
  • [44] Determination of free thyroid hormones in animal serum/plasma using ultrafiltration in combination with ultra-fast liquid chromatography-tandem mass spectrometry
    Tanoue, Rumi
    Kume, Imari
    Yamamoto, Yasuo
    Takaguchi, Kohki
    Nomiyama, Kei
    Tanabe, Shinsuke
    Kunisue, Tatsuya
    JOURNAL OF CHROMATOGRAPHY A, 2018, 1539 : 30 - 40
  • [45] Metabolite profiles of icariin in rat plasma by ultra-fast liquid chromatography coupled to triple-quadrupole/time-of-flight mass spectrometry
    Qian, Qian
    Li, Song-Lin
    Sun, E.
    Zhang, Ke-Rong
    Tan, Xiao-Bin
    Wei, Ying-Jie
    Fan, Hong-Wei
    Cui, Li
    Jia, Xiao-Bin
    JOURNAL OF PHARMACEUTICAL AND BIOMEDICAL ANALYSIS, 2012, 66 : 392 - 398
  • [46] Quantitation of Ursolic Acid in Human Plasma By Ultra-performance Liquid Chromatography Tandem Mass Spectrometry and Its Pharmacokinetic Study
    Wei, Guangli
    Xia, Yuanyuan
    Si, Duanyun
    Liu, Changxiao
    DRUG METABOLISM REVIEWS, 2010, 42 : 41 - 42
  • [47] Fast, very fast, and ultra-fast gas chromatography-mass spectrometry of thermally labile steroids, carbamates, and drugs in supersonic molecular beams
    Dagan, S
    Amirav, A
    JOURNAL OF THE AMERICAN SOCIETY FOR MASS SPECTROMETRY, 1996, 7 (08) : 737 - 752
  • [48] Ultra-fast ionization modeling in laser-plasma interaction
    d'Humieres, E.
    Sentoku, Y.
    Kemp, A.
    5TH INTERNATIONAL CONFERENCE ON INERTIAL FUSION SCIENCES AND APPLICATIONS (IFSA2007), 2008, 112
  • [49] Chemical profiling and multicomponents quantitative analysis of Panzerina lanata by ultra-fast liquid chromatography with tandem mass spectrometry
    Mi, Yingying
    Zheng, Wei
    Zhao, Xiangpei
    Zhang, Xun
    Yang, Kang
    Shen, Gangyi
    Liu, Weizhi
    Li, Wenrui
    Ma, Baiping
    Zhu, Dan
    JOURNAL OF SEPARATION SCIENCE, 2021, 44 (13) : 2524 - 2535
  • [50] Analysis of omarigliptin forced degradation products by ultra-fast liquid chromatography, mass spectrometry, and in vitro toxicity assay
    Mohr, Amanda
    Barbosa, Fabio de Souza
    Wingert, Nathalie Ribeiro
    Takeuchi, Carolina Kisara
    Garcia, Luiza
    Ribeiro, Maria Fernanda Nunes
    Arbo, Marcelo Dutra
    de Oliveira, Tiago Franco
    Steppe, Martin
    BIOMEDICAL CHROMATOGRAPHY, 2024, 38 (08)