Device-Related Infection Among Patients With Pacemakers and Implantable Defibrillators: Incidence, Risk Factors, and Consequences

被引:144
|
作者
Nery, Pablo B. [2 ]
Fernandes, Russell
Nair, Girish M.
Sumner, Glen L.
Ribas, Carlos S.
Menon, Syamkumar M. Divakara
Wang, Xiaoyin
Krahn, Andrew D. [3 ]
Morillo, Carlos A.
Connolly, Stuart J.
Healey, Jeff S. [1 ]
机构
[1] McMaster Univ, Dept Med, Hamilton, ON L8L2X2, Canada
[2] Univ Ottawa, Inst Heart, Ottawa, ON K1N 6N5, Canada
[3] Univ Western Ontario, London, ON N6A 3K7, Canada
基金
加拿大健康研究院;
关键词
complications; implantable cardioverter defibrillator; infection; pacemakers; surgery; CARDIOVERTER-DEFIBRILLATORS; PERMANENT PACEMAKER; MEDICARE BENEFICIARIES; COMPLICATIONS; ENDOCARDITIS; MANAGEMENT; DIAGNOSIS;
D O I
10.1111/j.1540-8167.2009.01690.x
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Methods: The authors analyzed the incidence and risk factors for cardiac device infection (CDI) among consecutive patients implanted with pacemaker (PM) or implantable cardioverter defibrillator (ICD) (including cardiac resynchronization therapy devices) at a tertiary health center in Hamilton, Ontario, Canada. Most patients with device-related infections were identified by an internal infection control system that reports any positive wound and blood cultures following surgery, between 2005 and the present. A retrospective review of patient records was also performed for all patients who received an ICD or PM between July 1, 2003 and March 20, 2007. Results: A total of 24 infections were identified among 2,417 patients having device surgery (1%). Fifteen of these infections (60%) were diagnosed within 90 days of the last surgical procedure. Univariate analysis showed that patients presenting with CDI were more likely to have had a device replacement, rather than a new implant, had more complex devices (dual/triple chamber vs single), and were more likely to have had a prior lead dislodgement. Multivariate analysis found device replacement (P = 0.02) and cardiac resynchronization therapy (CRT)/dual-chamber devices (P = 0.048) to be independent predictors of infection. One patient developed septic pulmonary emboli after having laser-assisted lead extraction. No patient died and 22 patients received a new device. Conclusion: CDI occurs in about 1% of cases in high volume facilities. Pulse generator replacement surgery and dual- or triple-chamber device implantation were associated with a significantly increased risk of infection. (J Cardiovasc Electrophysiol, Vol. pp. 786-790, July 2010).
引用
收藏
页码:786 / 790
页数:5
相关论文
共 50 条
  • [31] Risk Of Obstructive Sleep Apnea In Patients With Pacemakers And Implantable Cardioverter Defibrillators And Inicidence Of Ventricular Arrhythmia
    Clay, R.
    Kapur, V. K.
    Gronquist, J. M.
    Tereshchenko, L. G.
    Khan, A.
    [J]. AMERICAN JOURNAL OF RESPIRATORY AND CRITICAL CARE MEDICINE, 2016, 193
  • [32] Device-related complications in patients with cardiac resynchronization therapy and cardioverter defibrillators: experience of a center
    Brochado, B.
    Rodrigues, P.
    Anjo, D.
    Santos, M.
    Sousa, M. J.
    Vieira, P.
    Lagarto, V.
    Reis, H.
    Carvalho, H.
    Torres, S.
    [J]. EUROPEAN JOURNAL OF HEART FAILURE, 2013, 12 : S130 - S131
  • [33] INCIDENCE AND RISK FACTORS FOR MEDICAL DEVICE-RELATED PRESSURE INJURY IN INTENSIVE CARE UNIT UNSTABLE PATIENTS: A PROSPECTIVE COHORT STUDY
    Moraes, Juliano Teixeira
    Bueno, Luana Furtado
    Figueiredo, Manuela Mendonca
    Cortez, Daniel Nogueira
    [J]. JOURNAL OF WOUND OSTOMY AND CONTINENCE NURSING, 2022, 49 : S5 - S6
  • [34] Long-term risk of cardiovascular implantable electronic device-related infection after catheter ablation of atrial fibrillation
    Dinshaw, Leon
    Meyer, Christian
    [J]. JOURNAL OF CARDIOVASCULAR ELECTROPHYSIOLOGY, 2020, 31 (01) : 371 - 372
  • [35] Adaptation status and related factors in patients living with implantable cardioverter defibrillators
    Razavi, Mohammadreza
    Khatiban, Mahnaz
    Ahmadi, Fazloallah
    Oshvandi, Khodayar
    [J]. JOURNAL OF FAMILY MEDICINE AND PRIMARY CARE, 2022, 11 (08) : 4467 - 4472
  • [36] AneuRx device migration: Incidence, risk factors, and consequences
    Sampaio, SM
    Panneton, JM
    Mozes, G
    Andrews, JC
    Noel, AA
    Kalra, M
    Bower, TC
    Cherry, KJ
    Sullivan, TM
    Gloviczki, P
    [J]. ANNALS OF VASCULAR SURGERY, 2005, 19 (02) : 178 - 185
  • [37] FACTORS ASSOCIATED WITH DEPRESSION AMONG JAPANESE PATIENTS WITH IMPLANTABLE CARDIOVERTER DEFIBRILLATORS
    Ichikura, K.
    Matsuoka, S.
    Kobayashi, S.
    Suzuki, T.
    Nishimura, K.
    Shiga, T.
    Suzuki, S.
    Hagiwara, N.
    Ishigooka, J.
    [J]. INTERNATIONAL JOURNAL OF BEHAVIORAL MEDICINE, 2012, 19 : S46 - S46
  • [38] Breaking the Chain of Infection in Older Adults A Review of Risk Factors and Strategies for Preventing Device-Related Infections
    Schrank, Gregory
    Branch-Elliman, Westyn
    [J]. INFECTIOUS DISEASE CLINICS OF NORTH AMERICA, 2017, 31 (04) : 649 - +
  • [39] Risk factors for treatment failure in orthopedic device-related methicillin-resistant Staphylococcus aureus infection
    Ferry, T.
    Uckay, I.
    Vaudaux, P.
    Francois, P.
    Schrenzel, J.
    Harbarth, S.
    Laurent, F.
    Bernard, L.
    Vandenesch, F.
    Etienne, J.
    Hoffmeyer, P.
    Lew, D.
    [J]. EUROPEAN JOURNAL OF CLINICAL MICROBIOLOGY & INFECTIOUS DISEASES, 2010, 29 (02) : 171 - 180
  • [40] Risk factors for treatment failure in orthopedic device-related methicillin-resistant Staphylococcus aureus infection
    T. Ferry
    I. Uçkay
    P. Vaudaux
    P. François
    J. Schrenzel
    S. Harbarth
    F. Laurent
    L. Bernard
    F. Vandenesch
    J. Etienne
    P. Hoffmeyer
    D. Lew
    [J]. European Journal of Clinical Microbiology & Infectious Diseases, 2010, 29 : 171 - 180